Impella Heart Pump recallZ-0830-2025 · Abiomed, Inc. · initiated Dec 12, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
FDA root cause category: Under Investigation by firm
What the company is doing
On December 13, 2024, URGENT MEDICAL DEVICE CORRECTION (NOTIFICATION) letters were sent to customers. RECOMMENDATIONS: Product is not being removed and hospital inventory may continue to be used. It is recommended that users lower the pump flow level to P-2 and use imaging guidance when inserting, manipulating, or removing any guidewires, indwelling venous lines or devices to reduce the potential for interaction of the device tip with the pump inlet. If optical sensor damage occurs and Placement Signal Not Reliable (PSNR) alarm is displayed no additional actions are required. Hemodynamic support and Impella position are not affected and can be monitored as part of routine patient/pump management. In addition, calculated flow and suction monitoring are also unaffected when an optical sensor PSNR alarm is displayed. In the event of a pump stop, follow the displayed prompts of the Impella Stopped alarm. Damage to the guidewire, indwelling central venous line or other device involved in the unintended interaction should be evaluated, and users should assess if a new device is required. ACTIONS TO BE TAKEN BY CUSTOMER/USER: - Product is NOT being removed from the field and does not need to be returned. - Review, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 2) to [email protected]. - Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). - If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. - Post a copy of this notice in a visible area for awareness. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program as per below instructions: - Complete and submit the report online: www.fda.gov/medwatch/report.htm or Regula
Recall details
- Recall number
- Z-0830-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2024
- Posted by the FDA
- Jan 15, 2025
- Recalling firm
- Abiomed, Inc.
- Quantity
- 2,364 units
- Product code
- PYX
- FDA submissions
- P170011
- Distribution
- Domestic: Nationwide Distribution International: Canada
Product description
Impella RP Flex with SmartAssist; Product Number: 1000323;
Code information (models, lots, serial numbers)
Show the full code information (19k characters)
Product Number: 1000323; UDI-DI: 00813502012811; Serial Numbers: 551121, 551122, 551123, 551124, 551125, 551127, 551128, 551129, 537606, 537610, 537612, 537615, 537607, 537608, 537609, 537613, 537614, 548547, 548548, 548550, 548553, 548554, 548555, 548556, 548552, 548559, 548560, 548561, 548562, 548563, 548564, 548565, 549638, 549640, 549641, 549642, 549643, 548558, 546672, 546675, 546676, 546678, 546679, 546680, 547552, 547556, 547558, 547559, 547560, 546652, 546654, 546656, 546657, 546658, 546659, 547561, 549644, 549646, 549647, 546661, 547563, 547564, 547566, 547567, 547568, 547569, 547570, 547571, 546681, 547189, 547191, 547192, 547193, 547194, 547195, 548827, 548830, 548831, 548834, 548835, 548836, 543714, 543716, 543719, 543720, 543721, 543722, 547169, 547170, 547171, 547172, 547173, 547174, 547188, 548188, 548189, 548196, 545799, 545800, 545801, 545802, 545803, 545804, 548191, 548192, 548193, 548194, 545486, 545488, 545795, 545796, 545797, 529171, 529172, 529174, 529175, 529179, 545475, 545478, 545481, 545482, 545483, 545489, 545490, 545491, 545492, 545494, 539350, 539352, 539353, 539354, 539355, 543274, 543275, 543276, 543277, 543278, 543279, 543282, 529170, 543273, 543283, 543284, 543286, 544125, 544126, 544130, 539356, 539357, 539358, 539359, 539794, 539795, 539797, 539799, 544123, 544124, 544127, 538387, 538388, 538389, 539790, 539791, 539792, 539793, 538381, 538382, 538383, 538384, 538385, 