NMCT recallZ-1058-2023 · Ge Medical Systems Israel, Functional Imaging · initiated Dec 19, 2022

Class I Open, Classified All recalls posted in February 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ge Medical Systems Israel, Functional Imaging or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.

FDA root cause category: Process control

What the company is doing

GE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION on 12/19/2022 by mail. The notice explained the issue and the risk, and requested the following action by taken: .GE Healthcare asks that you STOP USE of your Nuclear Medicine system, if your system is listed in the Affected Product Details and Appendix of this letter, until an inspection of your system is completed by GE Healthcare and you have been provided with a letter indicating the system can be returned to clinical use. .GE Healthcare will correct all affected products at no cost to you. GE Healthcare will inspect your system to ensure that the appropriate mitigation is correctly implemented to address this potential life-threatening concern. A GE Healthcare representative will contact you to arrange for this correction. .If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-1058-2023
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Dec 19, 2022
Posted by the FDA
Feb 8, 2023
Recalling firm
Ge Medical Systems Israel, Functional Imaging
Quantity
380 units
Product code
KPS
FDA submissions
K173816
Distribution
Worldwide distribution. US nationwide including Puerto Rico, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Ethiopia, Finland, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uzbekistan, and Vietnam.

Product description

GE Nuclear Medicine 600/800 series systems, specifically: a) 850, b) 850 / Dismount, c) 850 3/8 inch NM Gantry, d) 860 / Dismount, e) 860 3/8 inch NM Gantry, f) 860 5/8 inch NM Gantry, g) 860 BJ, h) NM 830 3/8 inch BJ, i) NM/CT 850 3/8 inch BJ, j) NM/CT 850 3/8 inch ROW, k) NM/CT 850 3/8 inch ROW_, l) NM/CT 850 CT 850 3/8 inch, m) NM/CT 850 ES 3/8 inch, n) NM/CT 850 ES 3/8 inch ROW, o) NM/CT 860 3/8 inch, p) NM/CT 860 3/8 inch BJ, q) NM/CT 860 3/8 inch ROW, r) NM/CT 860 ADVANCE, s) NM/CT 860 advance 3/8 inch NM Gantry, t) NM/CT 860 BEYOND, u) NM/CT 860 beyond 3/8 inch NM Gantry, v) NM/CT 860 China for China, w) NM/CT 860 CT 860 3/8 inch, x) NM/CT 860 ES 3/8 inch, y) NM/CT 860 ES 3/8 inch ROW, z) NM/CT 850 3/8 inch, aa) NM/CT 850 ES 3/8 inch ROW, bb) NM/CT 860 3/8 inch CH, cc) XELERIS 4 DR WORKSTATION; Emission Computed Tomography System

Code information (models, lots, serial numbers)

Show the full code information (10k characters)

