V6-2 Transducer Probe recallZ-2403-2025 · Philips Ultrasound, Inc · initiated Jul 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Ultrasound, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
FDA root cause category: Device Design
What the company is doing
An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377.
Recall details
- Recall number
- Z-2403-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 3, 2025
- Posted by the FDA
- Aug 22, 2025
- Recalling firm
- Philips Ultrasound, Inc
- Quantity
- 171,322 units
- Product code
- IYN
- FDA submissions
- K062247, K132304, K153480, K160807, K081661
- Distribution
- US Nationwide distribution.
Product description
V6-2 Transducer Probe
Code information (models, lots, serial numbers)
Show the full code information (41k characters)
Model No. 989605415042, 989605437982, 989605438461, 989605357563, 989605357561, 989605387201, 989605415041, 989605449011, 989605395631, 989605454101; UDI: (01)00884838061460(21)B1VJQF, (01)00884838061460(21)B2RL4B, (01)00884838061651(21)B20WVG, (01)00884838061651(21)B28DGB, (01)00884838061651(21)B304XN, (01)00884838061651(21)B35CH3, (01)00884838061651(21)B20WW0, (01)00884838061651(21)B1YQQ6, (01)00884838061651(21)B1XLYC, (01)00884838061651(21)B1YQZR, (01)00884838061651(21)B1P7QY, (01)00884838061651(21)B2G0CC, (01)00884838061651(21)B2LN7C, (01)00884838061651(21)B2RL6L, (01)00884838061651(21)B1P7TG, (01)00884838061651(21)B2RL9K, (01)00884838061651(21)B1P7Q8, (01)00884838061651(21)B20WM1, (01)00884838061651(21)B20WKR, (01)00884838061651(21)B1P7PJ, (01)00884838061651(21)B1XLYW, (01)00884838061651(21)B1XLZ0, (01)00884838061651(21)B2NP4F, (01)00884838061651(21)B28CZW, (01)00884838061651(21)B2LMY0, (01)00884838061798(21)B05KK0, (01)00884838061798(21)B0QV98, (01)00884838061651(21)B20X0M, (01)00884838061798(21)B0FNKH, (01)00884838061798(21)B0XBVP, (01)00884838061798(21)B18KGT, (01)00884838061651(21)B2RKVM, (01)00884838061651(21)B1P7PC, (01)00884838061651(21)B3050G, (01)00884838061460(21)B2LN62, (01)00884838061798(21)B20WHB, (01)00884838061798(21)B2LN38, (01)00884838061798(21)B2NPC2, (01)00884838061651(21)B1YQTC, (01)00884838061651(21)B2NP40, (01)00884838061651(21)F086MZ, (01)00884838061651(21)B1P7V9, (01)00884838061651(21)B2G0KT, (01)00884838061651(21)B1RYXT, (01)00884838061651(21)B20WX4, (01)00884838061798(21)B0X420, (01)00884838061651(21)B20WXH, (01)00884838061651(21)F02BB3, (01)00884838061798(21)B0QM19, (01)00884838061651(21)B28D31, (01)00884838061651(21)B1YR13, (01)00884838061651(21)B28D9V, (01)00884838061651(21)B2RVBQ, (01)00884838061651(21)B2G0MR, (01)00884838061651(21)B28DDV, (01)00884838061651(21)B35C9D, (01)00884838061651(21)B35CC0, (01)00884838061651(21)B20X0L, (01)00884838061651(21)B28DK6, (01)00884838061651(21)B28DPR, (01)00884838061651(21)B2G0Q4, (01)00884838061651(21)B2G0QD, (01)00884838061460(21)F08468, (01)00884838061651(21)B28DCH, (01)00884838061651(21)B28DCR, (01)00884838061651(21)B28DDL, (01)00884838061651(21)B28DDP, (01)00884838061460(21)B1K6T9, (01)00884838061798(21)B136ZN, (01)00884838061651(21)F068GM, (01)00884838061651(21)F06JZR, (01)00884838061651(21)B1P7XM, (01)00884838061651(21)B2G0LR, (01)00884838061460(21)F01QRX, (01)00884838061798(21)B35DW6, (01)00884838061651(21)B28D3Y, (01)00884838061651(21)B20WZX, (01)00884838061651(21)B2G0YZ, (01)00884838061798(21)B0D654, (01)00884838061798(21)B0T291, (01)00884838061798(21)B0QJ7R, (01)00884838061651(21)B2RV95, (01)00884838061651(21)B28DMX, (01)00884838061651(21)B28DN2, (01)00884838061651(21)B1YQQL, (01)00884838061651(21)B2G0Q7, (01)00884838061651(21)B2RL68, (01)00884838061798(21)B0QLMF, (01)00884838061651(21)B32MZQ, (01)00884838061460(21)F07XT6, (01)00884838061651(21)B1XLY0, (01)00884838061798(21)B20WJG, (01)00884838061651(21)B2LN3C, (01)00884838061651(21)F00XP4, (01)00884838061651(21)B28D6Z, (01)00884838061798(21)B0NL5Y, (01)00884838061651(21)B2RL3Q, (01)00884838061798(21)B35CBY, (01)00884838061651(21)B1XM0V, (01)00884838061651(21)B1VT1T, (01)00884838061651(21)B1XLYB, (01)00884838061651(21)B1YR03, (01)00884838061651(21)B304YZ, (01)00884838061651(21)B28DDC, (01)00884838061651(21)B304Y5, (01)00884838061651(21)B2G0VM, (01)00884838061651(21)B2LMXB, (01)00884838061651(21)B3L57H, (01)00884838061651(21)B1P7NP, (01)00884838061651(21)B2G0WK, (01)00884838061798(21)B28DP5, (01)00884838061651(21)B20X91, (01)00884838061651(21)B2RKW3, (01)00884838061651(21)B2RV7W, (01)00884838061651(21)B1F7PB, (01)00884838061651(21)B2LN4X, (01)00884838061651(21)