Skip to content

CADD recallZ-1662-2025 · Smiths Medical ASD, Inc. · initiated Apr 10, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Pumps may experience Wireless Connection Modules intermittent connection alarms, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

FDA root cause category: Software design

What the company is doing

Firm began notifying customers on April 10, 2025 via letters titled "Urgent Medical Device Correction." Customers were instructed to validate any wireless network changes or updates to hospital network settings with the pump/communication module to ensure compatibility of the wireless connection before deploying network setting changes or updates to the production environment. If the alarm recurs or is unable to be cleared by power cycling the pump, the customer should revert to prior compatible network settings or turn off the wireless function of the communication module. The firm will also address the issue with a software update. Smiths Medical will contact customers to schedule implementation of the software update once it is available. ***Updated 5/28/2026*** On 5/27/26, firm revised the scope of the action to apply specifically to CADD customers using the wireless system. All current wireless customers, as well as any customers who upgrade to wireless, will receive the updated CADD firmware connection. All future implementations - both wireless and non-wireless - will include the corrected CADD firmware.

Recall details

Recall number
Z-1662-2025
Status
Open, Classified
Initiated by the company
Apr 10, 2025
Posted by the FDA
May 7, 2025
Recalling firm
Smiths Medical ASD, Inc.
Quantity
US: 83,833 items ; OUS: 71,843 items
Product code
MEA
FDA submissions
K170982
Distribution
Worldwide - US Nationwide distribution.

Product description

CADD Solis HSPCA Pump ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope: PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA 21-2111-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA 21-2101-51 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA 21-2111-0401-78 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA 21-2111-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, ENGLISH 1/EA 21-2111-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA 21-2112-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA 21-2111-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA 21-2112-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA 21-2111-0402-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA 21-2112-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA 21-2112-0402-51 PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA 21-2111-0100-51 PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA 21-2112-0100-51

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Affected CADD-Solis Software Versions: 4.0, 4.1, 4.2, 4.2.1, 4.3 UDI-DI: 4546027095675 4546027095682 10610586034787 10610586038778 10610586038785 10610586038808 10610586038822 10610586039232 10610586041037 10610586041051 15019517099690 15019517101003 15019517101010 15019517101034 15019517101041 15019517128697 15019517135343 15019517154290 15019517154887 15019517154924 15019517154979 15019517154986 15019517154993 15019517155006 15019517155013 15019517155044 15019517246735 15019517247008 15019517247169 15019517247176 15019517247183 15019517247206 00610586038900 00610586039310 04546027095620 04546027095637 04546027095644 04546027095668 04546027095699 04546027095705 04546027095712 04546027095729 04546027095736 04546027095897 04546027095903 04546027095910 04546027422983 04546027424116 04546027424130 04546027424147 04546027424178 04546027424185 04546027424192 04546027424208 04546027424598 05019517154934 05019517155108 10610586031854 10610586032950 10610586033933 10610586033940 10610586035357 10610586035395 10610586035401 10610586035418 10610586035432 10610586035456 10610586035494 10610586035548 10610586038792 10610586038839 10610586038860 10610586038891 10610586038914 10610586038921 10610586038938 10610586038952 10610586038969 10610586038983 10610586038990 10610586039010 10610586039041 10610586039065 10610586039072 10610586039089 10610586039195 10610586039256 10610586039270 10610586039409 10610586039416 10610586039430 10610586039447 10610586039454 10610586039461 10610586039478 10610586039485 10610586039492 10610586041822 30610586035337 30610586035344 30610586035368 30610586035511 30610586039205 30610586039397 35019517128677 35019517128684 35019517128691 35019517128721 35019517128738 35019517128745 35019517128769 35019517128783 35019517128837 35019517135316 35019517135323 35019517135330 35019517135354 35019517135361 35019517154850 35019517154867 35019517154874 35019517154904 35019517154911 35019517154942 35019517154959 35019517154966 35019517155055 35019517155062 35019517155079 35019517155086 35019517155130

Devices named in this recall

9 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

11 devices cleared under K170982 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page