a) 850: Catalog number 5376204-50-1, UDI/DI Not Applicable, Serial Numbers: NT3BG2200006BH; b) 850 / Dismount: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3BG2100003BH; c) 850 3/8 inch NM Gantry: Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 85EY67037, 850Z61033, 85EY67039, 85EY67048, 850Y61018, 850Y61009, 85EY67035, 85EX67001, 85EX67020, 85EY67045, 85EX67023, 85EY67032, 85EY67046, 850A61063, 850Z61047, 85EY67047, 850Z61053, 850Y61006, 850Y61010, 850X61005, 850Y61021, 85EZ67062, 85EX67019, 85EX67021, 85EA67063, 850A61069, 85EX67008, 85EY67040, 850Z61051, 850Z61038, 850A61075, 850Y61025, 850Z61030, 85EY67015, 85EX67017, 85EY67043, 850Z61042, 850Z61061, 85EY67026, 850A61065, 85EX67012, 850Y61023, 85EY67041, 850A61062, 850A61076, 850Y61012, 850A61064, 85EZ67058, 85EX67006, 850Z61043, 850Z61055, 85EY67042, 85EZ67055, 85EY67036, 85EX67024, 85EY67038, 85EX67022, 85EY67025, 85EZ67060, 850Z61058, 850Z61044, 850Z61045, 850Y61020, 850Z61032, 85EZ67056, 85EA67065, 85EY67028, 850Z61041, 85EX67002, 85EX67009, 850A61072, 850Z61036, 850A61068, 850A61071, 850Z61060, 850Z61034, 850Z61039, 850Z61035, 850Z61040, 85EY67031, 850Z61056, 850Z61057, 850Y61016, 85EX67016, 85EA67066, 85EX67004, 850Z61029, 850Z61031, 85EX67011, 85EX67010, 850Y61013, 850Y61015, 850Y61017, 850A61070, 85EX67005, 850A61073, 85EX67018, 85EY67033, 850Z61027, 85EX67013, 850Z61054, 850X61001, 85EZ67053, 85EZ67054, 85EZ67057, 850X61004, 850Z61059, 850Z61052, 850Y61014, 850Z61048, 850Y61026, 850Y61019, 85EY67044, 850Z61037, 850X61002, 85EX67003, 85EX67014, 850Y61007, 85EY67049, 85EZ67052, 850A61067, 850A61074, 85EY67050, 85EZ67051, 85EY67027; UDI/DI (01)00840682140775(21)850B61120(11)220700, Serial Numbers: 850B61120; UDI/DI 010084068214077521850Y6101111190300, Serial Numbers: 850Y61011; UDI/DI To be provided, Serial Numbers: 850A61083, 85EA67080, 850B61127, 850B61128, 850B61129, 850B61130, 85EB67087, 850B61133, 850B61134, 850B61135, 850B61101, 85EA67070, 850B61119, 850B61102, 850B61115, 85EA67076, CHBB65669, 850B61122, 850B61132, 850B61131, 85EA67075, 85EA67077, 85EB67088, 850A61093, 85EB67086, 850B61126, 850A61089, 85EA67078, 850B61112, 850B61111, 850B61110, 850B61107, 850B61109; Catalog number To be provided, UDI/DI To be provided, Serial Numbers: 60B62111; d) 860 / Dismount: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3CG2100006BH; e) 860 3/8 inch NM Gantry: Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 860Y62009, 860Y62010, 860Z62027, 860Y62005, 860A62049, 860Y62019, 860Y62014, 860Y62025, 860Z62028, 860Y62024, 860Z62038, 860A62063, 860Z62042, 860Y62017, 860A62065, 860A62058, 860Z62046, 860A62057, 860Z62041, 86EA76004, 860B62118, 860Z62043, 860A62060, 860Z62026, 860Y62006, 860Y62015, 860Y62004, 860Z62047, 860Y62020, 860A62050, 860Z62030, 860Y62007, 860X62002, 860Z62032, 860Y62021, 86EA76003, 860Z62040, 860Z62037, 860Z62031, 860A62062; , UDI/DI 010084068214075121860A6207711211000, Serial Numbers: 860A62077; UDI/DI (01)00840682140751(21)860B62097(11)220300, Serial Numbers: 860B62097; UDI/DI 010084068214075121860B6209611220300, Serial Numbers: 860B62096; UDI/DI 010084068214075121860A6205911210400, Serial Numbers: 860A62059; UDI/DI 010084068214075121860A6206811210600, Serial Numbers: 860A62068; UDI/DI 010084068214075121860Y6202311191100, Serial Numbers: 860Y62023; UDI/DI 010084068214075121860Y6201311190600, Serial Numbers: 860Y62013; UDI/DI To be provided, Serial Numbers: 860A62067, 860Y62016, 860B62117, 860B62100, 860B62105, 860B62090, 860A62072, 860B62108, 860B62110, 860A62078, 860A62083, 860B62094, 860A62066, 860A62084, 860A62071, 860A62070, 860A62073, 860A62069, 860B62095, 860A62086, 860A62081, 860A62085, 860B62099, 860B62115, 860A62082, 860B62114, 860A62074, 860Z62034, 86EB76006, 860B62122, 860B62119, 860B62121, 860B62089, 86EB76007, 860A62076, 860A62079, 860B62116, 86EB76009, 860B62120, 860B62093, 860A62080, 860B62098, 860B62091, 860B62103; f) 860 5/8 inch NM Gantry: Catalog number H3908AE, UDI/DI To be provided, Serial Numbers: 860X62001; , UDI/DI Not Applicable, Serial Numbers: 860Z62035; g) 860 BJ: Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3CG1900007BH, NT3CG1900008BH, NT3CG1900009BH; h) NM 830 3/8 inch BJ: Catalog number 5376204-50-1, UDI/DI To be provided, Serial Numbers: NT3AG1900007BH, NT3AG2100001BH, NT3AG2100006BH; i) NM/CT 850 3/8 inch BJ: Catalog number 5376204-50-1, UDI/DI Not Applicable, Serial Numbers: NT3BG2000003BH, NT3BG2100001BH, NT3BG2200001BH, NT3BG1900001BH, NT3BG2200002BH, NT3BG2000004BH, NT3BG2100004BH, NT3BG2200005BH; : Catalog number H2407AA, UDI/DI Not Applicable, Serial Numbers: NT3BG2000001BH; , UDI/DI To be provided, Serial Numbers: NT3BG2100002BH; : Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3BG2000002BH; j) NM/CT 850 3/8 inch ROW: