Skip to content

Mega Soft recallZ-2001-2023 · Megadyne Medical Products, Inc. · initiated Jun 1, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Megadyne Medical Products, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm has received reports of patient burns in surgical procedures where device was used.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm distributed a recall notification by mail starting 06/01/2023. Customers are to provide the notice to all users of MEGADYNE MEGA SOFT and MEGA 2000 Reusable Patient Return Electrodes that participate in the cleaning, setup, and operation of affected devices. Megadyne has been made aware of 63 reports of serious patient burns since April 2018. The firm has attributed the incidence of burns to some instances that the Mega Soft Pad Instructions For Use were not being properly followed. Devices must be thoroughly rinsed after cleaning to ensure any cleaning residue is removed. Customers are asked to distribute the recall notification to all users, confirm that personnel are following the Instructions For Use, and post the provided informational attachments near the OR for staff member reference. If devices have been further distributed customers are to share the provided recall notice. Completed Business Reply Forms are to be faxed to Sedgwich at 888-214-7430 or emailed [email protected]. Any questions about the Business Reply Form can be made by phone to 888-843-0254 (Reference Event 5627). Any questions about this recall or to report a complaint can be made by phone to 1-877-384-4266, Monday thru Friday from 8:00 AM to 5:00 PM. Medical Engagement can be requested at https://www.jnjmedtech.com/mir.

Recall details

Recall number
Z-2001-2023
Status
Open, Classified
Initiated by the company
Jun 1, 2023
Posted by the FDA
Jun 30, 2023
Recalling firm
Megadyne Medical Products, Inc.
Quantity
21,100 units
Product code
GEI
FDA submissions
K133726
Distribution
Distributed US Nationwide.

Product description

MEGADYNE MEGA SOFT Universal Dual Patient Return Electrode. used during electrosurgery

Code information (models, lots, serial numbers)

Model No. 0846; UDI-DI: 10614559104248; All Units.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

6 devices cleared under K133726 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92409).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page