538386, 543285, 543288, 543289, 543292, 535401, 535403, 535404, 535405, 535407, 539820, 539822, 539823, 539824, 539825, 539826, 539829, 535408, 535409, 535410, 537995, 539800, 539801, 539803, 539806, 539809, 528198, 528199, 535421, 535425, 535427, 535428, 537993, 539805, 537992, 539807, 539808, 531407, 531410, 531411, 531413, 537996, 537997, 537999, 528191, 528192, 528193, 528194, 528195, 531414, 531415, 533140, 533141, 533142, 536414, 536415, 536416, 536417, 536418, 536419, 532106, 533074, 533075, 533079, 533134, 533135, 533137, 533138, 533139, 537973, 537974, 537976, 537980, 538376, 544138, 544139, 544140, 538371, 538372, 538375, 538377, 538378, 538380, 539340, 539341, 539344, 539345, 539347, 539348, 536531, 536534, 536535, 536536, 536537, 539784, 533081, 533082, 533083, 539785, 539786, 539788, 539789, 544131, 544132, 544133, 544134, 544135, 544940, 544941, 544942, 544944, 544945, 544947, 544950, 544952, 544954, 544955, 544957, 544958, 518335, 518337, 518338, 518339, 518340, 518341, 518342, 518343, 544948, 544949, 537661, 537663, 537664, 516715A, 518326A, 518328A, 518329A, 518330A, 518332A, 518333A, 516709A, 516712A, 518476A, 518477A, 518479A, 518480A, 518481A, 518482A, 516706A, 516707A, 516708A, 516767A, 516768A, 516769A, 516770A, 516771A, 516773A, 516774A, 518474A, 518475A, 536560, 536563, 536566, 536568, 536711, 537665, 537666, 537667, 537668, 537669, 535272, 535273, 536712, 536713, 536714, 536715, 536717, 536718, 533798, 533800, 533801, 533804, 533805, 535274, 535275, 535276, 535277, 535278, 535279, 531698, 531702, 531703, 531704, 531705, 531706, 533806, 533807, 531707, 535155, 535156, 535157, 535158, 535160, 535162, 535164, 536719, 531425, 531426, 531417, 531418, 531419, 531420, 531421, 535793, 535794, 535796, 535797, 535800, 535761, 535763, 535765, 535767, 535770, 535791, 535792, 535768, 535769, 524385, 526612, 526614, 526615, 526617, 526618, 528640, 528641, 528643, 526611, 529503, 530051, 530052, 530054, 530055, 530056, 530057, 530059, 526609, 526610, 527878, 527879, 527875, 527876, 527877, 529915, 529916, 529919, 529920, 529921, 529922, 529923, 531728, 531730, 531731, 531732, 531733, 531734, 531735, 531736, 533405, 533406, 533407, 533408, 533409, 533410, 533412, 533413, 530680, 530681, 530682, 530684, 530685, 530687, 530688, 530689, 526179, 526185, 533044, 533045, 533046, 533047, 533049, 533520, 533522, 533523, 533524, 533529, 526180, 526181, 526182, 526184, 526186, 526187, 527870, 527871, 527872, 532099, 532100, 532101, 532102, 536392, 536393, 536394, 536396, 536398, 533050, 533051, 533052, 533053, 527945, 527949, 528646, 528647, 528648, 528649, 529726, 529727, 529729, 529730, 529731, 529732, 529733, 529734, 527940, 527941, 527943, 527946, 527947, 527948, 529502, 529505, 529507, 529509, 529500, 529501, 529504, 529506, 529508, 528181, 528184, 528185, 528186, 528187, 528189, 529181, 529182, 529183, 529184, 529185, 529186, 529187, 529188, 529189, 530040, 530041, 530048, 527003, 529071, 529072, 529073, 529074, 529075, 529076, 529077, 530046, 530049, 524378, 524386, 527001, 527002, 527004, 527005, 527006, 527007, 524377, 524379, 524380, 524381, 524383, 524384, 524804, 524807, 524808, 524809, 524810, 524811, 524812, 521358, 524803, 524805, 524806, 521356, 521357, 521360, 521361, 521362, 521363, 521354, 521355, 521359, 525187, 525188, 525189, 525192, 525038, 525039, 525041, 525045, 525046, 525191, 525193, 525194, 525195, 525196, 526620, 526624, 526621, 526623, 526626, 526628, 526619, 521314, 521315, 521316, 521317, 521318, 521320, 521321, 521322, 521323, 521872, 521874, 526173, 526174, 521875, 521876, 521877, 521878, 521879, 521880, 524181, 524182, 524183, 524184, 524185, 525824, 525830, 525833, 526175, 526176, 526177, 526178, 519279, 519282, 519284, 519286, 519287, 519288, 521245, 525825, 