MEDLINE recallZ-2180-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 18, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
FDA root cause category: Device Design
What the company is doing
Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline. Questions: please contact the Recall Department at 866-359-1704
Recall details
- Recall number
- Z-2180-2023
- Status
- Open, Classified
- Initiated by the company
- May 18, 2023
- Posted by the FDA
- Jul 20, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- # included in 3,740,814 units total
- Product code
- LRO
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
Product description
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) VEIN KIT, Model Number DYNJ55841B; b) FACELIFT PACK, Model Number DYNJ55954B; c) CLOT RETRIEVAL PACK, Model Number DYNJ56140B; d) VENOUS PACK, Model Number DYNJ56260B; e) WMC ULTRASOUND PACK, Model Number DYNJ56392; f) WMC ULTRASOUND PACK, Model Number DYNJ56392A; g) WMC ULTRASOUND PACK, Model Number DYNJ56392B; h) ULTRASOUND KIT-LF, Model Number DYNJ56507B; i) EVLT PACK, Model Number DYNJ56548B; j) RFA PACK, Model Number DYNJ57149; k) SPECIALS TRACE PACK, Model Number DYNJ57303B; l) CLOSURE PACK, Model Number DYNJ57514; m) EP PACK, Model Number DYNJ57594A; n) VEIN TRAY, Model Number DYNJ57736A; o) ENSEMBLE PICC LINE-LF, Model Number DYNJ57793; p) PICC LINE PACK-LF, Model Number DYNJ58160; q) ENDOVENOUS PACK, Model Number DYNJ58228; r) BIB AV FISTULA PACK, Model Number DYNJ58304; s) TRIPLE P KIT, Model Number DYNJ58424; t) LINQ TRAY, Model Number DYNJ58643A; u) TUBE CHANGE/SINOGRAM PACK, Model Number DYNJ59062; v) PICC PROCEDURE PACK, Model Number DYNJ59067; w) ENDOVASCULAR PACK, Model Number DYNJ59085; x) VEIN CLOSURE PACK, Model Number DYNJ59094; y) CLOSURE PROCEDURE, Model Number DYNJ59206; z) VNUS PACK, Model Number DYNJ59246; aa) DESERT CARDIOVASCULAR PACK, Model Number DYNJ59377; bb) EP LAB, Model Number DYNJ59392C; cc) VEIN PACK-LF, Model Number DYNJ59631; dd) PROCEDURE PACK, Model Number DYNJ59795A; ee) PACK ENDO AAA HYBRID CHRG, Model Number DYNJ60126A; ff) KF PACK, Model Number DYNJ60127A; gg) VENASEAL PACK, Model Number DYNJ60128A; hh) I.R. PACK, Model Number DYNJ60343A; ii) CLOSURE PACK, Model Number DYNJ60804; jj) CLOVIS MINOR IR PROCEDURE-LF, Model Number DYNJ61067; kk) CRMC MINOR IR PROCEDURE PK-LF, Model Number DYNJ61099A; ll) ECMO PACK, Model Number DYNJ61243; mm) VEIN PACK, Model Number DYNJ61264; nn) UNIVERSAL PACK, Model Number DYNJ61433; oo) OCEANA VEIN PROCEDURE PACK, Model Number DYNJ61517; pp) ENSEMBLE CHIMIO EMBOLISATION, Model Number DYNJ62295; qq) ENS. CHIMIO EMBOLISATION-LF, Model Number DYNJ62295A; rr) ENS. CHIMIO EMBOLISATION-LF, Model Number DYNJ62295B; ss) SCLEROTHERAPY KIT, Model Number DYNJ62434A; tt) ABLATION PACK, Model Number DYNJ62540; uu) PROCEDURE PACK, Model Number DYNJ62588; vv) VEIN ABLATION, Model Number DYNJ63268B; ww) TF TAVR PACK, Model Number DYNJ63340; xx) AFCH DRUG INFUSION PUMP PA, Model Number DYNJ63354C; yy) ALL PURPOSE TRAY FOAKS, Model Number DYNJ63576A; zz) NIR PACK, Model Number DYNJ63892; aaa) ABLATION PACK, Model Number DYNJ64068; bbb) LASER KIT, Model Number DYNJ64075A; ccc) EVLT, Model Number DYNJ64336; ddd) PROCEDURE PACK, Model Number DYNJ64372A; eee) PREMIER PROCEDURE PACK, Model Number DYNJ64393; fff) MINOR SET UP PACK, Model Number DYNJ64439; ggg) VIP PACK, Model Number DYNJ64666A; hhh) CT ABLATION PACK, Model Number DYNJ64928A; iii) ULTRASOUND KIT, Model Number DYNJ64930; jjj) VEIN CLOSURE PACK, Model Number DYNJ65189C; kkk) PACK,LTD BODY SSP, Model Number DYNJ65488B; lll) EVLT PACK, Model Number DYNJ65995; mmm) EVLT PACK, Model Number DYNJ65995A; nnn) EVLT PACK, Model Number DYNJ66005; ooo) JVL-PACK TUNNELED, Model Number DYNJ66065; ppp) EVLA PACK, Model Number DYNJ66258; qqq) EVLA PACK, Model Number DYNJ66258A; rrr) ANGIOGRAM SET UP PACK, Model Number DYNJ66260; sss) IR PACKER, Model Number DYNJ66620A; ttt) CLOSURE PROCEDURE PACK, Model Number DYNJ66741; uuu) PHYSIATRY PACK, Model Number DYNJ67414; vvv) FEM POP PACK, Model Number DYNJ67487A; www) SPECIALS PACK, Model Number DYNJ67504A; xxx) SPECIALS PACK, Model Number DYNJ67504B; yyy) RFA PACK, Model Number DYNJ67899; zzz) RR-BRACHY PACK, Model Number DYNJ68027; aaaa) COPS PROCEDURE PACK-LF, Model Number DYNJ68923A; bbbb) PROCEDURE PACK, Model Number DYNJ68968; cccc) PORT PACK, Model Number DYNJ68990; dddd) ACCESS PACK, Model Number DYNJ69017; eeee) KF PACK 2.0, Model Number DYNJ69619; ffff) NLK BASIC PACK, Model Number DYNJ69716A; gg
