Mega Soft 0848Recalls & reported problems
Megadyne Medical Products, Inc. · Reusable Patient Return Electrode, Mega Soft Universal Dual Plus · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10614559104859), received between Jan 20, 2023 and Aug 19, 2026 — classified as 7 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (3 reports).
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (K133726); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 8, 2023Open, Classified
Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
Z-0600-2024 · Root cause: Under Investigation by firm · FDA record ↗
- Jun 1, 2023Open, Classified
Firm has received reports of patient burns in surgical procedures where device was used.
Z-2000-2023 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Patient Device Interaction Problem3
- Appropriate Device Problem Term/Code Not Available2
- Human-Device Interface Problem2
- Unintended Electrical Shock2
- Arcing of Electrodes1
- Electromagnetic Interference1
- Temperature Problem1
Patient problems
- Burn(s)8
- Partial thickness (Second Degree) Burn3
- Electric Shock1
- Full thickness (Third Degree) Burn1
- No Clinical Signs, Symptoms or Conditions1
- Scar Tissue1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryAug 19, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Burn(s)
MAUDE report 50329900 ↗InjuryJul 9, 2024Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE PATIENT SUFFERED FROM A BURN.
Patient problems: Burn(s)
MAUDE report 19700755 ↗InjuryJul 9, 2024Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE PATIENT SUFFERED FROM A BURN.
Patient problems: Burn(s)
MAUDE report 19700680 ↗InjuryJul 9, 2024Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE PATIENT SUFFERED FROM A BURN.
Patient problems: Burn(s)
MAUDE report 19700645 ↗InjuryJul 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE PATIENT SUFFERED FROM A BURN.
Patient problems: Burn(s); Partial thickness (Second Degree) Burn
MAUDE report 19700624 ↗MalfunctionMar 11, 2024Electromagnetic Interference; Adverse Event Without Identified Device or Use Problem; Unintended Electrical Shock; Electromagnetic Interference; Adverse Event Without Identified Device or Use Problem; Unintended Electrical Shock
IT WAS REPORTED THAT DURING A RIGHT HIP ARTHROPLASTY SURGERY, THE NEURO STIMULATOR SEEMED TO NOT FUNCTION WELL AND/OR IT WOULD HAVE INTERFERENCE. THE ANESTHESIOLOGIST (REMOVED) THEN CHANGED THE NEURO STIMULATOR AND WHEN SHE CAME TO PLUG IN THE 2ND SHE FELT A SHOCK 4TH FINGER TO THE LEFT HAND AND WHEN SHE LOOKED AT HER FINGER SHE NOTICED THAT SHE HAD A 1ST-DEGREE BURN. SHE ALSO NOTED UNUSUAL DISTURBANCES IN ECG. IT ALWAYS HAS INTERFERENCE WHEN USING ELECTROCAUTERY DURING SURGERY BUT THIS WAS DIFFERENT ACCORDING TO DR S. LOOKS LIKE ASYTOLY WHICH IS USUALLY NOT THE CASE.
Patient problems: Partial thickness (Second Degree) Burn; Superficial (First Degree) Burn
MAUDE report 18874334 ↗MalfunctionSep 6, 2023Human-Device Interface Problem; Appropriate Device Problem Term/Code Not Available; Human-Device Interface Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN ENDOSCOPIC NEURO PROCEDURE THE OR STAFF WAS BURNED BY NASAL TEMPERATURE PROBE AND PROBE TEMPERATURE CONTINUED TO RISE, WHILE USING MEGASOFT PAD. IT WAS NOTED THAT THE PATIENT'S TEMPERATURE ELEVATED 1-2 DEGREES. IT WAS RECORDED VIA A NASAL TEMP PROBE AND THE STAFF CHANGED OUT THE MEGADYNE PAD WITH A STICKY PAD AND THE ISSUE STOPPED. THE PATIENT WAS UNDER FIVE KG WITH AN INFANT BAIR HUGGER BETWEEN THE MEGADYNE PAD AND THE PATIENT.
Patient problems: Burn(s)
MAUDE report 17690536 ↗MalfunctionAug 30, 2023Human-Device Interface Problem; Appropriate Device Problem Term/Code Not Available; Human-Device Interface Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN ORTHO SPINE PROCEDURE THAT AFTER COMPLETING THE PROCEDURE IT WAS DISCOVERED THAT THE PATIENT HAD AN ALTERNATE SITE BURN ABOUT THE SIZE AND THICKNESS OF A PENCIL ERASER ON THEIR UPPER LEG WHILE THE PATIENT WAY LAYING PRONE. UPON FURTHER INSPECTION PER THE ORTHO SPINE MANAGER IT WOULD APPEAR THAT THE BOVIE PENCIL MAY NOT HAVE BEEN HOLSTERED PROPERLY AND PUNCTURED THROUGH THE DRAPES INTO THE PATIENTS LEG AND ACTIVATED. NO DEGREE OF BURN WAS PROVIDED TO THE SALES REP. TREATMENT OF THE BURN WAS NOT PROVIDED TO THE SALES REP.
Patient problems: Burn(s)
MAUDE report 17649906 ↗InjuryAug 8, 2023Adverse Event Without Identified Device or Use Problem; Temperature Problem; Unintended Electrical Shock; Adverse Event Without Identified Device or Use Problem; Temperature Problem; Unintended Electrical Shock
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE AN EMPLOYEE OF ANESTHESIA RECEIVED A SLIGHT BLOW WHILE USING THE MONOPOLAR HOOK. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: Burn(s); Electric Shock
MAUDE report 17481459 ↗InjuryJun 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A REMOVAL OF ADENOIDS, A BURN WAS DISCOVERED ON THE PATIENTS BACK.
Patient problems: Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn; Scar Tissue
MAUDE report 17134048 ↗MalfunctionJan 20, 2023Arcing of Electrodes; Arcing of Electrodes
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WHEN THEY USED THE MEGASOFT BLANKET, SPARKS FLEW AT THE TIP OF THE ELECTROCAUTERY. THE ELECTROCAUTERY'S HAVE CHANGED AND IT HAS CONTINUED TO GIVE PROBLEMS. THEY HAVE CHANGED THE MEGASOFT, AND IT HAS WORKED CORRECTLY. THERE WAS A SURGICAL DELAY BUT THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16213877 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133726. One submission often covers many models, so check the product and lot details in each record.
- Z-0597-2024 · Dec 8, 2023 · Open, ClassifiedProduct Code 0845, MEGADYNE MEGA SOFT Universal Patient Return Electrode, Reusable Patient Return Electrode
- Z-1998-2023 · Jun 1, 2023 · Open, ClassifiedMEGADYNE MEGA SOFT Universal Patient Return Electrode. used during electrosurgery
- Z-2001-2023 · Jun 1, 2023 · Open, ClassifiedMEGADYNE MEGA SOFT Universal Dual Patient Return Electrode. used during electrosurgery
- Z-1999-2023 · Jun 1, 2023 · Open, ClassifiedMEGADYNE MEGA SOFT Universal Plus Patient Return Electrode. used during electrosurgery
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.