Impella Heart Pump recallZ-2128-2023 · Abiomed, Inc. · initiated May 26, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally introduced during manufacturing at the joint between the sidearm and the hub may cause the joint to detach.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT: VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) notification letter was delivered to customers on 5/23/23. Customers were also contacted via telephone on 5/26/23. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal). DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your Abiomed local clinical field staff to coordinate the return of the subject products. 3. Review, complete all fields, sign, and return the attached Business Response Form (BRF) on the last page of this letter to the Recall Coordinator identified in this document. 4. Forward this notice to any personnel in your facility who needs to be informed. 5. If any of the product subject to this recall (removal) has been forwarded to another facility, contact that facility immediately and provide a copy of this notice to the relevant personnel. 6. Post a copy of this field safety notice in a visible area for awareness. 7. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reportingprogram/reporting-serious-problems-fda If you have questions or concerns regarding this notice, please contact (Shashi Thoutam) directly at +1(734) 262-6255 and/or your local clinical field staff.
Recall details
- Recall number
- Z-2128-2023
- Status
- Open, Classified
- Initiated by the company
- May 26, 2023
- Posted by the FDA
- Jul 11, 2023
- Recalling firm
- Abiomed, Inc.
- Quantity
- 37 units
- Product code
- DYB
- FDA submissions
- K222113
- Distribution
- US Nationwide distribution in the states of MN, NJ, TN.
Product description
Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457
Code information (models, lots, serial numbers)
UDI-DI: 00813502013252; Lot Numbers: 1727033, 1725590
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Impella 1000457Abiomed, Inc. · Introducer, Catheter1 recall0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K222113 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Impella 1000434Abiomed, Inc. · Introducer, Catheter1 recall1 report
- Impella 1000534Abiomed, Inc. · Introducer, Catheter1 recall0 reports
- Impella 1000535Abiomed, Inc. · Introducer, Catheter1 recall0 reports
- Abiomed 14Fr Low Profile Introducer 1000458Abiomed, Inc. · Introducer, Catheter5 deaths46 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page