Medline Probe Cover Kits containing sterile Ultrasound Gel… recallZ-2159-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 18, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
FDA root cause category: Device Design
What the company is doing
Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline. Questions: please contact the Recall Department at 866-359-1704
Recall details
- Recall number
- Z-2159-2023
- Status
- Open, Classified
- Initiated by the company
- May 18, 2023
- Posted by the FDA
- Jul 20, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 55014 units
- Product code
- OEQ
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
Product description
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY DRP PK PEDI CATH, Model Number 00-401318M; b) ANGIOGRAPHY OR, Model Number CDS984944C; c) ANGIOGRAM, Model Number DYNJ0149444M; d) EP LAB DEVICE PACK-LF, Model Number DYNJ0160556D; e) VASCULAR ANGIO PACK-LF, Model Number DYNJ0201109AG; f) ANGIO PACK-LF, Model Number DYNJ0373279M; g) VISCERAL ANGIOGRAM PACK-LF, Model Number DYNJ0429121I; h) IPMC ANGIOGRAPHY PACK-LF, Model Number DYNJ0806254D; i) ANGIOGRAPHY PACK-LF, Model Number DYNJ0854485T; j) ANGIO DRAPE PACK, Model Number DYNJ17624B; k) DEPAUL ANGIO PACK, Model Number DYNJ19939M; l) EP ANGIOGRAPHY PACK, Model Number DYNJ24225K; m) ANGIOGRAPHY DRAPE PACK-LF, Model Number DYNJ28067G; n) DASHTI-YAO NEURO ANGIO-LF, Model Number DYNJ33457C; o) ANGIOGRAM TRAY, Model Number DYNJ33638K; p) ANGIO PACK, Model Number DYNJ34361; q) ANGIO/PICC PACK, Model Number DYNJ36049; r) IR MINOR ANGIO PACK, Model Number DYNJ37665B; s) ANGIO LEFT HEART PACK, Model Number DYNJ40632C; t) MV-IR ANGIO DRAPE PACK-LF, Model Number DYNJ41553D; u) CUSTOM ANGIO TRAY, Model Number DYNJ41649F; v) FS ANGIO, Model Number DYNJ42499C; w) ANGIO PACK, Model Number DYNJ42903B; x) ANGIO PACK, Model Number DYNJ43661B; y) ANGIOGRAPHY PACK, Model Number DYNJ44293D; z) ANGIO DRAPE PACK, Model Number DYNJ44320F; aa) ANGIOGRAM PACK, Model Number DYNJ45007; bb) ANGIO BASIN SET, Model Number DYNJ46043F; cc) ANGIO TRAY, Model Number DYNJ46153B; dd) ANGIOGRAM PACK, Model Number DYNJ47004C; ee) ANGIO PACK, Model Number DYNJ47226A; ff) ANGIOGRAPHY PACK, Model Number DYNJ47906; gg) ANGIO DRAPE SPEC PROC PACK-LF, Model Number DYNJ47951A; hh) MH HOOD CARDIAC ANGIO, Model Number DYNJ49884F; ii) ANGIO PACK, Model Number DYNJ50624A; jj) ANGIO PACK, Model Number DYNJ51239A; kk) ANGIO TRAY, Model Number DYNJ52121; ll) ANGIO PACK-LF, Model Number DYNJ52898C; mm) ANGIOGRAPHY PACK, Model Number DYNJ55252B; nn) ANGIOGRAPHY PACK, Model Number DYNJ57498; oo) PERCUTANEOUS ANGIO PACK-LF, Model Number DYNJ57645B; pp) GENERAL ANGIO PACK, Model Number DYNJ57760B; qq) ENSEMBLE ANGIOGRAPHIE-LF, Model Number DYNJ57792A; rr) ENSEMBLE ANGIOGRAPHIE-LF, Model Number DYNJ57792B; ss) ANGIO PACK, Model Number DYNJ58126A; tt) HH ANGIO PACK, Model Number DYNJ59465B; uu) OR ANGIO PACK, Model Number DYNJ60319A; vv) ANGIO PACK 2-A, Model Number DYNJ60507A; ww) FEMORAL ANGIOGRAPY SET UP, Model Number DYNJ61015B; xx) ANGIO PACK, Model Number DYNJ62299; yy) CLOVIS CATH LAB PACK, Model Number DYNJ62379; zz) ANGIO PACK, Model Number DYNJ62544; aaa) ANGIO TRAY W/CHLORAPREP, Model Number DYNJ62742; bbb) OR ANGIO PAC , Model Number DYNJ63976A; ccc) SUNSHINE PAC , Model Number DYNJ64830; ddd) ANGIO PAC , Model Number DYNJ65043A; eee) CATH LAB PAC , Model Number DYNJ65185; fff) ANGIOGRAPHY PACK, Model Number DYNJ65190C; ggg) PACK,BASIC ANGIO, Model Number DYNJ65482A; hhh) ANGIO PACK, Model Number DYNJ66398B; iii) ANGIOGRAPHY PACK, Model Number DYNJ66476; jjj) SURGERY ANGIO PACK-LF, Model Number DYNJ66500; kkk) OR ANGIOGRAPHY PACK, Model Number DYNJ67245; lll) ANGIOGRAPHY DRAPE PACK, Model Number DYNJ68065A; mmm) ANGIO TRAY, Model Number DYNJ68758; nnn) ANGIO, Model Number DYNJ902507B; ooo) ANGIOGRAPHY PACK, Model Number DYNJC2371G; ppp) ANGIOGRAPHY DRAPE PACK, Model Number DYNJT2344B; qqq) PEDI ANGIO PACK-LF, Model Number PHS41743D
