Skip to content

GE Nuclear Medicine 600/800 series systems, specifically recallZ-1057-2023 · Ge Medical Systems Israel, Functional Imaging · initiated Dec 19, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ge Medical Systems Israel, Functional Imaging or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.

FDA root cause category: Process control

What the company is doing

GE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION on 12/19/2022 by mail. The notice explained the issue and the risk, and requested the following action by taken: .GE Healthcare asks that you STOP USE of your Nuclear Medicine system, if your system is listed in the Affected Product Details and Appendix of this letter, until an inspection of your system is completed by GE Healthcare and you have been provided with a letter indicating the system can be returned to clinical use. .GE Healthcare will correct all affected products at no cost to you. GE Healthcare will inspect your system to ensure that the appropriate mitigation is correctly implemented to address this potential life-threatening concern. A GE Healthcare representative will contact you to arrange for this correction. .If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-1057-2023
Status
Open, Classified
Initiated by the company
Dec 19, 2022
Posted by the FDA
Feb 8, 2023
Recalling firm
Ge Medical Systems Israel, Functional Imaging
Quantity
86 units
Product code
KPS
FDA submissions
K153402
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Ethiopia, Finland, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uzbekistan, and Vietnam.

Product description

GE Nuclear Medicine 600/800 series systems, specifically: a) 870 CZT O540+ Gantry 7.25mm, b) D670 CZT Gantry 7.25mm, c) DISCOV NM/CT 870 CZT, d) Discovery NMCT 670 CZT O540+, e) NM/CT 870 CZT CYBER DEFENCE, f) NM/CT 870 CZT O540, g) NM/CT 870 CZT WSO, h) NM/CT 870 CZT, i) NM/CT 870 CZT O540+ 7.25MM; Emission Computed Tomography System

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

a) 870 CZT O540+ Gantry 7.25mm: Catalog number H3906CW , UDI/DI 01008406821408432187CZ6304211200300, Serial Numbers: 87CZ63042; UDI/DI 01008406821408432187CA6305311210100, Serial Numbers: 87CA63053; UDI/DI 01008406821408432187CX6301111181000, Serial Numbers: 87CX63011; UDI/DI 01008406821408432187CY6302011190300, Serial Numbers: 87CY63020; UDI/DI 01008406821408432187CY6303511191100, Serial Numbers: 87CY63035; UDI/DI 01008406821408432187CZ6304011200100, Serial Numbers: 87CZ63040; UDI/DI 01008406821408432187CX6300411180700, Serial Numbers: 87CX63004; UDI/DI 01008406821408432187CX6300911180900, Serial Numbers: 87CX63009; UDI/DI 01008406821408432187CB6307211220700, Serial Numbers: 87CB63072; UDI/DI 01008406821408432187CA6306111211100, Serial Numbers: 87CA63061; UDI/DI 01008406821408432187CY6303211191000, Serial Numbers: 87CY63032; UDI/DI To be provided, Serial Numbers: 87CB63075, 87CA63058, 87CB63070, 87CB63069, 87CB63071, 87CB63077, 87CA63063, 87CB63064, 87CB63078, 87CA63062, 87CB63067, 87CA63057, 87CA63060, 87CB63065, 87CB63068, 87CB63074; UDI/DI Not Applicable, Serial Numbers: 87CY63019, 87CA63054, 87CZ63043, 87CA63055, 87CY63028, 87CY63037, 87CZ63044, 87CZ63045, 87CZ63048, 87CZ63041, 87CY63024, 87CY63017, 87CX63007, 87CX63008, 87CX63006, 87CX63014, 87CY63022, 87CX63015, 87CY63025, 87CX63010, 87CZ63052, 87CY63027, 87CY63026, 87CY63030, 87CZ63046, 87CX63013, 87CY63036, 87CZ63047, 87CY63033, 87CY63016, 87CX63005, 87CZ63039, 87CX63002, 87CX63012, 87CX63001, 87CY63023, 87CY63018; b) D670 CZT Gantry 7.25mm: Catalog number H3906DD, UDI/DI Not Applicable, Serial Numbers: CZGA55003, CZGA55004, CZGY55001; , UDI/DI To be provided, Serial Numbers: CZGB55006, CZGA55005, CZGB55007; c) DISCOV NM/CT 870 CZT: Catalog number H3906CW, UDI/DI 01008406821408432187CY6303811191200, Serial Numbers: 87CY63038; d) Discovery NMCT 670 CZT O540+: Catalog number H3906DD , UDI/DI Not Applicable, Serial Numbers: CZGZ55002; e) NM/CT 870 CZT CYBER DEFENCE: Catalog number H3906FK, UDI/DI To be provided, Serial Numbers: 87DB82001; f) NM/CT 870 CZT O540: Catalog number H3906CW , UDI/DI Not Applicable, Serial Numbers: 87CY63021, 87CZ63050, 87CZ63049, 87CZ63051, 87CY63034; g) NM/CT 870 CZT WSO: Catalog number 5376204-70-20, UDI/DI 01008406821408432187CA6305911210800, Serial Numbers: 87CA63059; UDI/DI 01008406821408432187CB6307611221000, Serial Numbers: 87CB63076; UDI/DI 01008406821408432187CB6307311220800, Serial Numbers: 87CB63073; : Catalog number H3906CW, UDI/DI To be provided, Serial Numbers: 87CB63066; : Catalog number To be provided, UDI/DI To be provided, Serial Numbers: 87CA63056; h) NM/CT 870 CZT: Catalog number H3906CW, UDI/DI Not Applicable, Serial Numbers: 87CY63031; i) NM/CT 870 CZT O540+ 7.25MM: Catalog number H3906CV, UDI/DI 01008406821408432187CX6300311180700, Serial Numbers: 87CX63003; UDI/DI 01008406821408432187CY6302911190800, Serial Numbers: 87CY63029

Other devices under the same FDA submission

1 device cleared under K153402 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 91456).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page