B28DFQ, (01)00884838061651(21)B1VT24, (01)00884838061651(21)B1VT2F, (01)00884838061651(21)B1YQZB, (01)00884838061651(21)B1YR0H, (01)00884838061651(21)B28DQW, (01)00884838061651(21)B2G0X7, (01)00884838061651(21)B1RYT3, (01)00884838061798(21)B04HP6, (01)00884838061798(21)B0FMWM, (01)00884838061651(21)B1P7YB, (01)00884838061651(21)B20WQ6, (01)00884838061651(21)B20WQQ, (01)00884838061651(21)B35C30, (01)00884838061651(21)B28DQG, (01)00884838061651(21)B28DM9, (01)00884838061651(21)B2NP8Q, (01)00884838061651(21)B28DQR, (01)00884838061651(21)B28D5Q, (01)00884838061651(21)B1MB5G, (01)00884838061651(21)B3053P, (01)00884838061651(21)B28D8D, (01)00884838061651(21)B2RL7L, (01)00884838061651(21)B2RVHD, (01)00884838061651(21)B1RYWJ, (01)00884838061651(21)B2G0VW, (01)00884838061798(21)B304XV, (01)00884838061651(21)B2G0JR, (01)00884838061651(21)B2G0L0, (01)00884838061651(21)B2G0PC, (01)00884838061651(21)B1MB3Z, (01)00884838061651(21)B1VT1P, (01)00884838061651(21)B2LN0X, (01)00884838061651(21)B2LN2T, (01)00884838061651(21)B3058N, (01)00884838061798(21)B0TW4K, (01)00884838061651(21)B28D4X, (01)00884838061651(21)B1RYW1, (01)00884838061651(21)B1YR16, (01)00884838061651(21)B28DDX, (01)00884838061651(21)B2RL0G, (01)00884838061651(21)B32N1W, (01)00884838061651(21)B2RVGG, (01)00884838061651(21)B20X87, (01)00884838061651(21)B2NP4N, (01)00884838061651(21)B3056T, (01)00884838061651(21)B3056X, (01)00884838061651(21)B30575, (01)00884838061651(21)B32T1M, (01)00884838061651(21)B3057Y, (01)00884838061651(21)B20WWB, (01)00884838061651(21)F021PC, (01)00884838061651(21)B30595, (01)00884838061651(21)B32RZ1, (01)00884838061651(21)B3Q517, (01)00884838061651(21)F019DP, (01)00884838061460(21)F02KZL, (01)00884838061460(21)F02NL5, (01)00884838061651(21)B28D5V, (01)00884838061651(21)B28D6G, (01)00884838061651(21)B2RKZN, (01)00884838061651(21)B28DPW, (01)00884838061798(21)B35C03, (01)00884838061651(21)B2RVK2, (01)00884838061651(21)B304WM, (01)00884838061798(21)B1YR07, (01)00884838061651(21)B20WR5, (01)00884838061651(21)B13777, (01)00884838061651(21)B2RVBJ, (01)00884838061651(21)F07P4F, (01)00884838061651(21)B28DM7, (01)00884838061651(21)B28DMN, (01)00884838061651(21)B1XLY6, (01)00884838061651(21)B20WQ5, (01)00884838061651(21)F06K2F, (01)00884838061651(21)B2NP2J, (01)00884838061651(21)B2RVFG, (01)00884838061651(21)B2G0X2, (01)00884838061651(21)B20X09, (01)00884838061651(21)F087HC, (01)00884838061651(21)F00G0P, (01)00884838061651(21)B2G0DW, (01)00884838061651(21)B2NP56, (01)00884838061651(21)B28DFN, (01)00884838061651(21)B28D48, (01)00884838061651(21)B28D56, (01)00884838061651(21)B1VRXJ, (01)00884838061651(21)B28DNP, (01)00884838061651(21)B28DN0, (01)00884838061651(21)B2G0KV, (01)00884838061798(21)B35C2J, (01)00884838061651(21)B2LN0F, (01)00884838061651(21)B2RL9L, (01)00884838061798(21)B32RZ4, (01)00884838061651(21)B1XM2J, (01)00884838061651(21)B28DRK, (01)00884838061651(21)B28DB8, (01)00884838061798(21)B1YQQM, (01)00884838061798(21)B35CBZ, (01)00884838061651(21)F00M3W, (01)00884838061651(21)B1MB44, (01)00884838061651(21)B28DG5, (01)00884838061651(21)B2NP3H, (01)00884838061651(21)B2G0HP, (01)00884838061460(21)B1VL18, (01)00884838061460(21)B2NP8M, (01)00884838061651(21)B1RZ0G, (01)00884838061651(21)B3LFGL, (01)00884838061798(21)B2LN8L, (01)00884838061651(21)B2NP2K, (01)00884838061651(21)B2NP3Y, (01)00884838061651(21)B2LMZ1, (01)00884838061651(21)B1RYVC, (01)00884838061651(21)B30537, (01)00884838061651(21)F05VW0, (01)00884838061651(21)B28DB3, (01)00884838061651(21)B2NPBJ, (01)00884838061651(21)B2RL6G, (01)00884838061651(21)B28DBB, (01)00884838061651(21)B35HH6, (01)00884838061651(21)B28DPM, (01)00884838061798(21)B04R1N, (01)00884838061651(21)B1XM3Q, (01)00884838061651(21)B35CMF, (01)00884838061651(21)B2G0GH, (01)00884838061651(21)B1YQZJ, (01)00884838061651(21)B2G0LW, (01)00884838061651(21)B1P7XC, (01)00884838061651(21)B1P7XD, (01)00884838061651(21)B2RKWR, (01)00884838061651(21)B2RV71, (01)00884838061460(21)F05Q7X, (01)00884838061651(21)B2RV73, (01)00884838061651(21)B2NP2T, (01)00884838061651(21)B35H6B, (01)00884838061651(21)B1YR02, (01)00884838061651(21)B1YR0C, (01)00884838061651(21)B1YR0F, (01)00884838061651(21)B1YR0R, (01)00884838061460(21)B15MR1, (01)00884838061651(21)F08KF8, (01)00884838061651(21)F00J7M, (01)00884838061651(21)F00PCP, (01)00884838061651(21)F08PD2, (01)00884838061651(21)F08X0D, (01)00884838061651(21)B2LN0P, (01)00884838061651(21)B2G0MQ, (01)00884838061651(21)B28DCN, (01)00884838061651(21)B1YQZP, (01)00884