Skip to content

MiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715… recallZ-1259-2025 · Medtronic MiniMed, Inc. · initiated Jan 31, 2025

Class II Open, Classified All recalls posted in March 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic MiniMed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

FDA root cause category: Device Design

What the company is doing

On January 31, 2025, Medtronic MiniMed issued a "Urgent Medical Device Correction" notification to all affected consignees. The notification was distributed via: e-mail, certified mail, SMS text message or phone call. 1. Monitor Your Glucose Levels: 2. Discuss how to prepare for situations like this with your healthcare professional. 3. Respond to Alerts and Symptoms. 4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter by either completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting [https://info.medtronicdiabetes.com/pump-delivery-volume] to acknowledge this communication.

Recall details

Recall number
Z-1259-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jan 31, 2025
Posted by the FDA
Mar 3, 2025
Recalling firm
Medtronic MiniMed, Inc.
Quantity
70587 units
Product code
OZO
FDA submissions
P150001
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY and the countries of Algeria, Andorra, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State Of Bosnia, Herzegovina , Botswana, Brazil, Brunei, Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Cayman, Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic, Republic Of Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Malawi, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Republic Of Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion Romania, Russia, Saint Pierre And Miquelon, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Tanzania, United Republic Of Thailand, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic Of Vietnam, Yemen, Zimbabwe.

Product description

MiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755)

Code information (models, lots, serial numbers)

All Lot Numbers; All UDI numbers

Other devices under the same FDA submission

39 devices cleared under P150001 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 96200).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page