M/L Taper Hip Prosthesis Extended Offset, Reduced Neck… recallZ-2519-2023 · Zimmer, Inc. · initiated Jul 26, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The outer package labeling and product etch are for a Size 6, however, the implant is a Size 4, and vice versa.
FDA root cause category: Mixed-up of materials/components
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 7/26/23 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. You are receiving this notice because you have implanted one of the impacted stems based on our records. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. It is anticipated that the fit of the stem would have been highly detectable, for example a physical size 4 stem sitting lower than expected in a bone prepared for size 6, or vice versa and the stem sitting proud. During surgery, the available offsets in femoral head implants may have provided the necessary flexibility to appropriate
Recall details
- Recall number
- Z-2519-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 26, 2023
- Posted by the FDA
- Sep 7, 2023
- Recalling firm
- Zimmer, Inc.
- Quantity
- 24 units
- Product code
- JDI
- FDA submissions
- K200823
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, GA, MI, NJ, PA, TN, and WI. The countries of Brazil, Canada, Netherlands.
Product description
M/L Taper Hip Prosthesis Extended Offset, Reduced Neck Length, Size 4 (Item Number 00771100640) and Size 6 (Item Number 00771100440)
Code information (models, lots, serial numbers)
Item Number: 00771100640, UDI-DI: 00889024131699, Lot Number: 65236202; Item Number: 00771100440, UDI-DI: 00889024131613, Lot Number: 65249215
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- NA 00-7711-004-40Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 recall4 reports
- NA 00-7711-006-40Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 recall1 report
Other devices under the same FDA submission
27 devices cleared under K200823 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Continuum® Trabecular Metal™ 00-8757-056-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented8 reports
- Continuum® Trabecular Metal™ 00-8757-048-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented5 reports
- Continuum® Trabecular Metal™ 00-8757-054-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Trilogy® 00-8753-052-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- Continuum® Trabecular Metal™ 00-8757-060-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented4 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8758-009-28Zimmer, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer4 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8758-011-28Zimmer, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer4 reports
- Trilogy® 00-8753-048-01Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- Continuum® Trabecular Metal™ 00-8757-058-01Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 reports
- NA 802804004Zimmer, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer2 reports
- Continuum® Trilogy® Allofit® Longevity® 00-8755-010-36Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- Continuum® Trabecular Metal™ 00-8757-050-02Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page