525827, 525828, 525831, 525832, 519280, 519281, 519285, 521240, 521241, 521243, 521244, 521246, 521247, 525808, 525809, 526993, 526995, 526996, 526997, 525805, 525806, 525810, 525811, 525812, 526991, 526992, 519997, 525807, 525813, 526990, 519984, 519985, 519986, 519996, 520002, 520003, 519987, 519988, 519989, 519990, 519998, 519999, 520000, 520001, 519991, 519993, 524179, 524180, 522822, 522827, 522829, 522830, 524398, 524400, 524402, 524404, 517972, 517973, 517975, 517976, 517977, 517978, 517979, 517980, 524397, 524401, 524405, 523972, 523973, 523974, 523976, 523977, 523978, 523979, 523980, 523981, 519699, 519701, 522823, 522824, 522825, 517932, 517933, 517934, 517935, 519702, 519703, 519704, 519705, 519706, 519708, 522819, 522820, 522812, 522813, 522814, 522821, 519259, 519260, 521283, 522815, 522816, 522817, 522818, 521282, 521285, 521286, 521280, 521281, 521284, 521287, 519265, 519266, 519267, 519268, 519261, 519262, 519263, 514768, 514769, 514770, 516764, 516765, 519682, 519685, 519686, 519687, 519688, 514762, 514763, 514764, 514765, 514766, 516583, 516578, 516582, 516584, 516585, 514262, 514263, 516576, 516579, 516581, 500457A, 500459A, 500461A, 500462A, 500463A, 500464A, 500466A, 500685A, 500688A, 500734A, 500735A, 500736A, 500738A, 500739A, 500742A, 501001A, 501002A, 500994A, 500995A, 500997A, 500998A, 500999A, 501000A, 516757, 516758, 516759, 516760, 516763, 513149, 513150, 513151, 513152, 514241, 514242, 514243, 514244, 514245, 514246, 514248, 514264, 514265, 514266, 514267, 514268, 514269, 513153, 514782, 514783, 514786, 514789, 513154, 513155, 513156, 513157, 514781, 497569A, 497570A, 501475A, 501476A, 501480A, 501482A, 497568A, 497571A, 497574A, 501473A, 501477A, 501478A, 501479A, 501481A, 503309A, 503310A, 503313A, 503314A, 503316A, 503534A, 503535A, 503536A, 503537A, 503539A, 503540A, 503543A, 505090A, 505091A, 505092A, 505093A, 505094A, 505095A, 505096A, 496788A, 496789A, 496790A, 496791A, 496792A, 496793A, 496795A, 497572A, 497573A, 497575A, 497576A, 497577A, 499008A, 499009A, 499010A, 499011A, 499012A, 499013A, 499014A, 499015A, 499017A, 503315A, 503317A, 497129A, 497133A, 497135A, 497136A, 497137A, 497528A, 497529A, 497530A, 497531A, 497533A, 497535A, 497123A, 497124A, 497125A, 497126A, 497127A, 497130A, 497131A, 497132A, 497534A, 497536A, 497537A, 498591A, 498593A, 498594A, 498595A, 498597A, 498589A, 498590A, 498596A, 498258A, 498259A, 498262A, 498263A, 498266A, 498260A, 498261A, 498264A, 498265A, 498267A, 499938A, 499939A, 499940A, 499941A, 499943A, 499945A, 499947A, 495826A, 495827A, 495828A, 495829A, 497139A, 499944A, 500227A, 500228A, 500229A, 500230A, 495820A, 495824A, 495825A, 499951A, 499952A, 499953A, 499955A, 499957A, 496653A, 496659A, 497541A, 497543A, 499026A, 499948A, 499949A, 499956A, 500222A, 500224A, 500225A, 500226A, 496656A, 497544A, 497545A, 499021A, 499022A, 500223A, 496654A, 496657A, 496658A, 497540A, 497546A, 499018A, 499019A, 516570, 516572, 518984, 518985, 518986, 518987, 518989, 518990, 518991, 518992, 518993, 516573, 516574, 516566, 516567, 516568, 516569, 516513, 516515, 516571, 519669, 519670, 519672, 519673, 519675, 519676, 519677, 519679, 519680, 519681, 516514, 516517, 516519, 516536, 516542, 516544, 516545, 513137, 516692, 516693, 516694, 516695, 516697, 516699, 518429, 518430, 518433, 518435, 518436, 518437, 518438, 518965, 518966, 518967, 518968, 518969, 518970, 518971, 518972, 518973, 514760, 517883, 517888, 512881, 512883, 513129, 513130, 513131, 513132, 513133, 513134, 513136, 517882, 517885, 517887, 517889, 517891, 512812, 512813, 512814, 512815, 512816, 512817, 512818, 512874, 512875, 512876, 512877, 512878, 512879, 512880, 512809, 512810, 512811, 511439, 511461, 511826, 511827, 511828, 513148, 513494, 513497, 513498, 513499, 513501, 513502, 