Ti-Cron recallZ-1181-2023 · Covidien, LP · initiated Dec 2, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
FDA root cause category: Packaging process control
What the company is doing
Medtronic issued an "URGENT: MEDICAL DEVICE RECALL" letter to US consignees on 01-Dec-2022 via UPS 2-day delivery. OUS countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and actions to take: Actions: " Identify and quarantine all unused and non-expired affected non-absorbable sutures. " Please note the affected product may be located within Custom Suture Packs. Please reference Custom Suture Packs to identify affected product. " Please complete the enclosed Customer Confirmation Form (even if your account no longer has affected product on hand) and email to [email protected].> " Return all unused and non-expired affected product in your inventory to Medtronic as indicated in the Shipping and Return Instructions below. " Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed. Shipping and Return Instructions: " Product purchased directly from Medtronic please contact [email protected] for Return Good Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Credit or Replacement for the returned affected product will be issued based on the RGA number. " Return all unused and non-expired affected product(s) in your inventory to: Medtronic, Attn: Field Returns Dept. 195 McDermott Road, North Haven, CT 06473 USA. " If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. If you have any questions regarding this communication, please contact your Medtronic Representative or Customer Service at 800-962-9888.
Recall details
- Recall number
- Z-1181-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 2, 2022
- Posted by the FDA
- Feb 25, 2023
- Recalling firm
- Covidien, LP
- Quantity
- 40152 units
- Product code
- GAT
- FDA submissions
- K990952
- Distribution
- US Nationwide Distribution and OUS Foreign distribution to countries of: Albania Argentina Armenia Australia Austria Azerbaijan Belarus Belgium Bosnia And Herzegovina Brazil Bulgaria Cambodia Cameroon Canada Canary Islands Chile China Colombia Croatia Czech Republic Denmark Egypt Finland France Georgia Germany Ghana Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy JAPAN Jordan Kazakhstan Kenya Korea, Republic Of Kosovo Kuwait Latvia Lebanon Lithuania Macao Malaysia Malta Mauritius Mexico Moldova, Republic Of Morocco Myanmar Namibia Netherlands New Zealand North Macedonia Norway Pakistan Panama Paraguay Philippines Poland Portugal Reunion Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syrian Arab Republic Taiwan Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Viet Nam Zimbabwe
Product description
Ti-Cron Coated Braided Polyester Suture Product Description 88862775-31 TICRON* 4-0 BLU 75CM CV327DA 88862880-41 TICRON* 3-0 WHI 75CM Y31DA 88863012-61 TICRON* 0 BLU 90CM C16 X36 88863015-61 TICRON* 0 BLU 120CM V20X36 88863017-51 TICRON* 2-0 BLU 120CM SKX36 88863026-71 TICRON* 1 BLU 75CM HOS12X36 88863035-51 TICRON 2-0 BLU 105CM CV305DA 88863047-41 TICRON* 3-0 BLU 45CM C13X36 88863048-51 TICRON* 2-0 BLU 75CM C13X36 88863050-61 TICRON* 0 BLU 75CM C14 X36 88863054-51 TICRON* 2-0 BLU 75CM C17X36 88863056-89 TICRON* 5 BLU 45CM GS18X12 88863059-53 TICRON* 2-0 BLU 45CM P24X12 88863070-51 TICRON* 2-0 BLU 75CM C15X36 88863088-51 TICRON* 2-0 BLU 75CM Y5X36 88863090-51 TICRON 2-0 BLU 75CM GS21X36 88863092-71 TICRON* 1 BLU 75CM KV37X36 88863109-43 TICRON* 3-0 BLU 45CM P22X12 88863111-79 TICRON* 5 BLU 75CM HGS21X36 88863119-61 TICRON* 0 BLU 75CM KV34X36 88863154-09 TICRON* 4 BLU 75CM GS11X36 88863159-31 TICRON* 4-0 BLU 90CM KV5DA 88863160-41 TICRON* 3-0 BLU 90CM Y31DA 88863163-61 TICRON* 0 BLU 75CM SC1 X36 88863185-41 TICRON* 3-0 BLU 75CM CV305 88863186-41 TICRON* 3-0 BLU 75CM CV331 88863212-51 TICRON* 2-0 BLU 90CM Y5DA7P 88863226-41 TICRON* 3-0 BLU 60CM CV330DA 88863280-21 TICRON* 5-0 BLU 75CM CV301DA 88863280-31 TICRON* 4-0 BLU 75CM CV301DA 88863309-71 TICRON* 1 BLU 75CM GS11X36 88863369-31 TICRON* 4-0 BLU 90CM CV316DA 88863393-51 TICRON* 2-0 BLU 90CM CV300DA CD-3123K TICRON* 5-0 WHI 45CM SS24 DA
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
GTIN: 20884521062303, 20884521059570, 20884521060101, 20884521059006, 20884521059396, 20884521060682, 20884521059013, 10884521059290, 20884521059297, 20884521059280, 10884521061194, 20884521061191, 20884521062327, 20884521059204, 20884521060637, 20884521105192, 20884521062921, 10884521058620, 20884521058627, 20884521060170, 20884521059471, 10884521061293, 20884521061290, 20884521061641, 20884521059020, 10884521060432, 20884521060439, 20884521104300, 20884521063034, 20884521062655, 10884521062320, 10884521059900, 20884521059907, 20884521060361, 20884521059372, 20884521059990, 20884521060095, 20884521059464, 20884521062488, 10884521061507, 20884521061504. Model Number Lot # 8886277531 D1M0276Y 8886288041 D1L1698Y 8886301261 D1D3149FY 8886301561 D1E1922Y 8886301751 D2A0889Y 8886302671 D2C1802Y 8886303551 D1L0984Y 8886304741 D1G1844FY 8886304851 D1G0895FY 8886305061 D2C1969FY 8886305451 D1D3218FY 8886305689 D2D0199Y 8886305953 D1M1287FY 8886307051 D1D2540FY 8886308851 D2B2868Y 8886309051 D2B1332Y 8886309271 D2C2450Y 8886310943 D2C2451FY 8886311179 D1D2597Y 8886311961 D1M1166Y 8886315409 D2B2884Y 8886315931 D2B0542Y 8886316041 D1L1697Y D2B0543Y 8886316361 D1E0506Y 8886318541 D1J1639Y D1M2717Y D1M2851Y 8886318641 D1M2844Y D2A0933Y D2E0947Y D2E1925Y 8886321251 D2E3008Y 8886322641 D2A0451Y 8886328021 D1D0290Y 8886328031 D1L0721Y D1L2169Y D2B2861Y 8886330971 D2D2188Y 8886336931 D2B2738Y D2E0660Y D2F0767Y 8886339351 D1M0107Y D1M1299Y CD3123K D2B2213Y
