Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10).… recallZ-0208-2025 · Percussionaire Corporation · initiated Sep 16, 2024

Class I Open, Classified All recalls posted in October 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Percussionaire Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.

FDA root cause category: Process control

What the company is doing

On 10/03/2024, the firm emailed "Urgent Medical Device Recall" Letter informing customer that Percussionaire became aware of discoloration of fluid collecting in the Phasitron breathing circuit after extended Intrapulmonary Percussive Ventilation (IPV) therapy. I was determined that there is a reaction within the nebulizer cup between the combinations of saline mixed with albuterol and/or N-acetylcysteine, and an inadvertent nickel-coating on the spring. The reaction can aerosolize the nickel under normal clinical use. Customer are instructed to: " Complete and Return Acknowledgement form (see Appendix 1) after reviewing and implementing the requested actions below by October 18, 2024. " Return all impacted products " Until unaffected or replacement product is available, affected devices should not be used under the following conditions (Appendix 3): o For the delivery of medications o For use longer than 14 days when using normal saline o On patients with an inflammatory state or with a known nickel allergy o Immediately stop usage if any unexpected symptoms are encountered " Additionally, it is recommend to rinsing the Phasitron 5 after each use with sterile water and allowing it to air dry per the IFU. Percussionaire will: Replace the affected product and exchange with new, unaffected product. Until replacement or unaffected product is available, Customers have been advised to use the product only with saline and only under restricted conditions For questions/assistance - email [email protected]

Recall details

Recall number
Z-0208-2025
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Sep 16, 2024
Posted by the FDA
Oct 30, 2024
Recalling firm
Percussionaire Corporation
Quantity
43,830 devices
Product code
NHJ
FDA submissions
K905236
Distribution
U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI, O.U.S.: Canada, Switzerland (all products shipped OUS remained in Sentec AG's (distributor, parent company of Percussionaire) control, were quarantined, and did not leave the warehouse).

Product description

Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy

Code information (models, lots, serial numbers)

Model Numbers: P5-10 P5-HC P5-UC-10 UDI-DI codes: Lot Numbers: 240321 240329 240412 240430 240509 240516 240614 240625 240627

Other devices under the same FDA submission

14 devices cleared under K905236 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.