WaveWriter Alpha recallZ-2617-2024 · Boston Scientific Neuromodulation Corporation · initiated Jul 17, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Neuromodulation Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
FDA root cause category: Software design
What the company is doing
On 7/17/24, "URGENT MEDICAL DEVICE ADVISORY" Notices were sent to Physicians/Surgeons, Hospitals, Healthcare Professionals informing them of the following: 1) Review any patient reports of undesired sensations indicative of potential device reset behavior during IPG charging and report these observations to Boston Scientific. 2) An in-field IPG firmware update is available for the patient s device. This will eliminate the possibility of a coincident routine device system check during IPG charging, thus preventing a potential system reset. Append the patients medical record accordingly if a firmware update is completed. 3) Patient letter is enclosed with this communication. This letter is provided for sharing with your patient and/or to be included within the patient s medical record. The patient letter can also be accessed on the Boston Scientific Product Advisories page via this link: www.bostonscientific.com/en-US/pprc/product-advisories.html 4) Post this information in a visible location near the product(s) to ensure information is easily accessible to all users. 5) Distributor customers should notify all customers that have been shipped/sold affected product about the correction. 6) Complete and return the acknowledgement form to [email protected] Any adverse events or quality concerns associated with use of this product should be reported to the firm via email at [email protected] The patient letter advises the following: discuss this letter and the device software update with your doctor and continue to follow their instructions. If you have any additional questions after contacting your doctor, please contact the firm's Support at 1-833-327-8324 (Option 2) or email to: [email protected].
Recall details
- Recall number
- Z-2617-2024
- Status
- Open, Classified
- Initiated by the company
- Jul 17, 2024
- Posted by the FDA
- Aug 14, 2024
- Recalling firm
- Boston Scientific Neuromodulation Corporation
- Quantity
- 77,674
- Product code
- LGW
- FDA submissions
- P030017
- Distribution
- Worldwide Distribution: US (nationwide) including states of: TN, PA, WA, SC, MN, OH, NC, AK, NY, IL, CA, NJ, KY, IA, MO, UT, FL, IN, LA, AL, MS, MD, WI, GA, VA, TX, AZ, KS, OR, CO, CT, ID, AR, OK, MA, MI, ND, NE, SD, WV, NV, HI, NH, DC, MT, ME, VT, WY, DE, NM. OUS (foreign) countries of: Canada, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Lebanon, Luxembourg, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Portugal, Japan, Australia, New Zealand, Singapore, Argentina, Brazil, Colombia, Puerto Rico
Product description
WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
Code information (models, lots, serial numbers)
REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365SC12160/8714729985082/100188 - 758920, M365SC12320/8714729985099/100635 - 759404
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- WaveWriter Alpha™ SC-1232Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for pain relief1 recall41 deaths4,347 reports
- WaveWriter Alpha™ 16 SC-1216Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for pain relief1 recall29 deaths2,190 reports
Product lines
Other devices under the same FDA submission
92 devices cleared under P030017 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Spectra WaveWriter™ SC-1160Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF31 deaths7,221 reports
- Precision™ SC-1110-02Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF26 deaths6,519 reports
- Precision™ Montage™ MRI SC-1200Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF22 deaths4,316 reports
- Linear™ ST SC-2218-50Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2,729 reports
- Infinion™ CX SC-2317-70Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall1 death1,268 reports
- Linear™ ST SC-2218-70Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1,175 reports
- Precision Novi™ SC-1140Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF10 deaths1,001 reports
- Infinion™ 16 SC-2316-50EBoston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF15 deaths978 reports
- WaveWriter Alpha™ Prime SC-1432Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy7 deaths971 reports
- Infinion™ CX SC-2317-50Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall871 reports
- WaveWriter Alpha™ Prime 16 SC-1416Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy12 deaths911 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.