Why it was recalled

Pump not detected as connected to controller due to software issue.

FDA root cause category: Software design

What the company is doing

An Urgent Medical Device Correction was issued vail mail October 24. 2023. Customers are advised to examine inventory for product running AIC software version: V8.4 or V8.4.1. Do not to use the product unless software is updated. Contact [email protected] to schedule the software update and return the business reply form. Forward this information to relevant personnel and any facility product may have been transferred to.

Recall details

Recall number
Z-0441-2024
Status
Terminated
Initiated by the company
Oct 10, 2023
Posted by the FDA
Dec 1, 2023
Recalling firm
Abiomed, Inc.
Product code
OZD
FDA submissions
P140003
Distribution
US: AZ, CA, FL, NY & VA.

Product description

Automated Impella Controller (AIC), product number 0042-040-US with software versions V8.4 and V8.4.1.

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 00813502011401, serial numbers IC8386, IC8391 and IC8617.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

11 devices cleared under P140003 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93406).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page