LATITUDE NXT Remote Patient Management System, LATITUDE NXT… recallZ-1216-2023 · Boston Scientific Corporation · initiated Feb 2, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022. For these patients, the HeartLogic Index and any associated Yellow Alerts are currently not visible in LATITUDE as may have been intended by the healthcare provider..
FDA root cause category: Software Manufacturing/Software Deployment
What the company is doing
The recalling firm issued letters to LATITUDE Healthcare Professionals via third-party Federal Express with signature at the account required on 2/2/2023. The letter informed the HCP that records indicate they may have U.S. patients who do not have HeartLogic enabled as the HCP may have intended in the Model 6460 U.S. LATITUDE NXT Patient Management System. The patients included within the enclosed list are scheduled to have HeartLogic enabled in LATITUDE ON 2/14/2023. The letter provides the clinical impact and recommendations. The recommendations include: (1) Distribute this information to appropriate HCPs within your center who manage HeartLogic alerts for the patients in the enclosed list; and (2) If you do NOT want HeartLogic enabled in LATITUDE for patients in the enclosed list, contact Boston Scientific LATITUDE Customer Support via [email protected] before 2/14/2023 and identify which patients should NOT be enabled. On 2/14/2023, BSC will correct the data on the Model 6460 LATITUDE Server for affected patients, unless a request not to enable is received by the firm prior to this date. Update: on 02/14/2023, Boston Scientific sent an additional letter, due to the original letter being lost in the mail. The revised letter advised customers to email LATITUDE Custoemr Support to have HeartLogic disabled.
Recall details
- Recall number
- Z-1216-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 2, 2023
- Posted by the FDA
- Mar 3, 2023
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 1,531 patient data records
- Product code
- LWS
- FDA submissions
- P910077
- Distribution
- US Nationwide distribution to the healthcare facilities and healthcare providers are located in CA, CO, DE, FL, GA, ID, IL, IN, IA, KY, ME, MD, MA, NV, NH, NJ, NY, NC, OH, PA, RI, SC, TN TX, VA, WA, WV, and WI.
Product description
LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
Code information (models, lots, serial numbers)
UDI-DI: 00802526613876
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- LATITUDE® NXT 6460Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components1 recall0 reports
Other devices under the same FDA submission
33 devices cleared under P910077 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 death65 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls9 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls4 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 report
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls1 report
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- ZOOM® LATITUDE® 3120Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)4 recalls0 reports
- LATITUDE® Communicator 6290Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components142 reports
- LATITUDE™ Programming System 3300Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components141 reports
- LATITUDE™ Programming System 3300Boston Scientific Corporation · Pacemaker/icd/crt non-implanted components72 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page