Venclose digiRF Generators, Model VCRFG1, with software… recallZ-2638-2025 · Bard Peripheral Vascular Inc · initiated Aug 21, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.
FDA root cause category: Software Design Change
What the company is doing
On 8/21/2025, correction notices were mailed or emailed to Health Care Professionals (HCPs), Risk Manager, Head Nurse, Nurse Manager, Medical Device Safety Officer, Procurement Officer, Office Manager, and Distributors, were advised of the following: If the generator is operating with affected software and is connected to a catheter and the software check fails, a red "X" will be displayed on the generator screen without an accompanying error code Actions: 1) Firm Representative will contact your facility to arrange the restoration of your facility's generator to software version 3.17 upon receiving regulatory clearance of the 510(k) premarket notification. 2) Your facility will become eligible to receive future shipments of Venclose EVSRF Ablation Catheters only after the Venclose digiRF Generator has been verified to run software version 3.17. 3) Complete and return the customer response form via email to [email protected] 4) Post the Customer Notification Letter on or near the generator until the software has been restored to software version 3.17 by a firm representative. 5) Share this notification with all product users within your facility network to ensure they are also aware of this product correction. If the affected products have been distributed further, please ensure that these entities are contacted and informed of the recall. Please contact your firm representative if you require additional assistance with this process: North American Regional Complaint Center, 1-844-823-5433, [email protected]
Recall details
- Recall number
- Z-2638-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 21, 2025
- Posted by the FDA
- Sep 26, 2025
- Recalling firm
- Bard Peripheral Vascular Inc
- Quantity
- 1,725
- Product code
- GEI
- FDA submissions
- K160754, K211806, K250068
- Distribution
- Worldwide - US Nationwide distribution including in the states of NJ, NY, FL, LA, VA, CA, TX, NV, PA, NE, TN, IL, AZ, SC, AK, GA, CO, HI, OH, NC, WI, WV, MD, MO, AL, OK, MS, PR, KS, UT, CT, IA, MI, IN, MA, KY, WA, MT, MN, SD, ME, DE, NM, AR and the country of Malaysia.
Product description
Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).
Code information (models, lots, serial numbers)
UDI-DI: 00858254006022. Generator Software Version 3.35
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Venclose System digiRF Generator VCRFG1Venclose Inc · Electrosurgical, Cutting & Coagulation & Accessories1 recall63 reports
Other devices under the same FDA submission
4 devices cleared under K160754, K211806, K250068 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Venclose System EVSRF Catheter VC10A256F60Venclose Inc · Electrosurgical, Cutting & Coagulation & Accessories1 recall1 death68 reports
- Venclose System EVSRF Catheter VC10A256F100Venclose Inc · Electrosurgical, Cutting & Coagulation & Accessories1 recall52 reports
- Venclose Maven Catheter VC056FVenclose Inc · Electrosurgical, Cutting & Coagulation & Accessories3 reports
- Venclose™ EVSRF™ Catheter VC-10A2.5-6F-100 [10X]Venclose Inc · Electrosurgical, Cutting & Coagulation & Accessories2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page