Skip to content

Percussionaire F00012Recalls & reported problems

Percussionaire Corporation · Impulsator, 115V, English · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00849436000389), received between Jan 24, 2024 and Feb 13, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device Emits Odor” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K905236); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 18, 2026Class IIOpen, Classified

    Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

    Z-3323-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Emits Odor1
  • Infusion or Flow Problem1
  • Pressure Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Low Oxygen Saturation1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2024Pressure Problem; Pressure Problem

    PER COMPLAINT RECEIVED BY THE CUSTOMER, WHILE IN USE ON THE PATIENT, THE CAREGIVER NOTED THAT THE DEVICE WOULD NOT PERCUSS/ROVIDE REQUIRED THERAPY UNTIL IT REACHED 28 PSI.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18699247 ↗
  • MalfunctionJan 24, 2024Infusion or Flow Problem; Infusion or Flow Problem

    PER CUSTOMER COMPLAINT, THE IMPULSATOR AT HOME THERAPY DEVICE WAS IN USE ON PATIENT, WHEN IT STOPPED WORKING. IT WAS STATED BY THE CAREGIVER THAT A FUSE MAY HAVE MELTED. NO PATIENT HARM WAS STATED IN COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18573555 ↗
  • InjuryJan 24, 2024Device Emits Odor; Device Emits Odor

    PER COMPLAINT RECEIVED FROM THE CUSTOMER, WHILE PROVIDING TREATMENT TO THE CUSTOMER'S CHILD AT HOME, THE USER SMELLED A BURNT PLASTIC SMELL COMING FROM THE DEVICE. POST TREATMENT, THE CUSTOMER NOTICED THEIR CHILD'S OXYGEN LEVEL DROPPED SEVERLY, REQUIRING THE NEED TO CALL FOR EMERGENCY MEDICAL SERVICES. DURING A SECOND TREATMENT, OCCURRING THE FOLLOWING DAY, THE CUSTOMER NOTICED THEIR CHILD'S OXYGEN LEVEL DROPPED AGAIN. NO ADDITIONAL INFORMATION HAS BEEN RECEIVED FOR THIS INCIDENT.

    Patient problems: Low Oxygen Saturation

    MAUDE report 18573321 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K905236. One submission often covers many models, so check the product and lot details in each record.

  • Z-0208-2025 · Sep 16, 2024 · Open, Classified
    Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy

Other Device, Positive Pressure Breathing, Intermittent devices

All Device, Positive Pressure Breathing, Intermittent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page