ENDOSCOPE REPROCESSORRecalls & reported problems across all models

Olympus Medical Systems Corp. · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

8 open recalls name at least one ENDOSCOPE REPROCESSOR model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,497
Deaths reported
0
Injuries reported
39
Recalls
9

We found 3 models of ENDOSCOPE REPROCESSOR by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 1,497 adverse event reports in the FDA's MAUDE database, received between Oct 3, 2019 and Sep 30, 2026 — classified as 39 injuries, 1,458 malfunctions.

The most frequently reported specific device problem is “Break” (383 reports).

9 FDA recall records name at least one of these models by its device identifier, 8 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 13 reportsDec 2023: 21 reportsJan 2024: 13 reportsFeb 2024: 24 reportsMar 2024: 18 reportsApr 2024: 24 reportsMay 2024: 27 reportsJun 2024: 26 reportsJul 2024: 28 reportsAug 2024: 15 reportsSep 2024: 24 reportsOct 2024: 30 reportsNov 2024: 1 reportsDec 2024: 5 reportsJan 2025: 2 reportsFeb 2025: 9 reportsMar 2025: 3 reportsApr 2025: 5 reportsMay 2025: 7 reportsJun 2025: 5 reportsJul 2025: 8 reportsAug 2025: 6 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 6 reportsDec 2025: 8 reportsJan 2026: 8 reportsFeb 2026: 23 reportsMar 2026: 10 reportsApr 2026: 14 reportsMay 2026: 14 reportsJun 2026: 5 reportsJul 2026: 4 reportsAug 2026: 7 reportsSep 2026: 12 reportsOct 2026: 0 reports20242025202630
Last 36 months. The most recent month may be incomplete.

ENDOSCOPE REPROCESSOR recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Feb 12, 2026Open, Classified

    The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

    Z-1613-2026 · Root cause: Reprocessing Controls · Recall details · FDA record ↗

  2. Feb 12, 2026Open, Classified

    The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.

    Z-1612-2026 · Root cause: Reprocessing Controls · Recall details · FDA record ↗

  3. Oct 31, 2025Open, Classified

    Inform existing affected device customers about pertinent Warnings/Cautions, existing recommended maintenance schedules and that only properly trained personnel conduct repairs and maintenance.

    Z-0874-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Oct 31, 2025Open, Classified

    Inform existing affected device customers about pertinent Warnings/Cautions, existing recommended maintenance schedules and that only properly trained personnel conduct repairs and maintenance.

    Z-0873-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. Oct 31, 2025Open, Classified

    Inform existing affected device customers about pertinent Warnings/Cautions, existing recommended maintenance schedules and that only properly trained personnel conduct repairs and maintenance.

    Z-0872-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. May 26, 2023Open, Classified

    Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

    Z-2344-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  7. May 26, 2023Open, Classified

    Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

    Z-2343-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  8. May 26, 2023Open, Classified

    Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

    Z-2342-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  9. Dec 13, 2021Terminated

    The detergent and alcohol tank tubing may have been incorrectly assembled in the OER-Elite. If the tubing is incorrectly assembled, alcohol will be dispensed instead of detergent for the cleaning cycle and detergent will be dispensed instead of alcohol for the alcohol cycle, resulting in a ineffective endoscope reprocessing and patient exposure to residual detergent

    Z-0638-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported ENDOSCOPE REPROCESSOR problems

Device problems

  • Break383
  • Device Reprocessing Problem230
  • Crack229
  • Connection Problem125
  • Detachment of Device or Device Component55
  • Failure to Clean Adequately31
  • Fluid/Blood Leak29
  • Contamination27

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,148
  • No Patient Involvement286
  • No Consequences Or Impact To Patient26
  • Headache6
  • Chemical Exposure5
  • Dyspnea5
  • Viral Infection5
  • Dizziness4

ENDOSCOPE REPROCESSOR models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent ENDOSCOPE REPROCESSOR adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.