Code information (models, lots, serial numbers)
Show the full code information (9k characters)
all lots within expiry, distributed from Medline between December 2017 - May 2023: a) Model Number DYNJ55841B, UDI/DI (EA) 10888277757684, UDI/DI (CS) 40888277757685; b) Model Number DYNJ55954B, UDI/DI (EA) 10193489214239, UDI/DI (CS) 40193489214230; c) Model Number DYNJ56140B, UDI/DI (EA) 10193489547672, UDI/DI (CS) 40193489547673; d) Model Number DYNJ56260B, UDI/DI (EA) 10193489569827, UDI/DI (CS) 40193489569828; e) Model Number DYNJ56392, UDI/DI (EA) 10889942640676, UDI/DI (CS) 40889942640677; f) Model Number DYNJ56392A, UDI/DI (EA) 10195327261405, UDI/DI (CS) 40195327261406; g) Model Number DYNJ56392B, UDI/DI (EA) 10195327323677, UDI/DI (CS) 40195327323678; h) Model Number DYNJ56507B, UDI/DI (EA) 10193489443479, UDI/DI (CS) 40193489443470; i) Model Number DYNJ56548B, UDI/DI (EA) 10193489813180, UDI/DI (CS) 40193489813181; j) Model Number DYNJ57149, UDI/DI (EA) 10889942721603, UDI/DI (CS) 40889942721604; k) Model Number DYNJ57303B, UDI/DI (EA) 10193489360646, UDI/DI (CS) 40193489360647; l) Model Number DYNJ57514, UDI/DI (EA) 10889942747849, UDI/DI (CS) 40889942747840; m) Model Number DYNJ57594A, UDI/DI (EA) 10889942963218, UDI/DI (CS) 40889942963219; n) Model Number DYNJ57736A, UDI/DI (EA) 10888277779709, UDI/DI (CS) 40888277779700; o) Model Number DYNJ57793, UDI/DI (EA) 10889942819164, UDI/DI (CS) 40889942819165; p) Model Number DYNJ58160, UDI/DI (EA) 10889942828647, UDI/DI (CS) 40889942828648; q) Model Number DYNJ58228, UDI/DI (EA) 10889942835492, UDI/DI (CS) 40889942835493; r) Model Number DYNJ58304, UDI/DI (EA) 10889942843732, UDI/DI (CS) 40889942843733; s) Model Number DYNJ58424, UDI/DI (EA) 10889942851836, UDI/DI (CS) 40889942851837; t) Model Number DYNJ58643A, UDI/DI (EA) 10193489809213, UDI/DI (CS) 40193489809214; u) Model Number DYNJ59062, UDI/DI (EA) 10889942941650, UDI/DI (CS) 40889942941651; v) Model Number DYNJ59067, UDI/DI (EA) 10889942930043, UDI/DI (CS) 40889942930044; w) Model Number DYNJ59085, UDI/DI (EA) 10889942932016, UDI/DI (CS) 40889942932017; x) Model Number DYNJ59094, UDI/DI (EA) 10889942933501, UDI/DI (CS) 40889942933502; y) Model Number DYNJ59206, UDI/DI (EA) 10888277765573, UDI/DI (CS) 40888277765574; z) Model Number DYNJ59246, UDI/DI (EA) 10888277767294, UDI/DI (CS) 40888277767295; aa) Model Number DYNJ59377, UDI/DI (EA) 10888277736290, UDI/DI (CS) 40888277736291; bb) Model Number DYNJ59392C, UDI/DI (EA) 10193489282245, UDI/DI (CS) 40193489282246; cc) Model Number DYNJ59631, UDI/DI (EA) 10888277764026, UDI/DI (CS) 40888277764027; dd) Model Number DYNJ59795A, UDI/DI (EA) 10193489566987, UDI/DI (CS) 40193489566988; ee) Model Number DYNJ60126A, UDI/DI (EA) 10193489247084, UDI/DI (CS) 40193489247085; ff) Model Number DYNJ60127A, UDI/DI (EA) 10193489729047, UDI/DI (CS) 40193489729048; gg) Model Number DYNJ60128A, UDI/DI (EA) 10193489544510, UDI/DI (CS) 40193489544511; hh) Model Number DYNJ60343A, UDI/DI (EA) 10193489573831, UDI/DI (CS) 40193489573832; ii) Model Number DYNJ60804, UDI/DI (EA) 10193489532784, UDI/DI (CS) 40193489532785; jj) Model Number DYNJ61067, UDI/DI (EA) 10193489542226, UDI/DI (CS) 40193489542227; kk) Model Number DYNJ61099A, UDI/DI (EA) 10193489463965, UDI/DI (CS) 40193489463966; ll) Model Number DYNJ61243, UDI/DI (EA) 10193489550917, UDI/DI (CS) 40193489550918; mm) Model Number DYNJ61264, UDI/DI (EA) 10193489554472, UDI/DI (CS) 40193489554473; nn) Model Number DYNJ61433, UDI/DI (EA) 10193489565171, UDI/DI (CS) 40193489565172; oo) Model Number DYNJ61517, UDI/DI (EA) 10193489577570, UDI/DI (CS) 40193489577571; pp) Model Number DYNJ62295, UDI/DI (EA) 10193489641165, UDI/DI (CS) 40193489641166; qq) Model Number DYNJ62295A, UDI/DI (EA) 10193489845952, UDI/DI (CS) 40193489845953; rr) Model Number DYNJ62295B, UDI/DI (EA) 10195327209933, UDI/DI (CS) 40195327209934; ss) Model Number DYNJ62434A, UDI/DI (EA) 10193489366273, UDI/DI (CS) 40193489366274; tt) Model Number DYNJ62540, UDI/DI (EA) 10193489666472, UDI/DI (CS) 40193489666473; uu) Model Number DYNJ62588, UDI/DI (EA) 10193489669831, UDI/DI (CS) 40193489669832; vv) Model Number DYNJ63268B, UDI/DI (EA) 10193489305111, UDI/DI (CS) 40193489305112; ww) Model Number DYNJ63340, UDI/DI (EA) 10193489741162, UDI/DI (CS) 40193489741163; xx) Model Number DYNJ63354C, UDI/DI (EA) 10195327273255, UDI/DI (CS) 40195327273256; yy) Model Number DYNJ63576A, UDI/DI (EA) 10193489429251, UDI/DI (CS) 40193489429252; zz) Model Number DYNJ63892, UDI/DI (EA) 10193489771435, UDI/DI (CS) 40193489771436; aaa) Model Number DYNJ64068, UDI/DI (EA) 10193489783131, UDI/DI (CS) 40193489783132; bbb) Model Number DYNJ64075A, UDI/DI (EA) 10193489231328, UDI/DI (CS) 40193489231329; ccc) Model Number DYNJ64336, UDI/DI (EA) 10193489801873, UDI/DI (CS) 40193489801874; ddd) Model Number DYNJ64372A, UDI/DI (EA) 10193489884685, UDI/DI (CS) 40193489884686; eee) Model Number DYNJ64393, UDI/DI (EA) 10193489809633, UDI/DI (CS) 40193489809634; fff) Model Number DYNJ64439, UDI/DI (EA) 10193489200485, UDI/DI (CS) 40193489200486; ggg) Model Number DYNJ64666A, UDI/DI (EA) 10193489317268, UDI/DI (CS) 40193489317269; hhh) Model Number DYNJ64928A, UDI/DI (EA) 10195327329815, UDI/DI (CS) 40195327329816; iii) Model Number DYNJ64930, UDI/DI (EA) 10193489231687, UDI/DI (CS) 40193489231688; jjj) Model Number DYNJ65189C, UDI/DI (EA) 10195327072001, UDI/DI (CS) 40195327072002; kkk) Model Number DYNJ65488B, UDI/DI (EA) 10193489443646, UDI/DI (CS) 40193489443647; lll) Model Number DYNJ65995, UDI/DI (EA) 10193489311549, UDI/DI (CS) 40193489311540; mmm) Model Number DYNJ65995A, UDI/DI (EA) 10195327068127, UDI/DI (CS) 40195327068128; nnn) Model Number DYNJ66005, UDI/DI (EA) 10193489312546, UDI/DI (CS) 40193489312547; ooo) Model Number DYNJ66065, UDI/DI (EA) 10193489319187, UDI/DI (CS) 40193489319188; ppp) Model Number DYNJ66258, UDI/DI (EA) 10193489325478, UDI/DI (CS) 40193489325479; qqq) Model Number DYNJ66258A, UDI/DI (EA) 10195327194239, UDI/DI (CS) 40195327194230; rrr) Model Number DYNJ66260, UDI/DI (EA) 10193489325461, UDI/DI (CS) 40193489325462; sss) Model Number DYNJ66620A, UDI/DI (EA) 10193489914245, UDI/DI (CS) 40193489914246; ttt) Model Number DYNJ66741, UDI/DI (EA) 10193489374940, UDI/DI (CS) 40193489374941; uuu) Model Number DYNJ67414, UDI/DI (EA) 10193489964646, UDI/DI (CS) 40193489964647; vvv) Model Number DYNJ67487A, UDI/DI (EA) 10193489888997, UDI/DI (CS) 40193489888998; www) Model Number DYNJ67504A, UDI/DI (EA) 10195327009618, UDI/DI (CS) 40195327009619; xxx) Model Number DYNJ67504B, UDI/DI (EA) 10195327331023, UDI/DI (CS) 40195327331024; yyy) Model Number DYNJ67899, UDI/DI (EA) 10193489454499, UDI/DI (CS) 40193489454490; zzz) Model Number DYNJ68027, UDI/DI (EA) 10193489469158, UDI/DI (CS) 40193489469159; aaaa) Model Number DYNJ68923A, UDI/DI (EA) 10195327070793, UDI/DI (CS) 40195327070794; bbbb) Model Number DYNJ68968, UDI/DI (EA) 10193489870398, UDI/DI (CS) 40193489870399; cccc) Model Number DYNJ68990, UDI/DI (EA) 10193489872613, UDI/DI (CS) 40193489872614; dddd) Model Number DYNJ69017, UDI/DI (EA) 10193489878158, UDI/DI (CS) 40193489878159; eeee) Model Number DYNJ69619, UDI/DI (EA) 10193489959444, UDI/DI (CS) 40193489959445; ffff) Model Number DYNJ69716A, UDI/DI (EA) 10195327250577, UDI/DI (CS) 40195327250578; gggg) Model Number DYNJ69820, UDI/DI (EA) 10193489981438, UDI/DI (CS) 40193489981439; hhhh) Model Number