838061651(21)B1VRXN, (01)00884838061651(21)B2RVCG, (01)00884838061798(21)B304XW, (01)00884838061651(21)B2RVGY, (01)00884838061651(21)B20WH6, (01)00884838061651(21)B2LMX0, (01)00884838061651(21)B30520, (01)00884838061651(21)B28D39, (01)00884838061651(21)B28D89, (01)00884838061651(21)B28D91, (01)00884838061651(21)B20WRP, (01)00884838061798(21)B0JH88, (01)00884838061798(21)B32MVN, (01)00884838061651(21)B1XLXK, (01)00884838061651(21)B1XLY5, (01)00884838061651(21)B2RVFN, (01)00884838061651(21)F08JPM, (01)00884838061651(21)F0BTXK, (01)00884838061651(21)B20X15, (01)00884838061651(21)B2G0L4, (01)00884838061651(21)B304XK, (01)00884838061651(21)B28DRQ, (01)00884838061651(21)B28DHH, (01)00884838061651(21)B1F8CN, (01)00884838061651(21)B20WQX, (01)00884838061651(21)B20WR6, (01)00884838061651(21)B20WRJ, (01)00884838061651(21)B1F7LN, (01)00884838061651(21)B2NP3P, (01)00884838061651(21)B2NP69, (01)00884838061651(21)B2RL5Y, (01)00884838061651(21)B1P7X4, (01)00884838061651(21)B28DL8, (01)00884838061651(21)B2G0MM, (01)00884838061651(21)B28DM1, (01)00884838061651(21)B28D8X, (01)00884838061651(21)B28DQ0, (01)00884838061651(21)B2G0WT, (01)00884838061651(21)F04Y2L, (01)00884838061651(21)F080RN, (01)00884838061651(21)B2LN6V, (01)00884838061651(21)F0BTXD, (01)00884838061651(21)F0BTXF, (01)00884838061798(21)B1YQZ6, (01)00884838061651(21)B1XLXX, (01)00884838061651(21)B3PYGT, (01)00884838061798(21)B18KK4, (01)00884838061651(21)B2RKYQ, (01)00884838061651(21)B2G0XF, (01)00884838061651(21)B1YQX6, (01)00884838061651(21)B2G0JF, (01)00884838061651(21)B304XT, (01)00884838061651(21)B1XM42, (01)00884838061651(21)B32LJG, (01)00884838061651(21)B2RV8X, (01)00884838061651(21)B1P7V3, (01)00884838061651(21)B3057G, (01)00884838061460(21)B1WFYC, (01)00884838061798(21)B0WHCQ, (01)00884838061651(21)B32LXF, (01)00884838061651(21)B1VT2C, (01)00884838061651(21)B28DNT, (01)00884838061651(21)B1VRZR, (01)00884838061651(21)B20WWL, (01)00884838061651(21)B1374H, (01)00884838061651(21)B1F7ZD, (01)00884838061651(21)B2NP43, (01)00884838061651(21)F055HJ, (01)00884838061651(21)B2G0JG, (01)00884838061651(21)B2G0JK, (01)00884838061798(21)B0BN7K, (01)00884838061798(21)B1F8D0, (01)00884838061651(21)B2G0MH, (01)00884838061651(21)B1P7XP, (01)00884838061651(21)F07CN2, (01)00884838061651(21)B35CNW, (01)00884838061651(21)B28D7J, (01)00884838061651(21)B1YQZ7, (01)00884838061651(21)B2G0W5, (01)00884838061651(21)B304XZ, (01)00884838061798(21)B03D0P, (01)00884838061460(21)B1XP1R, (01)00884838061460(21)B1Z4CT, (01)00884838061460(21)B1XDT0, (01)00884838061651(21)F01534, (01)00884838061651(21)F00J7H, (01)00884838061651(21)F06K25, (01)00884838061651(21)B1P7PK, (01)00884838061651(21)B2G0PB, (01)00884838061651(21)B1VRZ8, (01)00884838061651(21)B1VT08, (01)00884838061798(21)B0GQDP, (01)00884838061651(21)B1P7YF, (01)00884838061651(21)B20WKK, (01)00884838061651(21)B20WW1, (01)00884838061651(21)B20WW8, (01)00884838061651(21)B20WW9, (01)00884838061651(21)B20WWQ, (01)00884838061651(21)B20WX2, (01)00884838061651(21)B20WXL, (01)00884838061651(21)B20WXP, (01)00884838061651(21)B28D9L, (01)00884838061651(21)B2RL4N, (01)00884838061651(21)B20X55, (01)00884838061651(21)B20X5F, (01)00884838061651(21)B2G0NK, (01)00884838061651(21)B20X06, (01)00884838061651(21)B28D4L, (01)00884838061651(21)B28D5K, (01)00884838061651(21)B2RVF0, (01)00884838061651(21)F01F5X, (01)00884838061651(21)B28DRP, (01)00884838061798(21)B0NLR6, (01)00884838061651(21)B1RYRD, (01)00884838061798(21)B0J7QV, (01)00884838061651(21)B304WF, (01)00884838061651(21)B1P7Y4, (01)00884838061651(21)B20WWR, (01)00884838061798(21)B0GX8X, (01)00884838061798(21)B0FWRJ, (01)00884838061798(21)B18KBF, (01)00884838061798(21)B28DGG, (01)00884838061651(21)B2G0JH, (01)00884838061798(21)B04HP5, (01)00884838061651(21)B304YN, (01)00884838061651(21)B1F7ZP, (01)00884838061651(21)B1P7XW, (01)00884838061798(21)B0D2Z9, (01)00884838061651(21)B2RV9F, (01)00884838061651(21)B1F7VG, (01)00884838061651(21)B32MFF, (01)00884838061651(21)F06JZN, (01)00884838061651(21)B28D6F, (01)00884838061651(21)B28D7W, (01)00884838061651(21)B32N02, (01)00884838061651(21)B3PYC7, (01)00884838061651(21)B2G0J1, (01)00884838061651(21)B1RYT0, (01)00884838061798(21)B0WPY0, (01)00884838061460(21)B1K7HY, (01)00884838061651(21)B20X5G, (01)00884838061651(21)B3L5C3, (01)00884838061651(21)B1RZ0M, (01)00884838061651(21)B1P7Q5, (01)00884838061798(21)B0HGHT, (01)00884838061651(21)B2G0JC, (01)00884838061798(21)B0K7GG, (01)00884838061651(21)B1VRZZ, (01)00884838061651(21)B1VRYY, (01)00884838061651(21)B1MB6Q, (01)00884838061651(21)B2RVD2, (01)00884838061651(21)B2RVFB, (01)00884838061651(21)B1P7PF, (01)00884838061651(21)B1VT00, (01)00884838061651(21)B1YQXM, (01)00884838061651(21)B1YR08, (01)00884838061651(21)B1YR0K, (01)00884838061798(21)B2G0KC, (01)00884838061651(21)B28DRY, (01)00884838061651(21)B1P7XJ, (01)00884838061651(21)B2LN4K, (01)00884838061651(21)B2G0LN, (01)00884838061651(21)B20WJR, (01)00884838061798(21)B28DC5, (01)00884838061651(21)B20WWY, (01)00884838061651(21)B20WTG, (01)00884838061651(21)B20WTT, (01)00884838061651(21)F06JZQ, (01)00884838061651(21)F076ZV, (01)00884838061651(21)B0WPJC, (01)00884838061651(21)B1XLYY, (01)00884838061460(21)B1K53X, (01)00884838061651(21)B20WR3, (01)00884838061651(21)B1YQXN, (01)00884838061651(21)B1YQXZ, (01)00884838061651(21)B1YQYY, (01)00884838061651(21)B1YQZF, (01)00884838061651(21)B1YQZN, (01)00884838061651(21)B304YR, (01)00884838061651(21)B1XM2Y, (01)00884838061798(21)B32MZY, (01)00884838061651(21)B1RYYW, (01)00884838061651(21)B1BR63, (01)00884838061651(21)B20WRH, (01)00884838061651(21)B3L4KJ, (01)00884838061651(21)B2RKXN, (01)00884838061651(21)B2LMWT, (01)00884838061651(21)B20WQC, (01)00884838061651(21)B3LD9H, (01)00884838061651(21)F00G02, (01)00884838067424(21)B25MX8, (01)00884838061651(21)B1P7W9, (01)00884838061651(21)B20WNC, (01)00884838061651(21)B35HF3, (01)00884838061460(21)B1VKY5, (01)00884838061651(21)B2NP44, (01)00884838061651(21)B2NPBY, (01)00884838061651(21)F08G8M, (01)00884838061651(21)F0BM1F, (01)00884838061651(21)B1F87F, (01)00884838061651(21)B28DNK, (01)00884838061651(21)B28DF7, (01)00884838061798(21)B32N1T, (01)00884838061651(21)B3FGWB, (01)00884838061651(21)B2RL71, (01)00884838061651(21)B1YQRJ, (01)00884838061651(21)B1YQW1, (01)00884838061651(21)B2G0H4, (01)00884838061651(21)B2G0HD, (01)00884838061651(21)B2G0JT, (01)00884838061651(21)B2G0JY, (01)00884838061798(21)B35CP5, (01)00884838061651(21)B28D8F, (01)00884838061651(21)B30596, (01)00884838061651(21)B28D0N, (01)00884838061651(21)B2LN12, (01)00884838061651(21)B1YR0W, (01)00884838061651(21)B30503, (01)00884838061651(21)B28D44, (01)00884838061651(21)B2G0GJ, (01)00884838061651(21)B2NPB3, (01)00884838061651(21)B20WY8, (01)00884838061651(21)B2G0VR, (01)00884838061798(21)B28DC2, (01)00884838061651(21)B20WK7, (01)00884838061798(21)B1YQQJ, (01)00884838061651(21)B1YQWT, (01)00884838061651(21)B28DPL, (01)00884838061651(21)B2RL63, (01)00884838061651(21)B3HWZF, (01)00884838061460(21)B2NP8L, (01)00884838061651(21)F09LFR, (01)00884838061651(21)B1YQYH, (01)00884838061651(21)B28D6D, (01)00884838061651(21)B28D53, (01)00884838061651(21)B28D66, (01)00884838061651(21)B28D6J, (01)00884838061651(21)B28D6W, (01)00884838061651(21)B28D6X, (01)00884838061651(21)B28D70, (01)00884838061651(21)B28D9W, (01)00884838061651(21)B28DCP, (01)00884838061651(21)B28DCT, (01)00884838061651(21)B2G0L1, (01)00884838061651(21)B3PZHZ, (01)00884838061651(21)B1RYYV, (01)00884838061651(21)B2G0FK, (01)00884838061460(21)B2RVBB, (01)00884838061798(21)B03L5G, (01)00884838061651(21)B2G0TV, (01)00884838061651(21)B2G0TX, (01)00884838061651(21)B2G0VP, (01)00884838061651(21)B2G0WF, (01)00884838061651(21)B2RKY4, (01)00884838061651(21)B3055V, (01)00884838061651(21)F05MQ0, (01)00884838061651(21)B2G0JW, (01)00884838061651(21)B2G0J6, (01)00884838061651(21)B1VRX6, (01)00884838061651(21)B1VRX7, (01)00884838061651(21)B1RYRR, (01)00884838061651(21)B3L4KB, (01)00884838061651(21)B2NPBV, (01)00884838061651(21)B20WRY, (01)00884838061798(21)B05JX9, (01)00884838061798(21)B2LMWC, (01)00884838061651(21)B2LN0K, (01)00884838061798(21)B32N05, (01)00884838061798(21)B0JH86, (01)00884838061651(21)B1RYZK, (01)00884838061651(21)F03MCV, (01)00884838061798(21)B32LX9, (01)00884838061651(21)B20WJD, (01)00884838061651(21)B1MB3Q, (01)00884838061651(21)B0NLC0, (01)00884838061651(21)B20WK2, (01)00884838061651(21)B20WLG, (01)00884838061651(21)B20WM5, (01)00884838061651(21)B28D3T, (01)00884838061651(21)B1RZ07, (01)00884838061651(21)B2RVM2, (01)00884838061651(21)F019DQ, (01)00884838061651(21)B2RVDG, (01)00884838061651(21)B2G0PN, (01)00884838061651(21)B1P7WC, (01)00884838061651(21)B1XLYK, (01)00884838061651(21)B1XLYL, (01)00884838061651(21)B20X73, (01)00884838061651(21)B2G0ZB, (01)00884838061651(21)B2RVC7, (01)00884838061651(21)B1XM3H, (01)00884838061798(21)B0T44Q, (01)00884838061651(21)B3Q04Y, (01)00884838061651(21)B28DC9, (01)00884838061651(21)B1P7Q6, (01)00884838061651(21)B32T7J, (01)00884838061651(21)B2G0Q3, (01)00884838061