Medline Probe Cover Kits containing sterile Ultrasound Gel… recallZ-2166-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 18, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
FDA root cause category: Device Design
What the company is doing
Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline. Questions: please contact the Recall Department at 866-359-1704
Recall details
- Recall number
- Z-2166-2023
- Status
- Open, Classified
- Initiated by the company
- May 18, 2023
- Posted by the FDA
- Jul 20, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- # included in 3,740,814 units total
- Product code
- OEZ
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
Product description
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) SILVER CROSS OPEN HEART, Model Number DYNJ906102G; b) SILVER CROSS OPEN HEART, Model Number DYNJ906102I; c) CUSTOM CARDIAC CABG, Model Number DYNJ906108C; d) VASCULAR STENT INOVA, Model Number DYNJ906440A; e) CABG, Model Number DYNJ906878A; f) VALVE, Model Number DYNJ906882A; g) VASCULAR, Model Number DYNJ906883A; h) DR WILLIAMS HEART, Model Number DYNJ907552A; i) PUH PERCUTANEOUS ECMO, Model Number DYNJ907554C; j) TAVR, Model Number DYNJ907561A; k) TAVR, Model Number DYNJ907561B; l) TAVR, Model Number DYNJ907561C; m) OPEN HEART CDS, Model Number DYNJ907607A; n) OPEN HEART CDS, Model Number DYNJ907607B; o) OPEN HEART CDS, Model Number DYNJ907607C; p) OPEN VASCULAR UE, Model Number DYNJ907858A; q) PACEMAKER, Model Number DYNJ907907; r) DEBOER HEART, Model Number DYNJ907914A; s) DEBOER HEART, Model Number DYNJ907914B; t) DEBOER HEART, Model Number DYNJ907914C; u) GENERAL HEART, Model Number DYNJ907915; v) AV FISTULA, Model Number DYNJ907975; w) AV FISTULA, Model Number DYNJ907975A; x) AV FISTULA, Model Number DYNJ907975B; y) SAC OPEN HEART DR. RUDERSDORF, Model Number DYNJ908074B; z) CARDIAC CABG B, Model Number DYNJ908121B; aa) OPEN HEART LINE SET UP, Model Number DYNJ908179A; bb) STEWARD VASCULAR SAS, Model Number DYNJ908563; cc) TAVR, Model Number DYNJ908646; dd) TAVR, Model Number DYNJ908646A; ee) ENDOVASCULAR, Model Number DYNJ908880; ff) ENDOVASCULAR, Model Number DYNJ908880A; gg) VASCULAR, Model Number DYNJ909091; hh) ENDOVASCULAR, Model Number DYNJ909253; ii) TAVR, Model Number DYNJ909270; jj) CABG, Model Number DYNJ9372030G; kk) OPEN HEART, Model Number DYNJ9718008AS; ll) ADULT CORONARY PACK, Model Number DYNJV0275M; mm) RADIOLOGY SPECIALS TRAY, Model Number DYNJVB1302A; nn) VASCULAR PACK, Model Number DYNJVB1307; oo) VASCULAR PACK, Model Number DYNJVB1307A; pp) VASCULAR EXTREMITY PPS, Model Number DYNJVB91063C; qq) ENDOVASCULAR #IMF 88370-LF, Model Number DYNJVB91067; rr) PERIPHERAL CONTINUOUS NERVE BL, Model Number PAIN1423A; ss) VEIN PACK-LF, Model Number PHS151095000; tt) E P PACK-LF, Model Number PHS41732B; uu) PICC VASCULAR ACCESS PACK, Model Number PICCNC0010; vv) TVS4000 ENDOVENOUS PACK, Model Number TVS4000L;
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
all lots within expiry, distributed from Medline between December 2017 - May 2023: a) Model Numbers DYNJ906102G, UDI/DI (EA)10195327290795, UDI/DI (CS)40195327290796; b) Model Numbers DYNJ906102I, UDI/DI (EA)10195327323349, UDI/DI (CS)40195327323340; c) Model Numbers DYNJ906108C, UDI/DI (EA)10195327118273, UDI/DI (CS)40195327118274; d) Model Numbers DYNJ906440A, UDI/DI (EA)10193489312874, UDI/DI (CS)40193489312875; e) Model Numbers DYNJ906878A, UDI/DI (EA)10193489391305, UDI/DI (CS)40193489391306; f) Model Numbers DYNJ906882A, UDI/DI (EA)10193489391237, UDI/DI (CS)40193489391238; g) Model Numbers DYNJ906883A, UDI/DI (EA)10193489398205, UDI/DI (CS)40193489398206; h) Model Numbers DYNJ907552A, UDI/DI (EA)10195327166052, UDI/DI (CS)40195327166053; i) Model Numbers DYNJ907554C, UDI/DI (EA)10195327183837, UDI/DI (CS)40195327183838; j) Model Numbers DYNJ907561A, UDI/DI (EA)10195327028848, UDI/DI (CS)40195327028849; k) Model Numbers DYNJ907561B, UDI/DI (EA)10195327290832, UDI/DI (CS)40195327290833; l) Model Numbers DYNJ907561C, UDI/DI (EA)10195327323271, UDI/DI (CS)40195327323272; m) Model Numbers DYNJ907607A, UDI/DI (EA)10193489873405, UDI/DI (CS)40193489873406; n) Model Numbers DYNJ907607B, UDI/DI (EA)10195327174194, UDI/DI (CS)40195327174195; o) Model Numbers DYNJ907607C, UDI/DI (EA)10195327335519, UDI/DI (CS)40195327335510; p) Model Numbers DYNJ907858A, UDI/DI (EA)10195327131357, UDI/DI (CS)40195327131358; q) Model Numbers DYNJ907907, UDI/DI (EA)10193489939736, UDI/DI (CS)40193489939737; r) Model Numbers DYNJ907914A, UDI/DI (EA)10195327149529, UDI/DI (CS)40195327149520; s) Model Numbers DYNJ907914B, UDI/DI (EA)10195327180607, UDI/DI (CS)40195327180608; t) Model Numbers DYNJ907914C, UDI/DI (EA)10195327329433, UDI/DI (CS)40195327329434; u) Model Numbers DYNJ907915, UDI/DI (EA)10193489943689, UDI/DI (CS)40193489943680; v) Model Numbers DYNJ907975, UDI/DI (EA)10193489965605, UDI/DI (CS)40193489965606; w) Model Numbers DYNJ907975A, UDI/DI (EA)10195327155070, UDI/DI (CS)40195327155071; x) Model Numbers DYNJ907975B, UDI/DI (EA)10195327376291, UDI/DI (CS)40195327376292; y) Model Numbers DYNJ908074B, UDI/DI (EA)10195327172688, UDI/DI (CS)40195327172689; z) Model Numbers DYNJ908121B, UDI/DI (EA)10195327095710, UDI/DI (CS)40195327095711; aa) Model Numbers DYNJ908179A, UDI/DI (EA)10195327222574, UDI/DI (CS)40195327222575; bb) Model Numbers DYNJ908563, UDI/DI (EA)10195327086428, UDI/DI (CS)40195327086429; cc) Model Numbers DYNJ908646, UDI/DI (EA)10195327123956, UDI/DI (CS)40195327123957; dd) Model Numbers DYNJ908646A, UDI/DI (EA)10195327314965, UDI/DI (CS)40195327314966; ee) Model Numbers DYNJ908880, UDI/DI (EA)10195327203436, UDI/DI (CS)40195327203437; ff) Model Numbers DYNJ908880A, UDI/DI (EA)10195327259198, UDI/DI (CS)40195327259199; gg) Model Numbers DYNJ909091, UDI/DI (EA)10195327268602, UDI/DI (CS)40195327268603; hh) Model Numbers DYNJ909253, UDI/DI (EA)10195327317027, UDI/DI (CS)40195327317028; ii) Model Numbers DYNJ909270, UDI/DI (EA)10195327317171, UDI/DI (CS)40195327317172; jj) Model Numbers DYNJ9372030G, UDI/DI (EA)10195327025434, UDI/DI (CS)40195327025435; kk) Model Numbers DYNJ9718008AS, UDI/DI (EA)10193489263121, UDI/DI (CS)40193489263122; ll) Model Numbers DYNJV0275M, UDI/DI (EA)10195327018627, UDI/DI (CS)40195327018628; mm) Model Numbers DYNJVB1302A, UDI/DI (EA)10193489928341, UDI/DI (CS)40193489928342; nn) Model Numbers DYNJVB1307, UDI/DI (EA)10193489872910, UDI/DI (CS)40193489872911; oo) Model Numbers DYNJVB1307A, UDI/DI (EA)10195327227333, UDI/DI (CS)40195327227334; pp) Model Numbers DYNJVB91063C, UDI/DI (EA)10195327257125, UDI/DI (CS)40195327257126; qq) Model Numbers DYNJVB91067, UDI/DI (EA)10193489801538, UDI/DI (CS)40193489801539; rr) Model Numbers PAIN1423A, UDI/DI (EA)10193489201826, UDI/DI (CS)40193489201827; ss) Model Numbers PHS151095000, UDI/DI (EA)10889942035601, UDI/DI (CS)40889942035602; tt) Model Numbers PHS41732B, UDI/DI (EA)10193489784817, UDI/DI (CS)40193489784818; uu) Model Numbers PICCNC0010, UDI/DI (EA)10195327231170, UDI/DI (CS)40195327231171; vv) Model Numbers TVS4000L, UDI/DI (EA)10195327223540, UDI/DI (CS)40195327223541;
Devices named in this recall
47 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ906878AMedline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
- Medline Industries, Inc. DYNJ908179AMedline Industries, Inc. · Cardiovascular procedure kit4 recalls0 reports
- Medline Industries, Inc. DYNJ908121BMedline Industries, Inc. · Cardiovascular procedure kit3 recalls0 reports
- Medline Industries, Inc. PAIN1423AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls4 reports
- Medline Industries, Inc. DYNJ9718008ASMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ9372030GMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907561AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907858AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907914AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907975AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ908074BMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907914BMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907554CMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ906102GMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907561BMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907561CMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ906102IMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907914CMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ906440AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ906882AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ906883AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. PHS41732BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJVB91067Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJVB1307Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ907607AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJVB1302AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ907907Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ907915Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ907975Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJV0275MMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ908563Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ906108CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ908646Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ907552AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ907607BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ908880Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJVB1307AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. PICCNC0010Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJVB91063CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ908880AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ909091Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ908646AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ909253Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ909270Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ907607CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ907975BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. PHS151095000Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92466).
