Skip to content

Mega Soft 0835Recalls & reported problems

Megadyne Medical Products, Inc. · Reusable Patient Return Electrode, Mega Soft, Dual Cord, Adult, For Patients Weighing over 25lbs. (11.4kg), 117cm Long x 51cm Wide x 1.25cm Thick · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10614559101872), received between Feb 23, 2023 and Feb 10, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Arcing of Electrodes” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 17, 2024Open, Classified

    To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

    Z-2250-2024 · Root cause: Unknown/Undetermined by firm · FDA record ↗

  2. Jun 1, 2023Open, Classified

    Firm has received reports of patient burns in surgical procedures where device was used.

    Z-1996-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Arcing of Electrodes1

Patient problems

  • Partial thickness (Second Degree) Burn3
  • Scar Tissue2
  • Full thickness (Third Degree) Burn1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 10, 2025Arcing of Electrodes; Arcing of Electrodes

    IT WAS REPORTED THAT DURING AN UNKNOWN WHILE USING AN ELECTROCAUTERY PENCIL, THE ELECTRICITY ARCED TO THE SURGEONS HAND HOLDING A LAPAROTOMY SPONGE. THE ARC CAUSED THE SPONGE TO LIGHT ON FIRE AND INITIALLY THE SURGEON THOUGHT HER GLOVE MAY HAVE HAD A HOLE BUT THERE WAS NO INDICATION OF A BURN OR INJURY TO HER FINGER. THE SURGEON BELIEVED THE MEGASOFT PAD TO BE THE CAUSE SO THEY CONTINUED THE CASE USING A STICKY PAD AND HAD NO FURTHER ISSUES. UPON INSPECTION, THE MEGASOFT PAD DID NOT APPEAR TO HAVE ANY DAMAGE, BURNS, OR OTHER ISSUES AND THE PATIENT DID NOT SUFFER ANY INJURIES EITHER. THE SURGERY WAS DELAYED 15 MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES. FIRE WAS PUT OUT AND A STICKY PAD WAS USED.,

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21343128 ↗
  • InjuryApr 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, THERE WAS A INJURY IN A PATIENT AFTER THE SURGERY, THE PATIENT SUFFERED FROM BURN.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 16712923 ↗
  • InjuryFeb 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN ADENOTONSILLECTOMY OF A PEDIATRIC PATIENT THE PATIENT RECEIVED BURNS.

    Patient problems: Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn; Scar Tissue

    MAUDE report 16429017 ↗
  • InjuryFeb 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN ADENOTONSILLECTOMY OF A PEDIATRIC PATIENT THE PATIENT RECEIVED BURNS.

    Patient problems: Partial thickness (Second Degree) Burn; Scar Tissue

    MAUDE report 16428796 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.