798(21)B20WD6, (01)00884838061651(21)F0870N, (01)00884838061460(21)B1TVQK, (01)00884838061651(21)F00QPT, (01)00884838061651(21)B2G0DL, (01)00884838061651(21)B1P7YP, (01)00884838061460(21)B1QX1V, (01)00884838061651(21)B35CWF, (01)00884838061798(21)B28DBN, (01)00884838061651(21)B35PZ4, (01)00884838061651(21)B1YQX5, (01)00884838061651(21)B35CWC, (01)00884838061651(21)B1XLZM, (01)00884838061651(21)B1VRWN, (01)00884838061651(21)B1YQWP, (01)00884838061651(21)B20WX1, (01)00884838061651(21)B2G0XT, (01)00884838061651(21)B28DDZ, (01)00884838061651(21)B35CT9, (01)00884838061651(21)B28D1G, (01)00884838061651(21)F00J7P, (01)00884838061798(21)B2RVCX, (01)00884838061651(21)F02NZD, (01)00884838061798(21)B078FR, (01)00884838061651(21)B2G0C2, (01)00884838061651(21)B35D5V, (01)00884838061651(21)B35CMM, (01)00884838061651(21)B20WZP, (01)00884838061651(21)B2RL7C, (01)00884838061651(21)F05NM1, (01)00884838061651(21)B28D05, (01)00884838061651(21)B2RVC3, (01)00884838061651(21)B28D2Y, (01)00884838061651(21)F00XP6, (01)00884838061460(21)B1VK49, (01)00884838061460(21)B20XQV, (01)00884838061651(21)B1F862, (01)00884838061798(21)B04H9W, (01)00884838061651(21)B2LMZ2, (01)00884838061651(21)B1VT2G, (01)00884838061651(21)B1YQW6, (01)00884838061651(21)B0WHR9, (01)00884838061651(21)B1XLXL, (01)00884838061651(21)B1YR1M, (01)00884838061651(21)B28DN1, (01)00884838061651(21)B28DN9, (01)00884838061651(21)B2LMYJ, (01)00884838061651(21)B1YQX7, (01)00884838061651(21)F01WV4, (01)00884838061651(21)B2LN1Q, (01)00884838061651(21)B3057V, (01)00884838061651(21)B3HMFM, (01)00884838061651(21)B1RZ0B, (01)00884838061651(21)B1RZ0C, (01)00884838061651(21)B20WRD, (01)00884838061651(21)B20WV8, (01)00884838061798(21)B1XM15, (01)00884838061651(21)B18KG6, (01)00884838061651(21)B1F80D, (01)00884838061798(21)B1XLZC, (01)00884838061798(21)B28DBP, (01)00884838061651(21)B1YQRB, (01)00884838061651(21)B20WVK, (01)00884838061651(21)B2LMYX, (01)00884838061651(21)B1P7XB, (01)00884838061651(21)B2RVG8, (01)00884838061460(21)F05MPX, (01)00884838061651(21)B2RL9M, (01)00884838061651(21)B28DFL, (01)00884838061651(21)B20WZL, (01)00884838061651(21)B20X03, (01)00884838061651(21)B1YQY3, (01)00884838061651(21)B2G0TQ, (01)00884838061651(21)B304YH, (01)00884838061460(21)B117WR, (01)00884838061651(21)B20WJV, (01)00884838061460(21)B1RY82, (01)00884838061651(21)B20WW5, (01)00884838061651(21)B3FGWC, (01)00884838061651(21)B1RYR7, (01)00884838061651(21)B2NP6L, (01)00884838061651(21)B32LPL, (01)00884838061651(21)B2RL3F, (01)00884838061651(21)B20WKD, (01)00884838061651(21)B2G0LC, (01)00884838061651(21)B20X6N, (01)00884838061651(21)B28D3M, (01)00884838061651(21)B28DPQ, (01)00884838061651(21)B28DRJ, (01)00884838061651(21)B1VRXC, (01)00884838061651(21)B28DK8, (01)00884838061651(21)B28DLM, (01)00884838061798(21)B1YQYL, (01)00884838061651(21)B2NP4X, (01)00884838061651(21)B28D5J, (01)00884838061651(21)B2G0WG, (01)00884838061651(21)F068GK, (01)00884838061798(21)B1P7TR, (01)00884838061798(21)B3057P, (01)00884838061651(21)B1BR3Z, (01)00884838061651(21)B1RYRV, (01)00884838061651(21)B2LMWX, (01)00884838061651(21)F063X5, (01)00884838061651(21)B1VT15, (01)00884838061651(21)B304WK, (01)00884838061651(21)B28DF0, (01)00884838061651(21)F06WHX, (01)00884838061651(21)B304X8, (01)00884838061651(21)B28D6H, (01)00884838061651(21)B1VRY7, (01)00884838061460(21)B0M27M, (01)00884838061651(21)F02BB7, (01)00884838061651(21)B1VRZB, (01)00884838061651(21)B28DF2, (01)00884838061651(21)B3HMC2, (01)00884838061651(21)B16WWB, (01)00884838061460(21)B16ZBK, (01)00884838061651(21)B28D3N, (01)00884838061798(21)B1XM3M, (01)00884838061651(21)B1MB6J, (01)00884838061651(21)B1YQQ7, (01)00884838061651(21)B1P7Z0, (01)00884838061651(21)B1YQRH, (01)00884838061651(21)B1YQYF, (01)00884838061651(21)B1P7PR, (01)00884838061651(21)B1P7RJ, (01)00884838061651(21)F05QVW, (01)00884838061651(21)B1P7T5, (01)00884838061460(21)B1FZNP, (01)00884838061651(21)B1P7TP, (01)00884838061460(21)B1K53W, (01)00884838061651(21)B1P7Y5, (01)00884838061651(21)B28DRD, (01)00884838061651(21)B1P7YM, (01)00884838061651(21)B28D5M, (01)00884838061651(21)B1F8D4, (01)00884838061651(21)B20WRG, (01)00884838061651(21)B2RL16, (01)00884838061798(21)B35D0J, (01)00884838061651(21)B1RYRP, (01)00884838061651(21)B28DNM, (01)00884838061798(21)B304X6, (01)00884838061651(21)B1VT1F, (01)00884838061460(21)B1QTWW, (01)00884838061651(21)B2G0P6, (01)00884838061651(21)B2LN1X, (01)00884838061651(21)B2RKWN, (01)00884838061651(21)B2RKX3, (01)00884838061651(21)B1YQQF, (01)00884838061651(21)B1YQT4, (01)00884838061651(21)B1XLXM, (01)00884838061651(21)B1RYTD, (01)00884838061460(21)B1ZRN7, (01)00884838061651(21)B1YR10, (01)00884838061651(21)B1YR0L, (01)00884838061798(21)B1VT1X, (01)00884838061798(21)B1YQYQ, (01)00884838061651(21)B1YQY1, (01)00884838061651(21)B1P7RQ, (01)00884838061651(21)B20WK0, (01)00884838061651(21)B1XLYV, (01)00884838061651(21)B1XLYZ, (01)00884838061651(21)B1XLZ7, (01)00884838061651(21)B1XM03, (01)00884838061651(21)B1XM07, (01)00884838061651(21)B2RVBR, (01)00884838061651(21)B2RVBV, (01)00884838061651(21)B28DT6, (01)00884838061651(21)B20WL1, (01)00884838061651(21)B20WJN, (01)00884838061651(21)B1YQRD, (01)00884838061798(21)B2G0XX, (01)00884838061651(21)B20WT9, (01)00884838061651(21)B20X14, (01)00884838061651(21)B20WRR, (01)00884838061651(21)B20WTQ, (01)00884838061651(21)B2G0D0, (01)00884838061651(21)B20X0D, (01)00884838061460(21)F05XHG, (01)00884838061651(21)B1VRX5, (01)00884838061651(21)B20WXF, (01)00884838061651(21)B20X7L, (01)00884838061651(21)B20X18, (01)00884838061651(21)B20X0W, (01)00884838061651(21)B28D7N, (01)00884838061651(21)B1XLZ2, (01)00884838061651(21)B20X04, (01)00884838061651(21)B28D38, (01)00884838061651(21)B28D3K, (01)00884838061651(21)B20X70, (01)00884838061651(21)B1VRZF, (01)00884838061651(21)B1P7PX, (01)00884838061651(21)B2RVFP, (01)00884838061651(21)B28D2Z, (01)00884838061651(21)B28D6C, (01)00884838061651(21)B28DCF, (01)00884838061651(21)F0153C, (01)00884838061651(21)B28D45, (01)00884838061651(21)B28DF1, (01)00884838061651(21)B28DFF, (01)00884838061651(21)B28DG3, (01)00884838061651(21)B28DD4, (01)00884838061460(21)B1GQJ3, (01)00884838061651(21)B28DMK, (01)00884838061651(21)B28DHM, (01)00884838061651(21)B28DLD, (01)00884838061651(21)B28DFZ, (01)00884838061651(21)B28DN3, (01)00884838061651(21)B1VRZP, (01)00884838061651(21)B1VT0L, (01)00884838061651(21)B2G0P9, (01)00884838061651(21)B2G0Y6, (01)00884838061460(21)B2RV6L, (01)00884838061460(21)F016BB, (01)00884838061651(21)B1VT02, (01)00884838061651(21)B2G0M4, (01)00884838061651(21)B2G0M3, (01)00884838061460(21)B1ZQNT, (01)00884838061460(21)F03BFT, (01)00884838061651(21)B2LN8W, (01)00884838061651(21)B2NP3X, (01)00884838061651(21)B2LMX8, (01)00884838061651(21)B2RVCZ, (01)00884838061651(21)B2G0L9, (01)00884838061651(21)B2LMZG, (01)00884838061651(21)F09C1Z, (01)00884838061651(21)B30517, (01)00884838061651(21)B2LN1D, (01)00884838061651(21)B304YC, (01)00884838061798(21)B2G0F4, (01)00884838061651(21)B2G0MT, (01)00884838061651(21)B2NP3W, (01)00884838061651(21)B2LN3Z, (01)00884838061460(21)B2ZQ92, (01)00884838061651(21)F05XHP, (01)00884838061651(21)B2NP3N, (01)00884838061651(21)B2LN27, (01)00884838061651(21)B28DG8, (01)00884838061651(21)B2NP68, (01)00884838061651(21)B2NP45, (01)00884838061651(21)B1MB5K, (01)00884838061651(21)B2LN94, (01)00884838061651(21)B2RV7Q, (01)00884838061651(21)B2RV8Y, (01)00884838061651(21)B1P7QQ, (01)00884838061651(21)B1P7QX, (01)00884838061460(21)B2G0RP, (01)00884838061651(21)B2RL6P, (01)00884838061651(21)F08V30, (01)00884838061798(21)B2LN3B, (01)00884838061651(21)B1RYRC, (01)00884838061651(21)B2RL59, (01)00884838061651(21)B1YQQB, (01)00884838061651(21)B2RVJM, (01)00884838061651(21)B2RVK9, (01)00884838061651(21)B2RVLX, (01)00884838061651(21)B2RL27, (01)00884838061651(21)B28CZX, (01)00884838061651(21)F093NN, (01)00884838061651(21)B2LN97, (01)00884838061651(21)B2RVDW, (01)00884838061651(21)B2RVFJ, (01)00884838061651(21)B2RV9P, (01)00884838061651(21)B2RVCR, (01)00884838061651(21)B2RVFY, (01)00884838061651(21)B2RVGK, (01)00884838061651(21)B28D49, (01)00884838061651(21)B28D7R, (01)00884838061651(21)B18K90, (01)00884838061651(21)B16WY3, (01)00884838061651(21)B304YL, (01)00884838061651(21)B304Y1, (01)00884838061651(21)B2RVGR, (01)00884838061651(21)B2RVHG, (01)00884838061798(21)B3052G, (01)00884838061651(21)B304XY, (01)00884838061651(21)B304Y0, (01)00884838061651(21)B32LK9, (01)00884838061651(21)B3052R, (01)00884838061651(21)B2LN2G, (01)00884838061651(21)B3053L, (01)00884838061651(21)B20WRN, (01)00884838061651(21)B20WT8, (01)00884838061651(21)B3051J, (01)00884838061651(21)B3058M, (01)00884838061651(21)B2G0WJ, (01)00884838061651(21)B30597, (01)00884838061651(21)F05VJB, (01)00884838061651(21)B3057H, (01)00884838061651(21)F00QPV, (01)00884838061460(21)B2RL08, (01)00884838061651(21)B28D41, (01)00884838061651(21)F096JH, (01)00884838061460(21)B2NP66, (01)00884838061651(21)F03GW0, (01)00884838061651(21)F03GW4, (01)00884838061651(21)F03GW5, (01)00884838061460