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
all lots within expiry, distributed from Medline between December 2017 - May 2023: a) Model Number 00-401318M, UDI/DI (EA)10193489348750, UDI/DI (CS)40193489348751 b) Model Number CDS984944C, UDI/DI (EA)10889942172955, UDI/DI (CS)40889942172956 c) Model Number DYNJ0149444M, UDI/DI (EA)10193489432824, UDI/DI (CS)40193489432825 d) Model Number DYNJ0160556D, UDI/DI (EA)10193489207552, UDI/DI (CS)40193489207553 e) Model Number DYNJ0201109AG, UDI/DI (EA)10193489497267, UDI/DI (CS)40193489497268 f) Model Number DYNJ0373279M, UDI/DI (EA)10193489519020, UDI/DI (CS)40193489519021 g) Model Number DYNJ0429121I, UDI/DI (EA)10193489897494, UDI/DI (CS)40193489897495 h) Model Number DYNJ0806254D, UDI/DI (EA)10889942471164, UDI/DI (CS)40889942471165 i) Model Number DYNJ0854485T, UDI/DI (EA)10193489864199, UDI/DI (CS)40193489864190 j) Model Number DYNJ17624B, UDI/DI (EA)10193489867299, UDI/DI (CS)40193489867290 k) Model Number DYNJ19939M, UDI/DI (EA)10193489393248, UDI/DI (CS)40193489393249 l) Model Number DYNJ24225K, UDI/DI (EA)10889942510702, UDI/DI (CS)40889942510703 m) Model Number DYNJ28067G, UDI/DI (EA)10889942744299, UDI/DI (CS)40889942744290 n) Model Number DYNJ33457C, UDI/DI (EA)10193489461800, UDI/DI (CS)40193489461801 o) Model Number DYNJ33638K, UDI/DI (EA)10193489343281, UDI/DI (CS)40193489343282 p) Model Number DYNJ34361, UDI/DI (EA)10884389348475, UDI/DI (CS)40884389348476 q) Model Number DYNJ36049, UDI/DI (EA)10884389498675, UDI/DI (CS)40884389498676 r) Model Number DYNJ37665B, UDI/DI (EA)10889942969739, UDI/DI (CS)40889942969730 s) Model Number DYNJ40632C, UDI/DI (EA)10193489461176, UDI/DI (CS)40193489461177 t) Model Number DYNJ41553D, UDI/DI (EA)10889942811335, UDI/DI (CS)40889942811336 u) Model Number DYNJ41649F, UDI/DI (EA)10193489641271, UDI/DI (CS)40193489641272 v) Model Number DYNJ42499C, UDI/DI (EA)10195327030865, UDI/DI (CS)40195327030866 w) Model Number DYNJ42903B, UDI/DI (EA)10888277798014, UDI/DI (CS)40888277798015 x) Model Number DYNJ43661B, UDI/DI (EA)10889942808700, UDI/DI (CS)40889942808701 y) Model Number DYNJ44293D, UDI/DI (EA)10193489833416, UDI/DI (CS)40193489833417 z) Model Number DYNJ44320F, UDI/DI (EA)10193489435511, UDI/DI (CS)40193489435512 aa) Model Number DYNJ45007, UDI/DI (EA)10888277085411, UDI/DI (CS)40888277085412 bb) Model Number DYNJ46043F, UDI/DI (EA)10193489856729, UDI/DI (CS)40193489856720 cc) Model Number DYNJ46153B, UDI/DI (EA)10193489853643, UDI/DI (CS)40193489853644 dd) Model Number DYNJ47004C, UDI/DI (EA)10193489696172, UDI/DI (CS)40193489696173 ee) Model Number DYNJ47226A, UDI/DI (EA)10888277561816, UDI/DI (CS)40888277561817 ff) Model Number DYNJ47906, UDI/DI (EA)10888277528376, UDI/DI (CS)40888277528377 gg) Model Number DYNJ47951A, UDI/DI (EA)10193489609844, UDI/DI (CS)40193489609845 hh) Model Number DYNJ49884F, UDI/DI (EA)10193489469578, UDI/DI (CS)40193489469579 ii) Model Number DYNJ50624A, UDI/DI (EA)10889942474639, UDI/DI (CS)40889942474630 jj) Model Number DYNJ51239A, UDI/DI (EA)10889942493548, UDI/DI (CS)40889942493549 kk) Model Number DYNJ52121, UDI/DI (EA)10889942243624, UDI/DI (CS)40889942243625 ll) Model Number