Novafil recallZ-1175-2023 · Covidien, LP · initiated Dec 2, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

FDA root cause category: Packaging process control

What the company is doing

Medtronic issued an "URGENT: MEDICAL DEVICE RECALL" letter to US consignees on 01-Dec-2022 via UPS 2-day delivery. OUS countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and actions to take: Actions: " Identify and quarantine all unused and non-expired affected non-absorbable sutures. " Please note the affected product may be located within Custom Suture Packs. Please reference Custom Suture Packs to identify affected product. " Please complete the enclosed Customer Confirmation Form (even if your account no longer has affected product on hand) and email to [email protected].> " Return all unused and non-expired affected product in your inventory to Medtronic as indicated in the Shipping and Return Instructions below. " Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed. Shipping and Return Instructions: " Product purchased directly from Medtronic please contact [email protected] for Return Good Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Credit or Replacement for the returned affected product will be issued based on the RGA number. " Return all unused and non-expired affected product(s) in your inventory to: Medtronic, Attn: Field Returns Dept. 195 McDermott Road, North Haven, CT 06473 USA. " If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. If you have any questions regarding this communication, please contact your Medtronic Representative or Customer Service at 800-962-9888.

Recall details

Recall number
Z-1175-2023
Status
Open, Classified
Initiated by the company
Dec 2, 2022
Posted by the FDA
Feb 25, 2023
Recalling firm
Covidien, LP
Quantity
48492 units
Product code
GAT
FDA submissions
K990952
Distribution
US Nationwide Distribution and OUS Foreign distribution to countries of: Albania Argentina Armenia Australia Austria Azerbaijan Belarus Belgium Bosnia And Herzegovina Brazil Bulgaria Cambodia Cameroon Canada Canary Islands Chile China Colombia Croatia Czech Republic Denmark Egypt Finland France Georgia Germany Ghana Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy JAPAN Jordan Kazakhstan Kenya Korea, Republic Of Kosovo Kuwait Latvia Lebanon Lithuania Macao Malaysia Malta Mauritius Mexico Moldova, Republic Of Morocco Myanmar Namibia Netherlands New Zealand North Macedonia Norway Pakistan Panama Paraguay Philippines Poland Portugal Reunion Romania Russian Federation Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Syrian Arab Republic Taiwan Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Viet Nam Zimbabwe

Product description

Novafil Monofilament Polybutester Suture Product Description: 8886 4002-41 NOVAFIL 3-0 BLU 90CM C14 8886 4400-13 NOVAFIL 6-0 BLU 45CM SBE2 8886 4400-23 NOVAFIL 5-0 BLU 45CM SBE2 8886 440123 NOVAFIL 5-0 BLU 45CM SBE3 8886 4402-33 NOVAFIL 4-0 BLU 45CM SBE4 8886 4403-33 NOVAFIL 4-0 BLU 45CM SBE6 8886 4410-13 NOVAFIL 6-0 BLU 45CM P10 8886 4410-03 NOVAFIL 7-0 BLU 45CM P10 8886 4422-43 NOVAFIL 3-0 BLU 45CM C14 8886 4424-51 NOVAFIL 2-0 BLU 75CM C16 8886 4450-41 NOVAFIL 3-0 BLU 75CM V26 8886 4452-51 NOVAFIL 2-0 BLU 75CM V20 8886 4459-61 NOVAFIL 0 BLU 75CM GS22 8886 4470-81 NOVAFIL* 2 BLU 150CM GS26 8886 4582-53 NOVAFIL* 2-0 BLU 75CM P14 SPB-1213G NOVAFIL* 6-0 BLU 45CM P13X12 SPB-1233G NOVAFIL* 4-0 BLU 45CM P13X12 SPB-1623G NOVAFIL* 5-0 BLU 45CM P12X12 SPB-5142G NOVAFIL* 5-0 BLU 45CM P11X12 SPB-5143G NOVAFIL* 4-0 BLU 45CM P11X12 SUT SPB-5223G NOVAFIL 5-0 BLU 45CM P13 SPB-5433G NOVAFIL* 4-0 CLR 45CM P24X12 SPB-5633G NOVAFIL* 4-0 BLU 45CM P12X12

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

GTIN 20884521100821, 20884521100098, 10884521100091, 10884521100299, 20884521100296, 20884521099910, 20884521100814, 20884521100791, 20884521100593, 20884521099965, 20884521099477, 20884521100760, 20884521118345, 10884521118348, 20884521099682, 20884521118406, 20884521099699, 20884521100838, 20884521099699, 20884521118390, 10884521118393, 20884521100890, 10884521100893, 20884521099569, 20884521100692, 20884521100074, 20884521099583, 20884521100876, 20884521100777. Model # Lot #: 8886400241 D1G1762FY, 8886440013 D2A0461Y, 8886440023 D1M0797Y, 8886440123 D1L3389Y, 8886440233 D2A0529Y, 8886440333 D2C0582Y, 8886441003 D1M1007FY, D1M2364FY 8886441013 D1M1007FY D1M2364FY, 8886441003 D1M0369FY D1M0796FY, 8886442243 D1D2485FY, 8886442451 D1E1240FY, 8886445041 D2B2109Y, 8886445251 D1D0295Y, D1E0843Y, 8886445961 D1D0529Y, 8886447081 D1D3190Y, 8886458253 D2B1686FY, SPB1213G D1L0169FY, D2B0773FY D2C0769FY, SPB1233G D2E0961FY, SPB1623G D1D2709FY, SPB5142G D1B2811FY, SPB5143G D1D2674FY, SPB5223G D1M1024FY, SPB5433G D2C1970FY, SPB5633G D2A0500FY, SPB-5223G D1M1024FY

Devices named in this recall

23 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

24 devices cleared under K990952 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 91324).