(21)B2NP4B, (01)00884838061460(21)B2RVC5, (01)00884838061651(21)F00J81, (01)00884838061651(21)F00J8F, (01)00884838061651(21)F05Q85, (01)00884838061651(21)F063X9, (01)00884838061651(21)F0846C, (01)00884838061651(21)B1RYTY, (01)00884838061651(21)B2RV9T, (01)00884838061460(21)B2RKWC, (01)00884838061460(21)B2RL0C, (01)00884838061460(21)B2G0NV, (01)00884838061460(21)B2G0PT, (01)00884838061651(21)B35CQV, (01)00884838061651(21)B35C45, (01)00884838061460(21)B2RKYF, (01)00884838061651(21)B1YQZM, (01)00884838061651(21)F04F2T, (01)00884838061460(21)B2RVD7, (01)00884838061651(21)B39LX6, (01)00884838061651(21)B3PYH3, (01)00884838061651(21)B28DR9, (01)00884838061651(21)B2RKVR, (01)00884838061651(21)B28DNZ, (01)00884838061460(21)B2LMXL, (01)00884838061651(21)B3FHD3, (01)00884838061651(21)B2G0WY, (01)00884838061651(21)B1P7XY, (01)00884838061651(21)B2RKYY, (01)00884838061460(21)B2ZQ90, (01)00884838061651(21)B1P7TV, (01)00884838061651(21)B1BR8R, (01)00884838061651(21)B28D9H, (01)00884838061651(21)B2RKXC, (01)00884838061460(21)B3L458, (01)00884838061651(21)B1F7Z9, (01)00884838061651(21)F00J7R, (01)00884838061651(21)B1RYX2, (01)00884838061651(21)B28DG9, (01)00884838061651(21)B20WY2, (01)00884838061460(21)B3L598, (01)00884838061651(21)F03K5J, (01)00884838061651(21)F03LBW, (01)00884838061651(21)F00QPK, (01)00884838061651(21)B28CZN, (01)00884838061460(21)B3PYC1, (01)00884838061460(21)B3LDCP, (01)00884838061651(21)B2LN11, (01)00884838061651(21)B2G0VY, (01)00884838061798(21)B32T7M, (01)00884838061651(21)F08R8T, (01)00884838061651(21)B2G0NG, (01)00884838061460(21)F0149G, (01)00884838061651(21)B2NP97, (01)00884838061460(21)F012WG, (01)00884838061460(21)F012WH, (01)00884838061651(21)F0846F, (01)00884838061651(21)F03LBQ, (01)00884838061651(21)B304XF, (01)00884838061460(21)B3LDCF, (01)00884838061460(21)F012W7, (01)00884838061460(21)F012VM, (01)00884838061460(21)F012WB, (01)00884838061651(21)B1YQQ5, (01)00884838061460(21)F02CN0, (01)00884838061651(21)F05VJC, (01)00884838061460(21)F02MX1, (01)00884838061651(21)F069DF, (01)00884838061460(21)F05N82, (01)00884838061460(21)F02MHG, (01)00884838061651(21)F063XL, (01)00884838061651(21)B2NP2G, (01)00884838061460(21)F0393T, (01)00884838061651(21)F068GJ, (01)00884838061651(21)F05QVY, (01)00884838061651(21)F0881R, (01)00884838061460(21)F03BFQ, (01)00884838061460(21)F01N4B, (01)00884838061651(21)B28D4R, (01)00884838061651(21)F0881N, (01)00884838061651(21)F06K2B, (01)00884838061651(21)F07DJF, (01)00884838061651(21)F02YBQ, (01)00884838061651(21)F01PRJ, (01)00884838061651(21)F07J8Q, (01)00884838061460(21)F02NL3, (01)00884838061460(21)F02KZM, (01)00884838061460(21)F05XHD, (01)00884838061460(21)F05M62, (01)00884838061651(21)F06K2H, (01)00884838061651(21)F06TT5, (01)00884838061651(21)F07Y4W, (01)00884838061651(21)F07DJD, (01)00884838061651(21)B1YQXX, (01)00884838061460(21)F05NJQ, (01)00884838061798(21)B0QHY9, (01)00884838061651(21)F06K23, (01)00884838061651(21)B0WHRB, (01)00884838061651(21)F077WP, (01)00884838061651(21)B20WWN, (01)00884838061651(21)B28DNJ, (01)00884838061460(21)B20Y21, (01)00884838061651(21)F07DJC, (01)00884838061651(21)B1P7W6, (01)00884838061651(21)B1YQWF, (01)00884838061651(21)F07CN4, (01)00884838061651(21)F0BZ00, (01)00884838061651(21)F08ZRD, (01)00884838061651(21)F07J8N, (01)00884838061651(21)F02YWK, (01)00884838061651(21)B2RVDM, (01)00884838061651(21)F032WZ, (01)00884838061651(21)F02WJ6, (01)00884838061651(21)F09R7L, (01)00884838061651(21)F09RYY, (01)00884838061651(21)F09VHD, (01)00884838061651(21)B1P7XT, (01)00884838061651(21)B1P7VL, (01)00884838061651(21)B1F84G, (01)00884838061651(21)F09LFP, (01)00884838061651(21)B32MDF, (01)00884838061651(21)B1P7Q0, (01)00884838061651(21)B20X20, (01)00884838061651(21)B20X2J, (01)00884838061651(21)B20X2T, (01)00884838061651(21)B20X3B, (01)00884838061651(21)B20X3D, (01)00884838061651(21)B20X3V, (01)00884838061651(21)B20X4L, (01)00884838061651(21)B1F8BF, (01)00884838061651(21)B1MB1Y, (01)00884838061651(21)B20WDH, (01)00884838061651(21)B3058Z, (01)00884838061651(21)B35P0N, (01)00884838061651(21)F0816G, (01)00884838061651(21)B32RZ6, (01)00884838061651(21)B304Y6, 010088483806165121F01ZGB, (01)00884838061651(21)B2RV9R, (01)00884838061651(21)B35D94, (01)00884838061651(21)B2NP9T, (01)00884838061651(21)F03RP9, (01)00884838061651(21)B35CMQ, (01)00884838061651(21)B2RV7V, (01)00884838061651(21)B35J5Q, (01)00884838061651(21)B39MCT, (01)00884838061651(21)B2NP80, (01)00884838061651(21)B2RL78; Serial