513503, 513139, 513140, 513142, 513143, 513144, 513145, 513146, 513147, 506304, 512760, 512762, 512763, 512767, 512768, 514751, 514752, 514753, 514754, 514755, 514756, 514757, 514758, 514759, 511819, 511820, 511821, 511822, 511823, 511824, 511825, 513604, 513605, 513606, 513607, 513608, 513611, 513612, 513613, 507534, 507535, 507536, 507537, 507538, 507539, 507540, 507542, 507543, 506305, 506306, 506307, 506308, 506309, 506311, 506312, 506313, 512769, 512771, 512772, 512776, 512777, 512773, 512774, 512775, 512778, 514232, 514233, 514234, 514235, 514236, 514237, 513616, 513618, 513619, 513620, 513621, 513623, 514238, 514239, 513614, 513615, 511443, 511830, 511831, 511832, 511833, 511834, 511835, 511836, 511837, 511437, 511440, 511441, 512461, 512462, 512457, 512458, 512459, 512460, 512464, 512465, 512467, 512472, 511435, 512455, 512456, 512469, 512470, 512471, 511434, 511436, 512463, 511444, 511445, 511448, 511451, 511454, 511455, 511456, 511458, 511459, 511460, 511463, 510914, 510915, 510916, 510899, 510900, 510902, 510903, 510905, 510907, 510908, 510909, 510911, 510912, 510913, 511446, 511447, 509348, 509349, 509350, 509351, 509354, 509356, 509357, 509355, 509921, 509922, 509923, 509930, 509924, 509929, 507810, 506536, 507804, 507805, 507806, 507807, 507808, 504771, 504772, 504773, 504775, 504776, 504777, 506529, 506531, 506532, 506535, 504769, 504770, 504018, 504019, 504020, 504021, 507334, 507335, 504015, 504016, 504022, 507330, 507332, 507337, 507811, 504023, 506537, 507338, 507812, 506538, 507329, 510571, 510573, 510574, 510575, 510580, 507604, 507606, 507609, 507611, 507613, 510546, 510547, 510550, 510552, 510554, 510577, 508406, 508409, 508411, 508412, 508413, 500468, 506824, 506825, 506832, 507143, 507144, 510141, 510142, 510143, 510145, 510146, 510147, 510155, 510170, 506827, 506831, 510150, 500470, 500469, 500472, 500475, 507146, 507147, 507148, 500467, 500471, 500473, 500474, 507145, 501658, 501659, 501661, 510151, 510152, 510169, 501660, 501662, 501663, 510157, 510159, 510160, 501664, 501665, 501667, 510162, 510165, 510167, 510168, 501666, 510532, 510526, 510528, 510529, 510530, 510535, 510533, 510534, 509952, 509953, 509958, 509959, 509954, 509957, 506294, 506295, 506298, 506299, 506300, 506301, 506303, 506297, 505731, 505733, 505734, 505735, 505736, 506545, 506546, 505732, 506407, 506409, 506540, 506541, 506542, 506406, 506408, 506539, 506544, 506411, 506412, 506413, 505087, 505088, 505894, 504033, 504429, 504443, 505427, 505431, 504029, 504030, 504031, 504032, 504430, 504432, 504433, 504435, 504442, 504025, 504028, 504438, 504440, 504791, 503837, 504444, 503834, 503892, 501314, 505081, 505085, 505896, 505899, 505900, 505905, 505424, 505902, 505904, 505906, 505907, 505908, 505909, 505911, 505912, 505913, 504024, 504026, 504436, 504437, 504793, 504795, 504796, 504797, 505079, 505428, 505429, 505432, 505433, 501308, 501309, 501310, 501311, 501312, 501316, 501317, 502041, 502042, 502043, 502044, 502046, 503836, 503889, 503890, 503891, 503893, 503894, 503895, 503896, 503897, 504446, 504447, 504789, 504792, 502037, 502038, 502039, 502040, 503838, 503839, 503840, 503841, 503842, 496231, 496232, 496233, 496234, 496235, 496236, 496237, 496238, 509256, 509257, 509248, 509249, 509250, 509251, 509253, 509340, 509341, 509343, 509344, 509345, 509346, 509347, 492444, 492445, 492446, 492447, 492449, 492451, 492452, 492453, 494355, 494357, 494359, 494356, 494358, 494363, 489561, 489563, 489564, 489565, 489566, 489567, 489568, 493861, 493862, 493863, 493865, 493866, 493867, 493868, 493869, 494344, 494345, 494346, 494349, 494351, 494353, 495835, 495836, 495837, 495838, 495839, 496665, 496667, 496668, 496669, 495162, 495164, 495168, 496243, 495163, 495830, 495070, 495075, 495076, 495161, 495831, 495073, 