Allura Xper FD recallZ-0558-2024 · Philips Medical Systems Nederland B.V. · initiated Nov 3, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
FDA root cause category: Device Design
What the company is doing
Philips issued Urgent Medical Device Correction (UMDC) letter US customers in the US on 11/3/23 and a Field Safety Notice (FSN) to customers outside the US. Letter states reason for recall, health risk and action to take: As a remedy, Philips is replacing the existing covers of all the affected Philips Allura and Philips Azurion with the monoplane fixed ceiling mounted systems with a redesigned L-arm rotation cover (bolted instead of spring clips) to make sure it is resistant to external collisions. Philips will contact all affected customers to arrange for a Field Service Engineer visit to replace the Larm rotation cover free of charge (reference FCO72200510 for the Philips Allura systems including the Philips Poly G H5000, FCO72200511 for the Philips Azurion systems). If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Recall details
- Recall number
- Z-0558-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 3, 2023
- Posted by the FDA
- Dec 15, 2023
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 9991 units
- Product code
- OWB
- FDA submissions
- K161563, K162859, K200917
- Distribution
- Worldwide - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, CANADA, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Latvia, Latvia, Republic of Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.
Product description
Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006 Allura Xper FD20 722010 Allura Xper FD10 722012 Allura Xper FD20 722015 Allura Xper FD20 OR Table 722016 Integris H5000 C / Allura 9C 722018 Integris Allura 9 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722026 Allura Xper FD10 722028 Allura Xper FD20 722033 Allura Xper FD10 OR Table 722035 Allura Xper FD20 OR Table 722043 Integris Allura 15 & 12 (monoplane) 72246 Poly G - OMCP - VISUB - CCD (H5000)
Code information (models, lots, serial numbers)
Model/Name/UDI-DI: 722001 Allura Xper FD10C; 722003 Allura Xper FD10; 722006 Allura Xper FD20 (01)00884838059115(21); 722010 Allura Xper FD10 (01)00884838059030(21); 722012 Allura Xper FD20 (01)00884838059054(21); 722015 Allura Xper FD20 OR Table; 722016 Integris H5000 C / Allura 9C ; 722018 Integris Allura 9; - 722022 Allura Xper FD10 OR Table; 722023 Allura Xper FD20 OR Table (01)00884838059085(21;) 722026 Allura Xper FD10 (01)00884838054189(21); 722028 Allura Xper FD20 (01)00884838054202(21); 722035 Allura Xper FD20 OR Table (01)00884838054240(21); 722043 Integris Allura 15 & 12 (monoplane).
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls2 deaths3,984 reports
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths3,614 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths1,837 reports
- Allura Xper FD Allura Xper FD20/15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths776 reports
- Allura Xper FD Allura Xper FD20 OR TablePhilips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System18 recalls411 reports
- Allura Xper FD Allura Xper FD20 OR TablePhilips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System15 recalls246 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93416).
- Z-0559-2024 · Open, ClassifiedPhilips Azurion systems with monoplane fixed ceiling mounts: 722078 Azurion 7 M12 722079 Azurion 7 M20 722223 Azurion 7 M12 722224 Azurion 7 M20 722227 Azurion 5 M12 722228 Azurion…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page