VITROS 6802914Recalls & reported problems

Ortho-Clinical Diagnostics, Inc. · VITROS 3600 Immunodiagnostic System - Refurbished · Class 1

All models of VITROS (71)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
6

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

6 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jan 26, 2024Class IIOpen, Classified

    A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

    Z-1229-2024 · Root cause: Software Design Change · Recall details · FDA record ↗

  2. Jan 26, 2024Class IIOpen, Classified

    A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

    Z-1224-2024 · Root cause: Software Design Change · Recall details · FDA record ↗

  3. Jan 22, 2024Class IIOpen, Classified

    Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS 3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems which may lead to erroneous test results.

    Z-1220-2024 · Root cause: Component change control · Recall details · FDA record ↗

  4. Oct 10, 2023Class IIOpen, Classified

    During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.

    Z-0543-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. Feb 20, 2019Class IITerminated

    Luminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS Systems

    Z-1290-2019 · Root cause: Software design · Recall details · FDA record ↗

  6. Oct 19, 2018Class IITerminated

    Software anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.

    Z-0652-2019 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VITROS 6802914

Has the VITROS 6802914 been recalled?

Yes. 6 FDA recall records name this device's identifier: 4 still open and 2 closed (terminated) by the FDA. The most recent, Z-1229-2024 (Class II), began on Jan 26, 2024: A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the VITROS 6802914 FDA approved?

Devices of this type (product code JJE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Ortho-Clinical Diagnostics, Inc.

Who makes the VITROS 6802914, and is it still sold?

The FDA's device database lists Ortho-Clinical Diagnostics, Inc. as the company responsible for it (brand VITROS, model 6802914). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices

All ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page