Shiley recallZ-1390-2025 · Covidien · initiated Feb 26, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.
FDA root cause category: Under Investigation by firm
What the company is doing
On 2/26/2025, recall notices were mailed and/or emailed to Risk Managers, Directors of Respiratory Care, Anesthesiology, Pulmonary/Intensive Care and ENT who were asked to do the following: 1) For patients with affected devices currently in place, a replacement is required. The patient's medical team should assess the overall patient risk when considering the timing of replacement. Clinicians should continue to follow current product Instructions For Use (IFU) along with facility specific policies and procedures. 2) Quarantine and return all unused product from the affected lot in your inventory to the firm. 3) Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product from the specified lot has been transferred or distributed. 4) Complete and return the enclosed Customer Confirmation Form via email to [email protected] If you have any questions regarding this communication, contact the firm's Customer Service at 800-962-9888, Option 2.
Recall details
- Recall number
- Z-1390-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 26, 2025
- Posted by the FDA
- Mar 25, 2025
- Recalling firm
- Covidien
- Quantity
- 780
- Product code
- JOH
- FDA submissions
- K150844
- Distribution
- OUS distribution only to countries of: France, Germany, Ireland, Italy, Lithuania, Portugal, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Taiwan
Product description
Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R
Code information (models, lots, serial numbers)
UDI-DI/GTIN: A8845212054401, 20884521205441, 10884521205444. Lot: 202405258X. Expiration: 06/01/2029
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
Product lines
Other devices under the same FDA submission
13 devices cleared under K150844 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Shiley 8CN85RCovidien LP · Tube tracheostomy and tube cuff1 recall86 reports
- Shiley 5CN70RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death22 reports
- Shiley 10CN10RCovidien LP · Tube tracheostomy and tube cuff1 recall16 reports
- Shiley 6UN75RCovidien LP · Tube tracheostomy and tube cuff1 recall14 reports
- Shiley 4CN65RCovidien LP · Tube tracheostomy and tube cuff1 recall11 reports
- Shiley 8UN85RCovidien LP · Tube tracheostomy and tube cuff1 recall6 reports
- Shiley 4UN65RCovidien LP · Tube tracheostomy and tube cuff1 recall5 reports
- Shiley 5UN70RCovidien LP · Tube tracheostomy and tube cuff1 recall3 reports
- Shiley 10UN10RCovidien LP · Tube tracheostomy and tube cuff1 recall3 reports
- Shiley 9UN90RCovidien LP · Tube tracheostomy and tube cuff1 recall1 report
- Shiley 9CN90RCovidien LP · Tube tracheostomy and tube cuff1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.