Catalog number 5376204-50-1, UDI/DI Not Applicable, Serial Numbers: NT3BG2100005BH, NT3BG1900002BH; Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 850Y61022, 850Z61046, 850Y61008; UDI/DI To be provided, Serial Numbers: 850B61100; k) NM/CT 850 3/8 inch ROW: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 860Y62008; l) NM/CT 850 CT 850 3/8 inch: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 850A61092, 850B61108, 850B61096, 850A61080, 850Z61050, 850B61106, 850B61097, 850A61077, 850B61118, 860B62112; Catalog number 5376204-50, UDI/DI 010084068214077521850A6109511211200, Serial Numbers: 850A61095; UDI/DI 010084068214077521850A6108811210900, Serial Numbers: 850A61088; UDI/DI 010084068214077521850B6112111220700, Serial Numbers: 850B61121; UDI/DI 010084068214077521850B6110411220400, Serial Numbers: 850B61104; UDI/DI 010084068214077521850A6108711210900, Serial Numbers: 850A61087; UDI/DI 010084068214077521850A6108211210800, Serial Numbers: 850A61082; UDI/DI 010084068214077521850B6110511220400, Serial Numbers: 850B61105; Catalog number H3907AA, UDI/DI To be provided, Serial Numbers: 850A61081; Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 850Z61049, 850A61066; UDI/DI To be provided, Serial Numbers: 850B61098, 850B61099, 850A61084, 850B61123, 850B61113, 850B61117, 850B61116, 850A61090, 850A61079, 850A61091, 850B61114, 850A61078, 850A61086; : Catalog number To be provided, UDI/DI To be provided, Serial Numbers: 850B61125, 850B61103; m) NM/CT 850 ES 3/8 inch: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 85EA67068; Catalog number 5376204-50, UDI/DI 01008406821407752185EA6707411210800, Serial Numbers: 85EA67074, 85EA67067, 85EA67073; Catalog number 5376204-50-1, UDI/DI To be provided, Serial Numbers: NT3BG1900003BH; Catalog number H3907AA, UDI/DI To be provided, Serial Numbers: 85EA67071; Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 85EX67007; UDI/DI To be provided, Serial Numbers: 85EA67069, 85EA67072, 85EA67079, 85EB67084; n) NM/CT 850 ES 3/8 inch ROW: Catalog number H3907AD, UDI/DI Not Applicable, Serial Numbers: 85EZ67059, 85EA67064, 85EY67034; o) NM/CT 860 3/8 inch: Catalog number H3908AD, UDI/DI To be provided, Serial Numbers: 00000860Z62033; p) NM/CT 860 3/8 inch BJ: Catalog number H2408AA, UDI/DI To be provided, Serial Numbers: NT3CG2000001BH; Catalog number 5376204-60-1, UDI/DI Not Applicable, Serial Numbers: NT3CG2100001BH, NT3CG1900010BH, NT3CG2100003BH, NT3CG1900005BH, NT3CG1900003BH, NT3CG2100004BH, NT3CG1900006BH, NT3CG2100005BH, NT3CG2000004BH, NT3CG1900002BH, NT3CG2100002BH; q) NM/CT 860 3/8 inch ROW: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 00000860Y62018; Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 860A62056, 860Y62011, 860A62048, 860Z62039, 860Z62044, 860A62052, 860A62055, 860Y62012, 860Y62022, 860A62054; Catalog number To be provided, UDI/DI Not Applicable, Serial Numbers: NT3CG2000003BH; UDI/DI To be provided, Serial Numbers: 860B62102; r) NM/CT 860 ADVANCE: Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 860Z62045, 860A62064, 860A62061; UDI/DI To be provided, Serial Numbers: 860A62075; Catalog number H3908BB , UDI/DI Not Applicable, Serial Numbers: 8BBB30002, 8BBB30001; s) NM/CT 860 advance 3/8 inch NM Gantry: Catalog number H3908BB, UDI/DI To be provided, Serial Numbers: 8BBB30003, 8BBB30004; t) NM/CT 860 BEYOND: Catalog number H3908BC, UDI/DI Not Applicable, Serial Numbers: 8BCB40001; u) NM/CT 860 beyond 3/8 inch NM Gantry: Catalog number H3908BC, UDI/DI To be provided, Serial Numbers: 8BCB40002; v) NM/CT 860 China for China: Catalog number 5376204-60-1, UDI/DI Not Applicable, Serial Numbers: NT3CG2200004BH, NT3CG2200003BH, NT3CG2200002BH; , UDI/DI To be provided, Serial Numbers: NT3CG2200001BH; w) NM/CT 860 CT 860 3/8 inch: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 860B62109, 860Z62029, 860B62101, 860B62113; Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 860Z62036, 86EZ76001, 860A62053; UDI/DI To be provided, Serial Numbers: 860B62092, 860A62088; x) NM/CT 860 ES 3/8 inch: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 86EB76005; y) NM/CT 860 ES 3/8 inch ROW: Catalog number H3908AD, UDI/DI Not Applicable, Serial Numbers: 86EA76002; z) NM/CT 850 3/8 inch: Catalog number H3907AA, UDI/DI Not Applicable, Serial Numbers: 850Y61024; aa) NM/CT 850 ES 3/8 inch ROW: Catalog number H3907AH, UDI/DI Not Applicable, Serial Numbers: 85EY67030; bb) NM/CT 860 3/8 inch CH: Catalog number H2408AA, UDI/DI Not Applicable, Serial Numbers: NT3DG1900007BH; cc) XELERIS 4 DR WORKSTATION: Catalog number 5782013, UDI/DI To be provided, Serial Numbers: 850B61124, 850A61094

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 91456).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page