DYNJ80299, UDI/DI (EA) 10195327021931, UDI/DI (CS) 40195327021932; iiii) Model Number DYNJ81343, UDI/DI (EA) 10195327113063, UDI/DI (CS) 40195327113064; jjjj) Model Number DYNJ83152, UDI/DI (EA) 10195327237080, UDI/DI (CS) 40195327237081; kkkk) Model Number DYNJ83524, UDI/DI (EA) 10195327273491, UDI/DI (CS) 40195327273492; llll) Model Number DYNJ83931, UDI/DI (EA) 10195327315351, UDI/DI (CS) 40195327315352; mmmm) Model Number DYNJ83976, UDI/DI (EA) 10195327323196, UDI/DI (CS) 40195327323197; nnnn) Model Number DYNJ84172, UDI/DI (EA) 10195327338909, UDI/DI (CS) 40195327338900; oooo) Model Number DYNJ901348D, UDI/DI (EA) 10193489594683, UDI/DI (CS) 40193489594684; pppp) Model Number DYNJ901392D, UDI/DI (EA) 10193489539516, UDI/DI (CS) 40193489539517; qqqq) Model Number DYNJ902168D, UDI/DI (EA) 10193489866155, UDI/DI (CS) 40193489866156; rrrr) Model Number DYNJ902281, UDI/DI (EA) 10888277981416, UDI/DI (CS) 40888277981417; ssss) Model Number DYNJ905002A, UDI/DI (EA) 10889942911653, UDI/DI (CS) 40889942911654; tttt) Model Number DYNJ905335, UDI/DI (EA) 10889942875894, UDI/DI (CS) 40889942875895; uuuu) Model Number DYNJ905651B, UDI/DI (EA) 10193489925067, UDI/DI (CS) 40193489925068; vvvv) Model Number DYNJ905936B, UDI/DI (EA) 10193489422818, UDI/DI (CS) 40193489422819; wwww) Model Number DYNJ908752, UDI/DI (EA) 10195327171360, UDI/DI (CS) 40195327171361; xxxx) Model Number DYNJRA0585, UDI/DI (EA) 10888277580695, UDI/DI (CS) 40888277580696; yyyy) Model Number DYNJRA1155B, UDI/DI (EA) 10193489583717, UDI/DI (CS) 40193489583718; zzzz) Model Number DYNJRA1578, UDI/DI (EA) 10193489272789, UDI/DI (CS) 40193489272780;
Devices named in this recall
104 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- Medline Industries, Inc. DYNJ66065Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ905936BMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ65488BMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ56260BMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69017Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ905651BMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69619Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ65189CMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ62295BMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69716AMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ63354CMedline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ83931Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ83976Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline DYNJRA0585Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ905335Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ59094Medline Industries, Inc. · General surgery tray2 recalls0 reports
- Medline Industries, Inc. DYNJ63576AMedline Industries, Inc. · General surgery tray1 recall1 report
- Medline Industries, Inc. DYNJ56507BMedline Industries, Inc. · General surgery tray1 recall1 report
- Medline Industries, Inc. DYNJ64439Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ55954BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ64075AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ64930Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ60126AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJRA1578Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ59392CMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ63268BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ65995Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ66005Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ64666AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ66260Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ66258Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57303BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ62434AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ66741Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ67899Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ61099AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ68027Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ60804Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ901392DMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ61067Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ60128AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ56140BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ61243Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ61264Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ61433Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ59795AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ60343AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ61517Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJRA1155BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ901348DMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ62295Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ62540Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ62588Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ60127AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ63340Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ63892Medline Industries, Inc. · General surgery tray1 recall0 reports
- MEDLINE EBSI1289ACenturion Medical Products Corporation · Dressing change tray1 recall0 reports
- Medline Industries, Inc. DYNJ64336Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ58643AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ64393Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ56548BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ62295AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ902168DMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ68968Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ68990Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ64372AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ67487AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ66620AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ67414Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ69820Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ67504AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ80299Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ65995AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ68923AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ81343Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ908752Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ66258AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ83152Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ56392AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ83524Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ56392BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ64928AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ67504BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ84172Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ59377Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ55841BMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ59631Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ59206Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ59246Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57736AMedline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ902281Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ56392Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57149Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57514Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ57793Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ58160Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ58228Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ58304Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ58424Medline Industries, Inc. · General surgery tray1 recall0 reports
- Medline Industries, Inc. DYNJ905002AMedline Industries, Inc. · General surgery tray1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92466).