Devices named in this recall
33 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ti-Cron 88863070-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls4 reports
- Ti-Cron 88863309-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls4 reports
- Ti-Cron 88863092-71Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls0 reports
- Ti-Cron 88863185-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls0 reports
- Ti-Cron 88863226-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE2 recalls0 reports
- Ti-Cron 88863015-61Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall18 reports
- Ti-Cron 88863047-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall2 reports
- Ti-Cron 88863088-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall2 reports
- Ti-Cron 88863186-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall1 report
- Ti-Cron 88863393-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall1 report
- Ti-Cron 88862775-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863012-61Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863017-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863035-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863048-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863050-61Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863054-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863056-89Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863059-53Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863109-43Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863111-79Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863119-61Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863154-09Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863159-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863160-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863163-61Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863212-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863280-21Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863280-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863369-31Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron CD-3123KCovidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88863090-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Ti-Cron 88862880-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
Product lines
Other devices under the same FDA submission
47 devices cleared under K990952 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Novafil 8886 4410-03Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall68 reports
- Novafil 8886 4401-23Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall12 reports
- Novafil 8886 4410-13Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall4 reports
- Novafil 8886 4452-51Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall4 reports
- Novafil SPB-1233GCovidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall4 reports
- Novafil SPB-1213GCovidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall2 reports
- Novafil SPB-5223GCovidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall2 reports
- Novafil 8886 4402-33Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall1 report
- Novafil SPB-5143GCovidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall1 report
- Novafil 8886 4002-41Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Novafil 8886 4403-33Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
- Novafil 8886 4422-43Covidien LP · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91324).
- Z-1172-2023 · Open, ClassifiedDermalon Monofilament Nylon Suture: (1) 88861741-11 DERMALON* 6-0 BLU 45CM C1 (2) 88861747-11 DERMALON* 6-0 BLU 45CM C12 (3) 88861749-24 DERMALON* 5-0 BLU 45CM P13 (4) 88861756-11…
- Z-1173-2023 · Open, ClassifiedMonofilament Nylon Suture (McKesson Branded) and Coated Braided Silk Suture (McKesson Branded) (1) S661GX 5-0 NYLON 18 DS18 (2)S683GX 4-0 SILK 18 DS18 (3) S683GX…
- Z-1174-2023 · Open, ClassifiedMonosof Monofilament Nylon sutures: Product Description CN-724 MONOSOF* 1 BLK 100CM GS25 X36 CN-791 MONOSOF* 2-0 BLK 100CM GCC90X36 CN-791-12 MONOSOF* 2-0 BLK 10…
- Z-1175-2023 · Open, ClassifiedNovafil Monofilament Polybutester Suture Product Description: 8886 4002-41 NOVAFIL 3-0 BLU 90CM C14 8886 4400-13 NOVAFIL 6-0 BLU 45CM SBE2 8886 4400-23 NOVAFIL 5-0 BLU 45CM SBE2 8…
- Z-1176-2023 · Open, ClassifiedSofsilk Coated Braided Silk Suture Product Description CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 CS211 CS-490 SOFSILK* 2-0 BLK 75CM GS11 X36 CS490 CS-562 SOFSILK* 2-0 BLK 75C…
- Z-1177-2023 · Open, ClassifiedSurgidac Uncoated Braided Polyester suture Product Description D-1764K SURGIDAC* 5-0 WHI 45CM SS24DA
- Z-1178-2023 · Open, ClassifiedSurgilon Braided Nylon suture Product Description: 88861883-51 SURGILON* 2-0 BLK 75CM C16 88861915-71 SURGILON* 1 BLK 75CM GS11 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 …
- Z-1179-2023 · Open, ClassifiedSurgipro Monofilament Polypropylene Sutures Product Description CP-535 SURGIPRO* 1 BLU 100CM GS24 X36 CP-824 SURGIPRO* 1 BLU 150CM GS26 X24 CP-825 SURGIPRO* 2 BLU 150CM GS26 X24 VP…
- Z-1180-2023 · Open, ClassifiedSurgipro II Monofilament Polypropylene Sutures Product Description Model Number VP-421X SURGIPRO*II 4-0 90CM V20 DA7P VP421X VP-435-X SURGIPRO II 4-0 90CM KV5 DA7P VP435X VP-541X S…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page