Medline Probe Cover Kits containing sterile Ultrasound Gel… recallZ-2164-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 18, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
FDA root cause category: Device Design
What the company is doing
Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline. Questions: please contact the Recall Department at 866-359-1704
Recall details
- Recall number
- Z-2164-2023
- Status
- Open, Classified
- Initiated by the company
- May 18, 2023
- Posted by the FDA
- Jul 20, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- # included in 3,740,814 units total
- Product code
- OEZ
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
Product description
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) MHC VASCULAR HYBRID PK-LF, Model Number DYNJ53433K; b) VEIN SPECIALISTS PACK, Model Number DYNJ53616A; c) SUMMIT SKIN AND VEIN PACK, Model Number DYNJ54014B; d) VEIN PACK, Model Number DYNJ54153B; e) RFD-ENDOVASCULAR OR PACK, Model Number DYNJ54265A; f) PEDI EP PACK, Model Number DYNJ54750G; g) EL PASO VERICOSE VEIN PACK, Model Number DYNJ54803C; h) LEE VEIN PACK, Model Number DYNJ54906B; i) HEART FAILURE PK, Model Number DYNJ55528A; j) HEART FAILURE PK, Model Number DYNJ55528B; k) HEART FAILURE PK, Model Number DYNJ55528C; l) ROCKWAY VASCULAR PACK-LF, Model Number DYNJ55757C; m) OPEN HEART ANGIO PACK, Model Number DYNJ55836B; n) DR NACKMAN VEIN PACK, Model Number DYNJ56235; o) KT TAVR PACK, Model Number DYNJ56283D; p) KT TAVR PACK, Model Number DYNJ56283F; q) VEIN PACK, Model Number DYNJ56724B; r) VASCULAR PACK, Model Number DYNJ56791A; s) VASCULAR PACK, Model Number DYNJ56791B; t) ECMO PACK, Model Number DYNJ56904; u) VEIN KIT, Model Number DYNJ57047; v) EH TAVR PACK, Model Number DYNJ57449F; w) EH TAVR PACK, Model Number DYNJ57449G; x) VASCULAR PACK, Model Number DYNJ57490A; y) VEIN PACK, Model Number DYNJ57512; z) DEVICE PACK, Model Number DYNJ57595A; aa) CAROTID PACK-LF, Model Number DYNJ57646B; bb) CAROTID PACK-LF, Model Number DYNJ57646C; cc) CATH LAB PACK, Model Number DYNJ57735F; dd) VEIN PACK, Model Number DYNJ57835; ee) VEIN PACK, Model Number DYNJ57917; ff) ENS. THROMBOLYSE/EMBOLISAT-LF, Model Number DYNJ58179C; gg) HEART VALVE PACK, Model Number DYNJ58365J; hh) VEIN PACK, Model Number DYNJ58609; ii) VASCULAR PACK, Model Number DYNJ58681A; jj) VASCULAR PACK, Model Number DYNJ58681B; kk) VASCULAR PACK, Model Number DYNJ58710; ll) DR HARRISON VEIN PACK II, Model Number DYNJ59028A; mm) VEIN PACK, Model Number DYNJ59151F; nn) EP LAB, Model Number DYNJ59392D; oo) EP LAB, Model Number DYNJ59392F; pp) VEIN PACK, Model Number DYNJ59662D; qq) VEIN PACK, Model Number DYNJ59681A; rr) VASCULAR PACK CHRG, Model Number DYNJ59778B; ss) VASCULAR ACCESS KIT, Model Number DYNJ59796C; tt) CATH PACK, Model Number DYNJ59830A; uu) VENOUS ACCESS ACCESSORY PACK, Model Number DYNJ61010A; vv) DR SIRAGUSA VEIN PACK, Model Number DYNJ61247; ww) MAIN OR CATH LAB, Model Number DYNJ61373; xx) VASCULAR PACK, Model Number DYNJ61528; yy) VASCULAR PACK, Model Number DYNJ61528A; zz) VEIN PACK, Model Number DYNJ61540C; aaa) NEURO VASCULAR PACK, Model Number DYNJ61584A; bbb) ABLATION PACK, Model Number DYNJ61957A; ccc) HEART GROIN PACK, Model Number DYNJ61981C; ddd) PERMA CATH PACK, Model Number DYNJ62007B; eee) PURPLE PACK, Model Number DYNJ62197A; fff) PURPLE PACK, Model Number DYNJ62197B; ggg) CLOVIS CATH LAB PACK, Model Number DYNJ62379A; hhh) VASCULAR PACK, Model Number DYNJ62399A; iii) NORTH TEXAS VASCULAR, Model Number DYNJ62735; jjj) VEIN TRAY, Model Number DYNJ62829A; kkk) HEART VEIN NYC LASER ABLATION, Model Number DYNJ62850A; lll) HEART VEIN NYC -ARTERIAL PACK, Model Number DYNJ62851B; mmm) TAVR PACK, Model Number DYNJ62915B; nnn) PACER SET UP PACK, Model Number DYNJ63069A; ooo) OPEN HEART TAVR PACK, Model Number DYNJ63099D; ppp) OPEN HEART TAVR PACK, Model Number DYNJ63099F; qqq) ELECTROPHYSIOLOGY PACK, Model Number DYNJ63209; rrr) TAVR PACK, Model