- Z-2154-2023 · Open, ClassifiedMedline Probe Cover Kits, Model Numbers: DYNJE5900, DYNJE5910, DYNJE5920, DYNJE5930, DYNJE5940, DYNJEX5900. for use during diagnostic ultrasound procedures
- Z-2155-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: ANESTHESIA OH TOTE, Model Number DYNJ61476
- Z-2156-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BLOCK PACK, Model Number DYNJ81722; b) PAIN BLOCK TRAY, Model Number DYNJRA0827A; c) PAI…
- Z-2157-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: PEDIATRIC EP PACK, Model Number DYNJ45158A
- Z-2158-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NERVE BLOCK ACCESSORY PACK, Model Number DYNJRA1220; b) PERI NEURO BLOCK PACK-LF, Model N…
- Z-2159-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY DRP PK PEDI CATH, Model Number 00-401318M; b) ANGIOGRAPHY OR, Model Number CD…
- Z-2160-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY DRAPE PACK, Model Number 00-400305R; b) ANGIO HOLLAND PACK-LF, Model Number D…
- Z-2161-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY PACK, Model Number DYNJ55252C; b) CATH LAB PACK, Model Number DYNJ55455A; c)…
- Z-2162-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATH PACK, Model Number DYNJ83105; b) TESIO, Model Number DYNJ83251; c) ANGIOGRAPHY PACK…
- Z-2163-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a. NORTON CHILDRENS CATH LAB-LF, Model Number 00-401185Q; b. PERCUTANEOUS VASCULAR CDS, Mod…
- Z-2164-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) MHC VASCULAR HYBRID PK-LF, Model Number DYNJ53433K; b) VEIN SPECIALISTS PACK, Model Numbe…
- Z-2165-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) EP PACER PACK, Model Number DYNJ67770A; b) EP PROCEDURE PACK, Model Number DYNJ67771; c)…
- Z-2166-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) SILVER CROSS OPEN HEART, Model Number DYNJ906102G; b) SILVER CROSS OPEN HEART, Model Numb…
- Z-2167-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ARTERIAL BUNDLE NO CATHETER, Model Number ART995; b) ART LINE KIT NO CATH, Model Number A…
- Z-2168-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) PORT PACK, Model Number DYNJ64675; b) PICC LINE TRAY, Model Number DYNJ64725B; c) VEIN O…
- Z-2169-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETAL SURGERY CDS-LF, Model Number CDS840214Q; b) MINOR PROCEDURE TRAY, Model Number DYND…
- Z-2170-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BREAST BIOPSY KIT, Model Number DYNDA1224B; b) U/S CORE BIOPSY KIT, Model Number DYNDH112…
- Z-2171-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: STERILE SET FOR IMPLANT-LF, Model Number OST011F
- Z-2172-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CHP COCHLEAR SUPPLEMENT CDS, Model Number CDS983069I; b) HEAD AND NECK PACK, Model Number…
- Z-2173-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETOSCOPY CDS-LF, Model Number CDS840215O; b) US BREAST BIOPSY TRAY, Model Number DYNDH…
- Z-2174-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: PACK ENDO AAA HYBRID CHRG, Model Number DYNJ60126B
- Z-2175-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATHETER INSERTION KIT CLABSI, Model Number CVI4880; b) CENTRAL CATHETER INSERTION TRA, …
- Z-2176-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CVC PACK, Model Number DYNJ46967F; b) ENDOVENOUS CORE PACK, Model Number DYNJ62557; c…
- Z-2177-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BREAST PACK, Model Number DYNDH1306; b) BRST BX TRAY, Model Number DYNDH1389B; c) PLAST…
- Z-2178-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: CATH LAB BASIN PACK, Model Number DYNJ69864A
- Z-2179-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NO-CATH RADIAL ARTERIAL PREMIU, Model Number ARTNC115; b) MAJOR VASCULAR, Model Number C…
- Z-2181-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BLOCK SUPPLIES NDL CNTR, Model Number DYNJRA1755; b) VNUS TEST KIT CFP-LF, Model Number …
- Z-2182-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) UNIVERSAL INSERTION TRAY, Model Number CVI4325; b) PPE BUNDLE KIT, Model Number DT13351A
- Z-2183-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) IV SECUREMENT KIT, Model Number DYNDV2407A; b) USG IV START KIT, Model Number DYNDV2435;…
- Z-2184-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) LACERATION TRAY, Model Number DYNDL1134A; b) ULTRASOUND GUIDED IV/MIDLINE K, Model Numbe…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page