495074, 495165, 495311, 495317, 496247, 495069, 495166, 495309, 495310, 495316, 495318, 496242, 496249, 495078, 495832, 495833, 495071, 495072, 494376, 495135, 495136, 495138, 495139, 495141, 495143, 495142, 494375, 494369, 494370, 494372, 494373, 494374, 494377, 494378, 495140, 495159, 493530, 493532, 493533, 493534, 493975, 493529, 493970, 493527, 493528, 493531, 493535, 493526, 493974, 493979, 488964, 492487, 492492, 494335, 492485, 493191, 494338, 493186, 492484, 493187, 493193, 494337, 492493, 492488, 492490, 492491, 494334, 494341, 492486, 492489, 493188, 493189, 493192, 493194, 492206, 492207, 492208, 492209, 492211, 490454, 490456, 490457, 490459, 490460, 490461, 490462, 492205, 442295A, 490455, 490458, 484758, 484763, 484765, 484766, 492252, 492627, 492629, 492253, 492633, 492246, 484761, 492244, 492245, 492248, 492625, 492626, 492630, 492631, 492251, 490250, 490247, 490243, 490244, 490246, 490248, 490249, 490245, 490251, 489554, 490446, 488989, 488992, 488993, 489545, 489546, 489547, 489548, 489549, 489551, 490254, 490255, 490256, 490257, 490258, 490259, 490447, 490448, 490452, 488987, 488988, 488991, 489553, 490260, 490449, 490450, 490262, 490444, 490445, 488968, 490434, 490437, 490438, 490441, 490442, 490439, 488966, 488967, 486478, 488608, 488732, 488965, 486475, 486477, 488607, 488609, 488733, 488734, 488962, 486476, 488725, 488729, 486470, 486471, 486473, 488600, 488601, 488602, 488726, 488727, 488728, 488730, 488731, 488960, 488604, 488605, 488606, 488603, 485106, 485108, 483967, 485107, 485111, 484291, 484292, 485110, 483968, 485109, 485387, 485388, 483961, 483964, 483965, 483966, 485102, 485103, 485105, 485382, 485383, 485390, 485391, 483959, 483960, 485104, 484290, 485386, 484284, 484285, 484288, 484289, 484293, 485384, 482360, 482361, 488583, 488587, 483299, 483300, 483301, 483305, 488585, 482356, 483302, 483312, 483313, 483315, 483963, 488581, 483307, 483317, 482352, 482358, 482359, 483298, 483308, 483309, 483310, 483316, 488584, 482354, 482355, 482357, 486114, 486120, 485692, 485694, 485695, 486109, 486112, 486115, 486117, 486121, 486125, 486129, 486130, 486131, 486135, 486136, 485697, 485699, 486118, 486122, 486124, 486128, 486132, 486133, 485693, 485700, 485701, 486107, 488586, 488589, 482203, 482346, 482350, 482351, 482198, 482206, 482338, 482348, 482204, 482205, 482197, 482199, 482334, 482344, 482202, 482333, 482337, 482332, 482335, 482339, 482341, 482345, 482347, 482142, 482145, 482147, 482193, 444471, 484756, 482146, 482148, 482225, 484747, 444470, 482190, 482191, 482217, 482223, 482224, 484751, 484754, 444477, 444469, 482150, 482192, 482218, 482220, 482189, 482219, 444472, 482149, 484749, 444473, 444474, 444475, 444476, 444478, 482187, 482221, 484750, 484752, 484753, 444241, 443183, 443184, 443188, 444246, 444247, 444788, 443185, 443187, 443191, 443492, 444243, 444248, 444784, 443192, 443487, 443488, 444244, 444245, 444816, 454662, 444810, 454653, 454654, 454661, 444811, 454604, 454648, 454649, 454650, 454657, 454659, 444818, 454643, 444815, 454655, 452533, 452539, 452457, 452458, 452459, 452461, 452462, 452463, 452464, 452465, 452530, 452536, 449676, 449679, 453939, 453940, 442458, 442459, 449675, 449682, 451977, 453933, 449890, 442453, 449895, 449898, 442202, 442203, 442204, 442455, 449896, 452037, 452039, 452041, 452042, 442206, 442207, 452035, 442209, 442210, 442211, 442457, 452034, 452036, 441272, 441273, 440011, 440167, 440168, 440169, 440171, 440172, 440173, 440174, 441276, 441279, 441281, 441275, 439837, 439840, 440927, 440929, 440931, 440935, 442258, 440932, 441322, 441324, 441326, 441327, 442292, 442293, 441323, 442294, 442298, 442299, 442300, 442301, 442421, 442257, 442416, 442417, 442419, 442420, 442422, 442413, 442414, 442418, 439833, 439839, 440005, 439835, 