Number DYNJ63340A; sss) VEIN PACK, Model Number DYNJ63823; ttt) NIR PACK, Model Number DYNJ63892A; uuu) VEIN ABLATION TRAY W/CHLOR, Model Number DYNJ64396; vvv) TORRES-VEIN PACK, Model Number DYNJ64449; www) VEIN PACK, Model Number DYNJ64471A; xxx) CVOR ENDOVASCULAR PACK, Model Number DYNJ64506A; yyy) HEART FAILURE PACK, Model Number DYNJ64877B; zzz) CT ABLATION PACK, Model Number DYNJ64928; aaaa) VEIN PROCEDURE, Model Number DYNJ65342A; bbbb) SMALL TRAY VASCULAR PACK, Model Number DYNJ65418; cccc) VEIN OHH, Model Number DYNJ65560A; dddd) PORT PICC TUNNEL PACK, Model Number DYNJ65775A; eeee) VARITHENA PACK, Model Number DYNJ65
Code information (models, lots, serial numbers)
Show the full code information (9k characters)
all lots within expiry, distributed from Medline between December 2017 - May 2023: a) Model Numbers DYNJ53433K, UDI/DI (EA)10195327339487, UDI/DI (CS)40195327339488; b) Model Numbers DYNJ53616A, UDI/DI (EA)10889942688562, UDI/DI (CS)40889942688563; c) Model Numbers DYNJ54014B, UDI/DI (EA)10889942760329, UDI/DI (CS)40889942760320; d) Model Numbers DYNJ54153B, UDI/DI (EA)10889942714421, UDI/DI (CS)40889942714422; e) Model Numbers DYNJ54265A, UDI/DI (EA)10195327152338, UDI/DI (CS)40195327152339; f) Model Numbers DYNJ54750G, UDI/DI (EA)10195327021351, UDI/DI (CS)40195327021352; g) Model Numbers DYNJ54803C, UDI/DI (EA)10193489202618, UDI/DI (CS)40193489202619; h) Model Numbers DYNJ54906B, UDI/DI (EA)10193489289633, UDI/DI (CS)40193489289634; i) Model Numbers DYNJ55528A, UDI/DI (EA)10889942976638, UDI/DI (CS)40889942976639; j) Model Numbers DYNJ55528B, UDI/DI (EA)10195327046491, UDI/DI (CS)40195327046492; k) Model Numbers DYNJ55528C, UDI/DI (EA)10195327248208, UDI/DI (CS)40195327248209; l) Model Numbers DYNJ55757C, UDI/DI (EA)10195327116538, UDI/DI (CS)40195327116539; m) Model Numbers DYNJ55836B, UDI/DI (EA)10193489951172, UDI/DI (CS)40193489951173; n) Model Numbers DYNJ56235, UDI/DI (EA)10889942621552, UDI/DI (CS)40889942621553; o) Model Numbers DYNJ56283D, UDI/DI (EA)10195327093877, UDI/DI (CS)40195327093878; p) Model Numbers DYNJ56283F, UDI/DI (EA)10195327211851, UDI/DI (CS)40195327211852; q) Model Numbers DYNJ56724B, UDI/DI (EA)10193489466317, UDI/DI (CS)40193489466318; r) Model Numbers DYNJ56791A, UDI/DI (EA)10193489594119, UDI/DI (CS)40193489594110; s) Model Numbers DYNJ56791B, UDI/DI (EA)10195327045678, UDI/DI (CS)40195327045679; t) Model Numbers DYNJ56904, UDI/DI (EA)10889942693450, UDI/DI (CS)40889942693451; u) Model Numbers DYNJ57047, UDI/DI (EA)10889942710485, UDI/DI (CS)40889942710486; v) Model Numbers DYNJ57449F, UDI/DI (EA)10193489939200, UDI/DI (CS)40193489939201; w) Model Numbers DYNJ57449G, UDI/DI (EA)10195327064884, UDI/DI (CS)40195327064885; x) Model Numbers DYNJ57490A, UDI/DI (EA)10193489641998, UDI/DI (CS)40193489641999; y) Model Numbers DYNJ57512, UDI/DI (EA)10889942748273, UDI/DI (CS)40889942748274; z) Model Numbers DYNJ57595A, UDI/DI (EA)10193489622157, UDI/DI (CS)40193489622158; aa) Model Numbers DYNJ57646B, UDI/DI (EA)10193489828504, UDI/DI (CS)40193489828505; bb) Model Numbers DYNJ57646C, UDI/DI (EA)10195327287139, UDI/DI (CS)40195327287130; cc) Model Numbers DYNJ57735F, UDI/DI (EA)10193489629705, UDI/DI (CS)40193489629706; dd) Model Numbers DYNJ57835, UDI/DI (EA)10889942782338, UDI/DI (CS)40889942782339; ee) Model Numbers DYNJ57917, UDI/DI (EA)10889942794294, UDI/DI (CS)40889942794295; ff) Model Numbers DYNJ58179C, UDI/DI (EA)10195327209926, UDI/DI (CS)40195327209927; gg) Model Numbers DYNJ58365J, UDI/DI (EA)10195327019099, UDI/DI (CS)40195327019090; hh) Model Numbers DYNJ58609, UDI/DI (EA)10889942871261, UDI/DI (CS)40889942871262; ii) Model Numbers DYNJ58681A, UDI/DI (EA)10888277798489, UDI/DI (CS)40888277798480; jj) Model Numbers DYNJ58681B, UDI/DI (EA)10195327293376, UDI/DI (CS)40195327293377; kk) Model Numbers DYNJ58710, UDI/DI (EA)10889942880850, UDI/DI (CS)40889942880851; ll) Model Numbers DYNJ59028A, UDI/DI (EA)10193489736595, UDI/DI (CS)40193489736596; mm) Model Numbers DYNJ59151F, UDI/DI (EA)10193489487565, UDI/DI (CS)40193489487566; nn) Model Numbers DYNJ59392D, UDI/DI (EA)10195327053574, UDI/DI (CS)40195327053575; oo) Model Numbers DYNJ59392F, UDI/DI (EA)10195327135980, UDI/DI (CS)40195327135981; pp) Model Numbers DYNJ59662D, UDI/DI (EA)10195327030872, UDI/DI (CS)40195327030873; qq) Model Numbers DYNJ59681A, UDI/DI (EA)10193489469875, UDI/DI (CS)40193489469876; rr) Model Numbers DYNJ59778B, UDI/DI (EA)10193489695922, UDI/DI (CS)40193489695923; ss) Model Numbers DYNJ59796C, UDI/DI (EA)10193489904215, UDI/DI (CS)40193489904216; tt) Model Numbers DYNJ59830A, UDI/DI (EA)10193489570236, UDI/DI (CS)40193489570237; uu) Model Numbers DYNJ61010A, UDI/DI (EA)10193489307146, UDI/DI (CS)40193489307147; vv) Model Numbers DYNJ61247, UDI/DI (EA)10193489551709, UDI/DI (CS)40193489551700; ww) Model Numbers DYNJ61373, UDI/DI (EA)10193489576696, UDI/DI (CS)40193489576697; xx) Model Numbers DYNJ61528, UDI/DI (EA)10193489578980, UDI/DI (CS)40193489578981; yy) Model Numbers DYNJ61528A, UDI/DI (EA)10195327336714, UDI/DI (CS)40195327336715; zz) Model Numbers DYNJ61540C, UDI/DI (EA)10193489495126, UDI/DI (CS)40193489495127; aaa) Model Numbers DYNJ61584A, UDI/DI (EA)10193489812817, UDI/DI (CS)40193489812818; bbb) Model Numbers DYNJ61957A, UDI/DI (EA)10193489699074, UDI/DI (CS)40193489699075; ccc) Model Numbers DYNJ61981C, UDI/DI (EA)10195327106027, UDI/DI (CS)40195327106028; ddd) Model Numbers DYNJ62007B, UDI/DI (EA)10195327161965, UDI/DI (CS)40195327161966; eee) Model Numbers DYNJ62197A, UDI/DI (EA)10193489994094, UDI/DI (CS)40193489994095; fff) Model Numbers DYNJ62197B, UDI/DI (EA)10195327167240, UDI/DI (CS)40195327167241; ggg) Model Numbers DYNJ62379A, UDI/DI (EA)10195327196639, UDI/DI (CS)40195327196630; hhh) Model Numbers DYNJ62399A, UDI/DI (EA)10193489250268, UDI/DI (CS)40193489250269; iii) Model Numbers DYNJ62735, UDI/DI (EA)10193489685282, UDI/DI (CS)40193489685283; jjj) Model Numbers DYNJ62829A, UDI/DI (EA)10193489306620, UDI/DI (CS)40193489306621; kkk) Model Numbers DYNJ62850A, UDI/DI (EA)10193489784237, UDI/DI (CS)40193489784238; lll) Model Numbers DYNJ62851B, UDI/DI (EA)10193489271218, UDI/DI (CS)40193489271219; mmm) Model Numbers DYNJ62915B, UDI/DI (EA)10193489365351, UDI/DI (CS)40193489365352; nnn) Model Numbers DYNJ63069A, UDI/DI (EA)10193489423617, UDI/DI (CS)40193489423618; ooo) Model Numbers DYNJ63099D, UDI/DI (EA)10193489932690, UDI/DI (CS)40193489932691; ppp) Model Numbers DYNJ63099F, UDI/DI (EA)10195327150037, UDI/DI (CS)40195327150038; qqq) Model Numbers DYNJ63209, UDI/DI (EA)10193489727883, UDI/DI (CS)40193489727884; rrr) Model Numbers DYNJ63340A, UDI/DI (EA)10195327273583, UDI/DI (CS)40195327273584; sss) Model Numbers DYNJ63823, UDI/DI (EA)10193489763423, UDI/DI (CS)40193489763424; ttt) Model Numbers DYNJ63892A, UDI/DI (EA)10195327176518, UDI/DI (CS)40195327176519; uuu) Model Numbers DYNJ64396, UDI/DI (EA)10193489810509, UDI/DI (CS)40193489810500; vvv) Model Numbers DYNJ64449, UDI/DI (EA)10193489201574, UDI/DI (CS)40193489201575; www) Model Numbers DYNJ64471A, UDI/DI (EA)10193489414158, UDI/DI (CS)40193489414159; xxx) Model Numbers DYNJ64506A, UDI/DI (EA)10193489971064, UDI/DI (CS)40193489971065; yyy) Model Numbers DYNJ64877B, UDI/DI (EA)10193489415896, UDI/DI (CS)40193489415897; zzz) Model Numbers DYNJ64928, UDI/DI (EA)10193489231502, UDI/DI (CS)40193489231503; aaaa) Model Numbers DYNJ65342A, UDI/DI (EA)10193489369663, UDI/DI (CS)40193489369664; bbbb) Model Numbers DYNJ65418, UDI/DI (EA)10193489273717, UDI/DI (CS)40193489273718; cccc) Model Numbers DYNJ65560A, UDI/DI (EA)10193489397185, UDI/DI (CS)40193489397186; dddd) Model Numbers DYNJ65775A, UDI/DI (EA)10195327140229, UDI/DI (CS)40195327140220; eeee) Model Numbers DYNJ65842, UDI/DI (EA)10193489296310, UDI/DI (CS)40193489296311; ffff) Model Numbers DYNJ66076, UDI/DI (EA)10193489319231, UDI/DI (CS)40193489319232; gggg) Model Numbers DYNJ66152A, UDI/DI (EA)10193489820218, UDI/DI (CS)40193489820219; hhhh) Model Numbers DYNJ66152B, UDI/DI (EA)10195327090425, UDI/DI (CS)40195327090426; iiii) Model Numbers DYNJ66271C, UDI/DI (EA)10195327098506, UDI/DI (CS)40195327098507; jjjj) Model Numbers DYNJ66356C, UDI/DI (EA)10193489991055, UDI/DI (CS)40193489991056; kkkk) Model Numbers DYNJ66356D, UDI/DI (EA)10195327060299, UDI/DI (CS)40195327060290; llll) Model Numbers DYNJ66356G, UDI/DI (EA)10195327357306, UDI/DI (CS)40195327357307; mmmm) Model Numbers DYNJ66427, UDI/DI (EA)10193489354683, UDI/DI (CS)40193489354684; nnnn) Model Numbers DYNJ66445, UDI/DI (EA)10193489356083, UDI/DI (CS)40193489356084; oooo) Model Numbers DYNJ66743A, UDI/DI (EA)10193489435559, UDI/DI (CS)40193489435550; pppp) Model Numbers DYNJ66769, UDI/DI (EA)10193489380521, UDI/DI (CS)40193489380522; qqqq) Model Numbers DYNJ66916A, UDI/DI (EA)10193489993226, UDI/DI (CS)40193489993227; rrrr) Model Numbers DYNJ66926B, UDI/DI (EA)10195327340261, UDI/DI (CS)40195327340262; ssss) Model Numbers DYNJ67487B, UDI/DI (EA)10195327166144, UDI/DI (CS)40195327166145; tttt) Model Numbers DYNJ67487C, UDI/DI (EA)10195327260743, UDI/DI (CS)40195327260744; uuuu) Model Numbers DYNJ67517A, UDI/DI (EA)10193489462197, UDI/DI (CS)40193489462198; vvvv) Model Numbers DYNJ67725, UDI/DI (EA)10193489425840, UDI/DI (CS)40193489425841; wwww) Model Numbers DYNJ67750A, UDI/DI (EA)10193489832242, UDI/DI (CS)40193489832243; xxxx) Model Numbers DYNJ67750B, UDI/DI (EA)10193489912043, UDI/DI (CS)40193489912044; yyyy) Model Numbers DYNJ67750C, UDI/DI (EA)10195327341367, UDI/DI (CS)40195327341368; zzzz) Model Numbers DYNJ67770, UDI/DI (EA)10193489430714, UDI/DI (CS)40193489430715;
Devices named in this recall
106 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- Medline Industries, Inc. DYNJ56235Medline Industries, Inc. · Cardiovascular procedure kit3 recalls0 reports
- Medline Industries, Inc. DYNJ64877BMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ59830AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ57646BMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ64506AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ54750GMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ57449GMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ63892AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ67487CMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ53433KMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ57512Medline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ59151FMedline Industries, Inc. · Cardiovascular procedure kit1 recall1 report
- Medline Industries, Inc. DYNJ64449Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ54803CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ64928Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62399AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62851BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ65418Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ54906BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ65842Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62829AMedline Industries, Inc. · Central venous catheter tray1 recall0 reports
- Medline Industries, Inc. DYNJ61010AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66076Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66427Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66445Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62915BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ65342AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66769Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ65560AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ64471AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ63069AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ67725Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ67770Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66743AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ67517AMedline Industries, Inc. · Central venous catheter tray1 recall0 reports
- Medline Industries, Inc. DYNJ56724BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ59681AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61540CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61247Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61373Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61528Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ56791AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ57595AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ57735FMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ57490AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62735Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ59778BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61957AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ63209Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ59028AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ63823Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62850AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ64396Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61584AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66152AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ67750AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ59796CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ67750BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ63099DMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ57449FMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ55836BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66356CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66885AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ67820AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ66916AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62197AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ58365JMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ59662DMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ56791BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ55528BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ59392DMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66356DMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66152BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ56283DMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66271CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61981CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ55757CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ59392FMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ65775AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ63099FMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ54265AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62007BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ67487BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62197BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62379AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ58179CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ56283FMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ55528CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ63340AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ57646CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ58681BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ61528AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66926BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ67750CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ66356GMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ58681AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ53616AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ56904Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ57047Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ54153BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92466).