- Z-2154-2023 · Open, ClassifiedMedline Probe Cover Kits, Model Numbers: DYNJE5900, DYNJE5910, DYNJE5920, DYNJE5930, DYNJE5940, DYNJEX5900. for use during diagnostic ultrasound procedures
- Z-2155-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: ANESTHESIA OH TOTE, Model Number DYNJ61476
- Z-2156-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BLOCK PACK, Model Number DYNJ81722; b) PAIN BLOCK TRAY, Model Number DYNJRA0827A; c) PAI…
- Z-2157-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: PEDIATRIC EP PACK, Model Number DYNJ45158A
- Z-2158-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NERVE BLOCK ACCESSORY PACK, Model Number DYNJRA1220; b) PERI NEURO BLOCK PACK-LF, Model N…
- Z-2159-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY DRP PK PEDI CATH, Model Number 00-401318M; b) ANGIOGRAPHY OR, Model Number CD…
- Z-2160-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY DRAPE PACK, Model Number 00-400305R; b) ANGIO HOLLAND PACK-LF, Model Number D…
- Z-2161-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ANGIOGRAPHY PACK, Model Number DYNJ55252C; b) CATH LAB PACK, Model Number DYNJ55455A; c)…
- Z-2162-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATH PACK, Model Number DYNJ83105; b) TESIO, Model Number DYNJ83251; c) ANGIOGRAPHY PACK…
- Z-2163-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a. NORTON CHILDRENS CATH LAB-LF, Model Number 00-401185Q; b. PERCUTANEOUS VASCULAR CDS, Mod…
- Z-2164-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) MHC VASCULAR HYBRID PK-LF, Model Number DYNJ53433K; b) VEIN SPECIALISTS PACK, Model Numbe…
- Z-2165-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) EP PACER PACK, Model Number DYNJ67770A; b) EP PROCEDURE PACK, Model Number DYNJ67771; c)…
- Z-2167-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) ARTERIAL BUNDLE NO CATHETER, Model Number ART995; b) ART LINE KIT NO CATH, Model Number A…
- Z-2168-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) PORT PACK, Model Number DYNJ64675; b) PICC LINE TRAY, Model Number DYNJ64725B; c) VEIN O…
- Z-2169-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETAL SURGERY CDS-LF, Model Number CDS840214Q; b) MINOR PROCEDURE TRAY, Model Number DYND…
- Z-2170-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BREAST BIOPSY KIT, Model Number DYNDA1224B; b) U/S CORE BIOPSY KIT, Model Number DYNDH112…
- Z-2171-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: STERILE SET FOR IMPLANT-LF, Model Number OST011F
- Z-2172-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CHP COCHLEAR SUPPLEMENT CDS, Model Number CDS983069I; b) HEAD AND NECK PACK, Model Number…
- Z-2173-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) FETOSCOPY CDS-LF, Model Number CDS840215O; b) US BREAST BIOPSY TRAY, Model Number DYNDH…
- Z-2174-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: PACK ENDO AAA HYBRID CHRG, Model Number DYNJ60126B
- Z-2175-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CATHETER INSERTION KIT CLABSI, Model Number CVI4880; b) CENTRAL CATHETER INSERTION TRA, …
- Z-2176-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CVC PACK, Model Number DYNJ46967F; b) ENDOVENOUS CORE PACK, Model Number DYNJ62557; c…
- Z-2177-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BREAST PACK, Model Number DYNDH1306; b) BRST BX TRAY, Model Number DYNDH1389B; c) PLAST…
- Z-2178-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: CATH LAB BASIN PACK, Model Number DYNJ69864A
- Z-2179-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) NO-CATH RADIAL ARTERIAL PREMIU, Model Number ARTNC115; b) MAJOR VASCULAR, Model Number C…
- Z-2180-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) VEIN KIT, Model Number DYNJ55841B; b) FACELIFT PACK, Model Number DYNJ55954B; c) CLOT R…
- Z-2181-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) BLOCK SUPPLIES NDL CNTR, Model Number DYNJRA1755; b) VNUS TEST KIT CFP-LF, Model Number …
- Z-2182-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) UNIVERSAL INSERTION TRAY, Model Number CVI4325; b) PPE BUNDLE KIT, Model Number DT13351A
- Z-2183-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) IV SECUREMENT KIT, Model Number DYNDV2407A; b) USG IV START KIT, Model Number DYNDV2435;…
- Z-2184-2023 · Open, ClassifiedMedline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) LACERATION TRAY, Model Number DYNDL1134A; b) ULTRASOUND GUIDED IV/MIDLINE K, Model Numbe…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page