Medtronic CareLink SmartSync Device Manager, Model Number… recallZ-2169-2025 · Medtronic, Inc. · initiated Jun 18, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.
FDA root cause category: Software design
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letter dated 6/18/25 was sent to customers. Customer Actions: "Update SmartSync (to version 4.2.3 or higher). Refer to release notes provided by your local Medtronicrepresentative for update instructions. If needed, Medtronic representatives are available to work with you toinstall or update the SmartSync application(s) on your tablet. See Appendix for instructions on how to verify thesoftware update is complete. "Sign and return the enclosed confirmation form and share this notice with those who need to be aware withinyour organization or with any organization where SmartSync may be in use. Questions regarding this informationshould be directed to the local Medtronic Representative or Medtronic Programmer Technical Services at 800-638-1991, Option 1.
Recall details
- Recall number
- Z-2169-2025
- Status
- Open, Classified
- Initiated by the company
- Jun 18, 2025
- Posted by the FDA
- Jul 23, 2025
- Recalling firm
- Medtronic, Inc.
- Quantity
- 22091 total devices
- Product code
- NVZ
- FDA submissions
- P890003
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- CareLink SmartSync™ Azure Astra Application D00U003Medtronic, Inc. · ANALYZER, PACEMAKER GENERATOR FUNCTION4 recalls0 reports
- CareLink SmartSync™ Percepta Serena Solara Application D00U004Medtronic, Inc. · Pacemaker/icd/crt non-implanted components4 recalls0 reports
Other devices under the same FDA submission
83 devices cleared under P890003 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA 229047Medtronic, Inc. · Pacemaker/icd/crt non-implanted components1 recall662 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls559 reports
- Medtronic CareLink® Programmer 2090WMedtronic, Inc. · Programmer, pacemaker3 recalls489 reports
- CareLink Encore™ 29901AMedtronic, Inc. · Pacemaker/icd/crt non-implanted components2 recalls474 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls422 reports
- CareLink Encore™ 29901Medtronic, Inc. · implantable pacemaker Pulse-generator3 recalls265 reports
- Medtronic CareLink® Programmer 2090WMedtronic, Inc. · Programmer, pacemaker3 recalls208 reports
- CareLink SmartSync™ 24970AMedtronic, Inc. · TESTER, PACEMAKER ELECTRODE FUNCTION308 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls151 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Programmer, pacemaker2 recalls198 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Programmer, pacemaker3 recalls108 reports
- Medtronic CareLink® Programmer 2090AA4Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls82 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97130).
- Z-2168-2025 · Open, ClassifiedMedtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page