DYNJ52898C, UDI/DI (EA)10193489876314, UDI/DI (CS)40193489876315 mm) Model Number DYNJ55252B, UDI/DI (EA)10193489870152, UDI/DI (CS)40193489870153 nn) Model Number DYNJ57498, UDI/DI (EA)10889942745746, UDI/DI (CS)40889942745747 oo) Model Number DYNJ57645B, UDI/DI (EA)10193489828658, UDI/DI (CS)40193489828659 pp) Model Number DYNJ57760B, UDI/DI (EA)10193489342581, UDI/DI (CS)40193489342582 qq) Model Number DYNJ57792A, UDI/DI (EA)10193489530896, UDI/DI (CS)40193489530897 rr) Model Number DYNJ57792B, UDI/DI (EA)10193489845945, UDI/DI (CS)40193489845946 ss) Model Number DYNJ58126A, UDI/DI (EA)10193489484755, UDI/DI (CS)40193489484756 tt) Model Number DYNJ59465B, UDI/DI (EA)10193489470932, UDI/DI (CS)40193489470933 uu) Model Number DYNJ60319A, UDI/DI (EA)10193489276299, UDI/DI (CS)40193489276290 vv) Model Number DYNJ60507A, UDI/DI (EA)10193489830552, UDI/DI (CS)40193489830553 ww) Model Number DYNJ61015B, UDI/DI (EA)10193489368888, UDI/DI (CS)40193489368889 xx) Model Number DYNJ62299, UDI/DI (EA)10193489641516, UDI/DI (CS)40193489641517 yy) Model Number DYNJ62379, UDI/DI (EA)10193489651706, UDI/DI (CS)40193489651707 zz) Model Number DYNJ62544, UDI/DI (EA)10193489666526, UDI/DI (CS)40193489666527 aaa) Model Number DYNJ62742, UDI/DI (EA)10193489685923, UDI/DI (CS)40193489685924 bbb) Model Number DYNJ63976A, UDI/DI (EA)10193489377965, UDI/DI (CS)40193489377966 ccc) Model Number DYNJ64830, UDI/DI (EA)10193489223132, UDI/DI (CS)40193489223133 ddd) Model Number DYNJ65043A, UDI/DI (EA)10193489379327, UDI/DI (CS)40193489379328 eee) Model Number DYNJ65185, UDI/DI (EA)10193489253146, UDI/DI (CS)40193489253147 fff) Model Number DYNJ65190C, UDI/DI (EA)10193489433111, UDI/DI (CS)40193489433112 ggg) Model Number DYNJ65482A, UDI/DI (EA)10193489416824, UDI/DI (CS)40193489416825 hhh) Model Number DYNJ66398B, UDI/DI (EA)10193489891928, UDI/DI (CS)40193489891929 iii) Model Number DYNJ66476, UDI/DI (EA)10193489356779, UDI/DI (CS)40193489356770 jjj) Model Number DYNJ66500, UDI/DI (EA)10193489357912, UDI/DI (CS)40193489357913 kkk) Model Number DYNJ67245, UDI/DI (EA)10193489393057, UDI/DI (CS)40193489393058 lll) Model Number DYNJ68065A, UDI/DI (EA)10193489888720, UDI/DI (CS)40193489888721 mmm) Model Number DYNJ68758, UDI/DI (EA)10193489837568, UDI/DI (CS)40193489837569 nnn) Model Number DYNJ902507B, UDI/DI (EA)10889942781188, UDI/DI (CS)40889942781189 ooo) Model Number DYNJC2371G, UDI/DI (EA)10889942494675, UDI/DI (CS)40889942494676 ppp) Model Number DYNJT2344B, UDI/DI (EA)10889942107803, UDI/DI (CS)40889942107804 qqq) Model Number PHS41743D, UDI/DI (EA)10193489453270, UDI/DI (CS)40193489453271
Devices named in this recall
69 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ65185Medline Industries, Inc. · Angiography/angioplasty kit3 recalls1 report
- Medline Industries, Inc. DYNJC2371GMedline Industries, Inc. · Angiography/angioplasty kit3 recalls0 reports
- Medline Industries, Inc. DYNJ40632CMedline Industries, Inc. · Angiography/angioplasty kit2 recalls1 report
- Medline Industries, Inc. DYNJ57760BMedline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ33638KMedline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ61015BMedline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ62742Medline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ44293DMedline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ46153BMedline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ0854485TMedline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ50624AMedline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ57498Medline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
- Medline Industries, Inc. DYNJ0160556DMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ64830Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ60319AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. 00-401318MMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ66476Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ66500Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ63976AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ65043AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ67245Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ19939MMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ65482AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ0149444MMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ65190CMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ44320FMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. PHS41743DMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ33457CMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ49884FMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ59465BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ58126AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ0201109AGMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ0373279MMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ57792AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ47951AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ41649FMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ62299Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ62379Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ62544Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ47004CMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ57645BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ60507AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ68758Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ57792BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ46043FMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ17624BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ55252BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ52898CMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ68065AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ66398BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ0429121IMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ42499CMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ34361Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ36049Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ45007Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ47906Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ47226AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ42903BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJT2344BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. CDS984944CMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ52121Medline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ0806254DMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ51239AMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ24225KMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ28067GMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ902507BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ43661BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ41553DMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
- Medline Industries, Inc. DYNJ37665BMedline Industries, Inc. · Angiography/angioplasty kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92466).
