Skip to content

Abbott Navitor Titan, Vision Technology, Transcatheter… recallZ-0491-2025 · Abbott Medical · initiated Oct 17, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.

FDA root cause category: Under Investigation by firm

What the company is doing

A Medical Device Recall letter was hand delivered Oct. 21, 2024 to the one US consignee for 1 Impacted Valve. Nine foreign consignees were also notified for 26 Impacted Valves distributed outside of the United States (OUS). Customers are advised to return any remaining unused Impacted Valves, if applicable, and complete and return the accompanying Acknowledgement Form to Abbott. Abbott is executing further testing to verify that long-term durability of the Impacted Valves has not been affected and will communicate any changes to the recommendations set forth in this letter if needed by Q4 2025. In the interim, Abbott recommends continuing the same local standard of care in patients implanted with the Impacted Valve as would be used with patients implanted with any other transcatheter aortic valve implant. If you have any questions, please contact your local Sales Representative or Abbott Support at 1-800-544-1664 (Option 2).

Recall details

Recall number
Z-0491-2025
Status
Open, Classified
Initiated by the company
Oct 17, 2024
Posted by the FDA
Nov 20, 2024
Recalling firm
Abbott Medical
Quantity
1 US, 9 OUS
Product code
NPT
FDA submissions
P190023
Distribution
New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

Product description

Abbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve, 35 mm, NVRO-35, Sterile

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 05415067045768, Serial Numbers: 20078589, 20357219, 20357226, 20357685, 20358319, 20358343, 20358351, 20358363, 20358368, & 20358370.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

16 devices cleared under P190023 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95627).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page