- Z-2154-2023 · Open, ClassifiedMedline Probe Cover Kits, Model Numbers: DYNJE5900, DYNJE5910, DYNJE5920, DYNJE5930, DYNJE5940, DYNJEX5900. for use during diagnostic ultrasound procedures
- Z-2155-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: ANESTHESIA OH TOTE, Model Number DYNJ61476
- Z-2156-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BLOCK PACK, Model Number DYNJ81722; b) PAIN BLOCK TRAY, Model Number DYNJRA0827A; c) PAI…
- Z-2157-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: PEDIATRIC EP PACK, Model Number DYNJ45158A
- Z-2158-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NERVE BLOCK ACCESSORY PACK, Model Number DYNJRA1220; b) PERI NEURO BLOCK PACK-LF, Model N…
- Z-2159-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY DRP PK PEDI CATH, Model Number 00-401318M; b) ANGIOGRAPHY OR, Model Number CD…
- Z-2160-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY DRAPE PACK, Model Number 00-400305R; b) ANGIO HOLLAND PACK-LF, Model Number D…
- Z-2161-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY PACK, Model Number DYNJ55252C; b) CATH LAB PACK, Model Number DYNJ55455A; c)…
- Z-2162-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATH PACK, Model Number DYNJ83105; b) TESIO, Model Number DYNJ83251; c) ANGIOGRAPHY PACK…
- Z-2163-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a. NORTON CHILDRENS CATH LAB-LF, Model Number 00-401185Q; b. PERCUTANEOUS VASCULAR CDS, Mod…
- Z-2165-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) EP PACER PACK, Model Number DYNJ67770A; b) EP PROCEDURE PACK, Model Number DYNJ67771; c)…
- Z-2166-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) SILVER CROSS OPEN HEART, Model Number DYNJ906102G; b) SILVER CROSS OPEN HEART, Model Numb…
- Z-2167-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ARTERIAL BUNDLE NO CATHETER, Model Number ART995; b) ART LINE KIT NO CATH, Model Number A…
- Z-2168-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) PORT PACK, Model Number DYNJ64675; b) PICC LINE TRAY, Model Number DYNJ64725B; c) VEIN O…
- Z-2169-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETAL SURGERY CDS-LF, Model Number CDS840214Q; b) MINOR PROCEDURE TRAY, Model Number DYND…
- Z-2170-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BREAST BIOPSY KIT, Model Number DYNDA1224B; b) U/S CORE BIOPSY KIT, Model Number DYNDH112…
- Z-2171-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: STERILE SET FOR IMPLANT-LF, Model Number OST011F
- Z-2172-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CHP COCHLEAR SUPPLEMENT CDS, Model Number CDS983069I; b) HEAD AND NECK PACK, Model Number…
- Z-2173-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETOSCOPY CDS-LF, Model Number CDS840215O; b) US BREAST BIOPSY TRAY, Model Number DYNDH…
- Z-2174-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: PACK ENDO AAA HYBRID CHRG, Model Number DYNJ60126B
- Z-2175-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATHETER INSERTION KIT CLABSI, Model Number CVI4880; b) CENTRAL CATHETER INSERTION TRA, …
- Z-2176-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CVC PACK, Model Number DYNJ46967F; b) ENDOVENOUS CORE PACK, Model Number DYNJ62557; c…
- Z-2177-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BREAST PACK, Model Number DYNDH1306; b) BRST BX TRAY, Model Number DYNDH1389B; c) PLAST…
- Z-2178-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: CATH LAB BASIN PACK, Model Number DYNJ69864A
- Z-2179-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NO-CATH RADIAL ARTERIAL PREMIU, Model Number ARTNC115; b) MAJOR VASCULAR, Model Number C…
- Z-2180-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) VEIN KIT, Model Number DYNJ55841B; b) FACELIFT PACK, Model Number DYNJ55954B; c) CLOT R…
- Z-2181-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BLOCK SUPPLIES NDL CNTR, Model Number DYNJRA1755; b) VNUS TEST KIT CFP-LF, Model Number …
- Z-2182-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) UNIVERSAL INSERTION TRAY, Model Number CVI4325; b) PPE BUNDLE KIT, Model Number DT13351A
- Z-2183-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) IV SECUREMENT KIT, Model Number DYNDV2407A; b) USG IV START KIT, Model Number DYNDV2435;…
- Z-2184-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) LACERATION TRAY, Model Number DYNDL1134A; b) ULTRASOUND GUIDED IV/MIDLINE K, Model Numbe…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page