440003, 440006, 439538, 439541, 439543, 439495, 439497, 439498, 439499, 439500, 439535, 439537, 440984, 440985, 439536, 438640, 438642, 438646, 438837, 438838, 438841, 440191, 438835, 438840, 439006, 439007, 439008, 439009, 436502, 439004, 439005, 438644, 438648, 438843, 439002, 440189, 438647, 440187, 436875A, 438842, 439000, 439001, 438607, 438606, 437526, 438608, 437531, 438777, 438773, 437528, 437530, 437799, 438645, 438771, 437796, 438600, 438776, 437794, 438770, 438775, 437800, 438778, 437792, 437801, 438769, 437793, 436512, 436508, 436510, 436515, 436877, 436879, 436880, 436873, 436874, 436506, 436507, 436509, 436514, 436876, 436881, 437523, 436511, 436501, 436504, 436505, 436500, 436503, 436498, 437788, 437791, 434516A, 437540, 437783, 437784, 437786, 437561, 437558, 434512, 437535, 434510, 434511, 436093, 433954, 434517, 433949, 436101, 433952, 433953, 436094, 437541, 434508, 434831, 436096, 436098, 436100, 437537, 437536, 437539, 437538, 436092, 436097, 434513, 437529, 433955, 434514, 434515, 434805, 436148, 433751, 433748, 434808, 436146, 434799, 436151, 436149, 436150, 436143, 434837, 436145, 434832, 434838, 434834, 434836, 436142, 434833, 433923, 433926, 434830, 433925, 434806, 433927, 434801, 434800, 434802, 434803, 433753, 432652, 433749, 433752, 433754, 434543, 434545, 434546, 433757, 434538, 434541, 434542, 433755, 433756, 434540, 432645, 432646, 432648, 432643, 432644, 429654, 429652, 432650, 432651, 429653A, 432400, 432401, 432402, 432397, 432398, 429646, 429648, 429650, 429649, 429651, 430363, 430366, 430362, 430367, 430365, 432394, 432396, 430360, 432393, 432395, 431848, 431847, 431846, 431845, 431536, 431844, 431851, 431868, 431870, 431537, 431849, 431850, 431869, 431535, 431843, 431531, 431529, 431530, 431532, 431533, 431534, 431099, 431101, 431098, 431103, 428725, 431106, 431105, 431104, 428730, 428731, 428733, 428726, 428729, 428734, 425591, 425592, 425593, 425595, 425596, 425599, 430927, 430929, 430931, 425600, 429616, 429617, 429619, 429620, 430922, 430923, 430924, 429621, 430926, 429623, 424002, 424006, 424000, 424004, 424007, 422974, 422975, 422978, 422979, 422972, 423999, 423998, 424001, 422981, 422980, 418933, 418935, 418937, 418939, 418934, 418938, 418941, 418940, 419490, 419491, 419496, 421953, 419489, 419495, 421955, 419494, 420695, 420696, 420697, 421960, 420698, 420699, 421958, 421952, 421957, 421956, 409537, 409541, 409543, 409544, 409546, 416645, 416646, 416653, 416644, 416650, 409542, 416647, 416649, 416651, 411373, 411374, 411376, 411377, 411372, 411378, 411379, 411380, 410413, 410411, 410412, 410408, 410409, 410405, 410410, 410406, 410407, 407255A, 407988A, 407991A, 407989A, 407994A, 407253A, 407992A;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Impella 1000323Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump6 recalls189 deaths943 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under P170011 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Impella 0042-0000-USAbiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump12 recalls120 deaths2,456 reports
- Impella 0042-0040-USAbiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump8 recalls41 deaths1,305 reports
- Impella 004334Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump3 recalls61 deaths408 reports
- Impella 0046-0035Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump4 recalls37 deaths258 reports
- Impella 1000200Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump9 deaths84 reports
- Impella 0043-0003Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump1 recall7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95780).
- Z-0829-2025 · Open, ClassifiedImpella RP with SmartAssist; Product Number: 0046-0035;
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page