2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
8 other recalls cover devices cleared under the same FDA submission (K173816); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Dec 19, 2022Open, Classified
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
Z-1058-2023 · Root cause: Process control · FDA record ↗
- Feb 7, 2021Open, Classified
GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
Z-1545-2021 · Root cause: Device Design · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K173816. One submission often covers many models, so check the product and lot details in each record.
- Z-1546-2021 · Feb 7, 2021 · Open, ClassifiedNM/CT 860 Nuclear Medicine / CT Scanners
- Z-0853-2021 · Dec 10, 2020 · TerminatedGE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
- Z-0854-2021 · Dec 10, 2020 · TerminatedGE NM/CT 860 system Model Number H3908AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
- Z-2991-2020 · Aug 14, 2020 · TerminatedNM/CT 850 Model Number: H3907AD
- Z-2992-2020 · Aug 14, 2020 · TerminatedNM/CT 860, SPECT/CT
- Z-1262-2019 · Mar 20, 2019 · TerminatedNM/CT 850 Product Usage: The GE NM/CT 850 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It i
- Z-1263-2019 · Mar 20, 2019 · TerminatedNM/CT 860 Product Usage: The GE NM/CT 860 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It
- Z-1261-2019 · Mar 20, 2019 · TerminatedNM 830 Product Usage: The Discovery NM 630 is an all-purpose dual detector nuclear imaging system intended for general nuclear medicine imaging procedures for detection of radioi
Other System, x-ray, tomography, computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.