- Z-2154-2023 · Open, ClassifiedMedline Probe Cover Kits, Model Numbers: DYNJE5900, DYNJE5910, DYNJE5920, DYNJE5930, DYNJE5940, DYNJEX5900. for use during diagnostic ultrasound procedures
- Z-2155-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: ANESTHESIA OH TOTE, Model Number DYNJ61476
- Z-2156-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BLOCK PACK, Model Number DYNJ81722; b) PAIN BLOCK TRAY, Model Number DYNJRA0827A; c) PAI…
- Z-2157-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: PEDIATRIC EP PACK, Model Number DYNJ45158A
- Z-2158-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NERVE BLOCK ACCESSORY PACK, Model Number DYNJRA1220; b) PERI NEURO BLOCK PACK-LF, Model N…
- Z-2160-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY DRAPE PACK, Model Number 00-400305R; b) ANGIO HOLLAND PACK-LF, Model Number D…
- Z-2161-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY PACK, Model Number DYNJ55252C; b) CATH LAB PACK, Model Number DYNJ55455A; c)…
- Z-2162-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATH PACK, Model Number DYNJ83105; b) TESIO, Model Number DYNJ83251; c) ANGIOGRAPHY PACK…
- Z-2163-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a. NORTON CHILDRENS CATH LAB-LF, Model Number 00-401185Q; b. PERCUTANEOUS VASCULAR CDS, Mod…
- Z-2164-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) MHC VASCULAR HYBRID PK-LF, Model Number DYNJ53433K; b) VEIN SPECIALISTS PACK, Model Numbe…
- Z-2165-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) EP PACER PACK, Model Number DYNJ67770A; b) EP PROCEDURE PACK, Model Number DYNJ67771; c)…
- Z-2166-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) SILVER CROSS OPEN HEART, Model Number DYNJ906102G; b) SILVER CROSS OPEN HEART, Model Numb…
- Z-2167-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ARTERIAL BUNDLE NO CATHETER, Model Number ART995; b) ART LINE KIT NO CATH, Model Number A…
- Z-2168-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) PORT PACK, Model Number DYNJ64675; b) PICC LINE TRAY, Model Number DYNJ64725B; c) VEIN O…
- Z-2169-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETAL SURGERY CDS-LF, Model Number CDS840214Q; b) MINOR PROCEDURE TRAY, Model Number DYND…
- Z-2170-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BREAST BIOPSY KIT, Model Number DYNDA1224B; b) U/S CORE BIOPSY KIT, Model Number DYNDH112…
- Z-2171-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: STERILE SET FOR IMPLANT-LF, Model Number OST011F
- Z-2172-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CHP COCHLEAR SUPPLEMENT CDS, Model Number CDS983069I; b) HEAD AND NECK PACK, Model Number…
- Z-2173-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETOSCOPY CDS-LF, Model Number CDS840215O; b) US BREAST BIOPSY TRAY, Model Number DYNDH…
- Z-2174-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: PACK ENDO AAA HYBRID CHRG, Model Number DYNJ60126B
- Z-2175-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATHETER INSERTION KIT CLABSI, Model Number CVI4880; b) CENTRAL CATHETER INSERTION TRA, …
- Z-2176-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CVC PACK, Model Number DYNJ46967F; b) ENDOVENOUS CORE PACK, Model Number DYNJ62557; c…
- Z-2177-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BREAST PACK, Model Number DYNDH1306; b) BRST BX TRAY, Model Number DYNDH1389B; c) PLAST…
- Z-2178-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: CATH LAB BASIN PACK, Model Number DYNJ69864A
- Z-2179-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NO-CATH RADIAL ARTERIAL PREMIU, Model Number ARTNC115; b) MAJOR VASCULAR, Model Number C…
- Z-2180-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) VEIN KIT, Model Number DYNJ55841B; b) FACELIFT PACK, Model Number DYNJ55954B; c) CLOT R…
- Z-2181-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BLOCK SUPPLIES NDL CNTR, Model Number DYNJRA1755; b) VNUS TEST KIT CFP-LF, Model Number …
- Z-2182-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) UNIVERSAL INSERTION TRAY, Model Number CVI4325; b) PPE BUNDLE KIT, Model Number DT13351A
- Z-2183-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) IV SECUREMENT KIT, Model Number DYNDV2407A; b) USG IV START KIT, Model Number DYNDV2435;…
- Z-2184-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) LACERATION TRAY, Model Number DYNDL1134A; b) ULTRASOUND GUIDED IV/MIDLINE K, Model Numbe…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page