Navitor recallZ-0493-2025 · Abbott Medical · initiated Oct 17, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
FDA root cause category: Under Investigation by firm
What the company is doing
A Medical Device Recall letter was hand delivered Oct. 21, 2024 to the one US consignee for 1 Impacted Valve. Nine foreign consignees were also notified for 26 Impacted Valves distributed outside of the United States (OUS). Customers are advised to return any remaining unused Impacted Valves, if applicable, and complete and return the accompanying Acknowledgement Form to Abbott. Abbott is executing further testing to verify that long-term durability of the Impacted Valves has not been affected and will communicate any changes to the recommendations set forth in this letter if needed by Q4 2025. In the interim, Abbott recommends continuing the same local standard of care in patients implanted with the Impacted Valve as would be used with patients implanted with any other transcatheter aortic valve implant. If you have any questions, please contact your local Sales Representative or Abbott Support at 1-800-544-1664 (Option 2).
Recall details
- Recall number
- Z-0493-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 17, 2024
- Posted by the FDA
- Nov 20, 2024
- Recalling firm
- Abbott Medical
- Quantity
- 3 OUS
- Product code
- NPT
- FDA submissions
- P190023
- Distribution
- New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
Product description
Abbot Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 29 mm, NVRO-29, Sterile
Code information (models, lots, serial numbers)
UDI-DI (GTIN): 5415067045775, Serial Numbers: 20150441, 20156804, & 20170535
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Navitor™ NVRO-29St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered1 recall19 deaths318 reports
Product lines
Other devices under the same FDA submission
16 devices cleared under P190023 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Navitor™ NVRO-27St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered1 recall35 deaths471 reports
- Navitor™ NVTR-27St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered1 recall27 deaths360 reports
- Navitor™ NVRO-25St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered32 deaths385 reports
- FlexNav™ FNAV-DS-LGSt. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered15 deaths347 reports
- Navitor™ NVTR-29St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered1 recall28 deaths249 reports
- Navitor™ NVTR-25St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered21 deaths279 reports
- FlexNav™ FNAV-DS-SMSt. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered8 deaths212 reports
- Navitor Titan™ NVRO-35St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered1 recall12 deaths143 reports
- Navitor™ NVRO-23St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered11 deaths139 reports
- Navitor™ NVTR-23St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered11 deaths129 reports
- Portico™ PRT-27St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered8 deaths57 reports
- Portico™ PRT-25St. Jude Medical, Inc. · Aortic valve, prosthesis, percutaneously delivered8 deaths53 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95627).
- Z-0491-2025 · Open, ClassifiedAbbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve, 35 mm, NVRO-35, Sterile
- Z-0492-2025 · Open, ClassifiedAbbott Navitor Transcatheter Aortic Heart Valve, 29 mm, NVTR-29, Sterile
- Z-0494-2025 · Open, ClassifiedAbbott Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 27 mm, NVRO-27, Sterile
- Z-0495-2025 · Open, ClassifiedAbbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page