No. B1VJQF, B2RL4B, B20WVG, B28DGB, B304XN, B35CH3, B1F81Y, B1F8CJ, B20WW0, B1YQQ6, 034GRX, B1XLYC, B1YQZR, B1P7QY, B2G0CC, B2LN7C, B2RL6L, B1P7TG, 0397R3, B04QBF, B0D043, B2RL9K, 02ZBPR, B1P7Q8, B20WM1, B20WKR, B1P7PJ, B1F805, B1XLYW, B1XLZ0, B2NP4F, B28CZW, B03DLC, B2LMY0, B05KK0, B0QV98, B18KG3, B1F89N, 034B85, B20X0M, B0FNKH, B0XBVP, B18KGT, B1F7YC, B2RKVM, B1F896, B1P7PC, B13763, B3050G, B2LN62, B20WHB, B2LN38, B2NPC2, 03BYGX, 030YGM, B1F82Q, B1YQTC, B2NP40, F086MZ, B1P7V9, B2G0KT, B1RYXT, 0323VN, 03363L, B20WX4, B0X420, B20WXH, B1F83Z, F02BB3, B0QM19, B28D31, B1YR13, B28D9V, B2RVBQ, B1F7ZC, B2G0MR, B28DDV, B35C9D, B35CC0, B20X0L, B28DK6, B28DPR, B2G0Q4, B2G0QD, F08468, B28DCH, B28DCR, B28DDL, B28DDP, B1K6T9, B136ZN, 03939D, 033F0F, F068GM, F06JZR, B1P7XM, B2G0LR, F01QRX, 035J0G, 0320PD, B35DW6, B28D3Y, B20WZX, B1F7L8, B2G0YZ, 036ZR7, B0D654, B0T291, B0QJ7R, B2RV95, B28DMX, B28DN2, B1YQQL, B2G0Q7, B2RL68, B0QLMF, B32MZQ, B1F7XJ, F07XT6, B1XLY0, B20WJG, B2LN3C, F00XP4, B1F7PV, B28D6Z, B1F81L, B0NL5Y, B2RL3Q, B35CBY, B1XM0V, 02WB9Z, B1VT1T, B1XLYB, B1YR03, B304YZ, B28DDC, B304Y5, B2G0VM, B2LMXB, B3L57H, B1BRBV, B1F85T, B1P7NP, B2G0WK, B28DP5, B20X91, B2RKW3, B2RV7W, B1F7PB, B2LN4X, B28DFQ, B1VT24, B1VT2F, B1YQZB, B1YR0H, B28DQW, B2G0X7, 03BN84, B1RYT3, 036M1T, 038ZGT, B0GQGW, 03H2ZL, B04HP6, B0FMWM, 02Z87V, 038WHF, 036GN3, B1F7WM, B1P7YB, B20WQ6, B20WQQ, B35C30, B28DQG, B1F7W3, 0314RV, B28DM9, B2NP8Q, B28DQR, B28D5Q, B1MB5G, B3053P, B28D8D, B2RL7L, B2RVHD, B1RYWJ, B2G0VW, B304XV, B2G0JR, B2G0L0, B2G0PC, B1MB3Z, B1VT1P, B2LN0X, B2LN2T, B3058N, B0TW4K, B28D4X, B1RYW1, B1YR16, B1P7NY, B1F83P, B28DDX, B2RL0G, B32N1W, B16WYW, B2RVGG, B20X87, B2NP4N, B3056T, B3056X, B30575, B32T1M, B3057Y, B1F85B, B20WWB, F021PC, B30595, B32RZ1, B3Q517, F019DP, F02KZL, F02NL5, B28D5V, B28D6G, B2RKZN, B28DPW, B35C03, B2RVK2, B304WM, B1YR07, B1F817, B20WR5, B13777, B2RVBJ, F07P4F, B28DM7, B28DMN, B18KG4, B1XLY6, B1F874, B1F7MT, B20WQ5, F06K2F, B2NP2J, B2RVFG, B2G0X2, B20X09, F087HC, F00G0P, 02VLJQ, B2G0DW, B1F7RT, B1BR7Y, B1F7QM, B2NP56, B28DFN, 032C6N, B28D48, B28D56, B1VRXJ, B28DNP, B28DN0, B2G0KV, B35C2J, 03GKH0, 02Y0RL, B2LN0F, B2RL9L, B32RZ4, 036LZZ, B1XM2J, B1F7X0, B28DRK, B1F7L3, B28DB8, B1YQQM, B1F812, B35CBZ, B1F884, B1F7K6, B1F7ZN, B1P7R9, F00M3W, B1MB44, B28DG5, B2NP3H, B1F7R0, B2G0HP, B1F7KB, B1VL18, B2NP8M, B1RZ0G, B3LFGL, B2LN8L, B2NP2K, B2NP3Y, B2LMZ1, B1RYVC, B30537, B18K9J, F05VW0, B28DB3, 03BMXM, B2NPBJ, B2RL6G, B28DBB, B09R6J, B35HH6, B28DPM, B04R1N, B1XM3Q, B35CMF, 02TX9Z, B1F88C, B1F88W, B1F87N, B0WNR1, B2G0GH, B1YQZJ, 034GH8, B2G0LW, B1P7XC, B1P7XD, B2RKWR, B2RV71, F05Q7X, B2RV73, B2NP2T, B35H6B, B1YR02, B1YR0C, B1YR0F, B1YR0R, B1BRCZ, B1F7RV, B15MR1, B1ZRN4, B1CYVT, 038RG5, F08KF8, B1F7P0, F00J7M, F00PCP, F08PD2, F08X0D, B1BR9Y, B2LN0P, B2G0MQ, B28DCN, B1YQZP, B1VRXN, B2RVCG, B304XW, B1F83G, B2RVGY, B20WH6, B2LMX0, B30520, B28DN4, B2RVB5, B28D39, B28D89, B28D91, B20WRP, B0JH88, B32MVN, B1XLXK, B1XLY5, B2RVFN, F08JPM, F0BTXK, B1F7Q8, B20X15, B1F7MN, B2G0L4, B304XK, 038X19, B28DRQ, B28DHH, B1F8CN, B20WQX, B20WR6, B20WRJ, B1F858, B1F7Q0, B1F7LN, B2NP3P, B2NP69, B2RL5Y, B1P7X4, B18K8Y, B28DL8, B2G0MM, B28DM1, B28D8X, B28DQ0, B2G0WT, F04Y2L, 03GCP8, F080RN, B2LN6V, F0BTXD, F0BTXF, B1YQZ6, 02VFCK, B1BRCH, B1XLXX, B3PYGT, B1F7X7, B18KK4, B2RKYQ, B2G0XF, B1YQX6, B2G0JF, 038RCT, B304XT, B1XM42, 030DB8, B32LJG, B2RV8X, B1P7V3, B3057G, B1F7Q7, B1MB3W, B1WFYC, 033GC6, B0WHCQ, B32LXF, B1VT2C, B28DNT, B1VRZR, B20WWL, B1374H, B1F7ZD, B2NP43, F055HJ, 0326LQ, B2G0JG, B2G0JK, B0BN7K, B1F8D0, B2G0MH, B1P7XP, B1F80V, B1F7R5, F07CN2, B35CNW, B28D7J, B1YQZ7, B0K0GV, B2G0W5, B1F7WG, B1F7P8, B304XZ, B03D0P, B1XP1R, B1Z4CT, B1XDT0, F01534, F00J7H, F06K25, B1F7Y2, B1P7PK, B2G0PB, B107QD, 035BWT, B1VRZ8, B1VT08, B0GQDP, B1P7YF, B20WKK, B20WW1, B20WW8, B20WW9, B20WWQ, B20WX2, B20WXL, B20WXP, B28D9L, B2RL4N, B1MB6Z, B0QVHN, B20X55, B20X5F, B2G0NK, B20X06, B1F87M, B28D4L, B28D5K, B2RVF0, F01F5X, 03FVXV, B28DRP, B1F84J, B0NLR6, 033K44, B1RYRD, 036J82, B0J7QV, B1F88F, B304WF, B1BRCP, B1F8DM, B1P7Y4, B20WWR, B0GX8X, 031RB7, B0FWRJ, B18KBF, B08VNW, B28DGG, 034B9J, B2G0JH, B04HP5, B304YN, B1F7ZP, B1P7XW, B0D2Z9, 02WQD4, B2RV9F, 0366D7, B1F7VG, B32MFF, F06JZN, B1MB32, B1F857, B28D6F, B28D7W, B1MB7D, B32N02, B3PYC7, B1MB4M, 03D9J2, B2G0J1, B1F82M, B1RYT0, B0WPY0, B1K7HY, B20X5G, B3L5C3, B1RZ0M, B1P7Q5, B0HGHT, 02XZZ9, B0C29B, B1BRC7, B1BR93, B1BRC3, B1F846, B2G0JC, B0K7GG, B0ZNJN, B1VRZZ, B1VRYY, B1MB41, B1MB6Q, B1F854, B2RVD2, B2RVFB, 0330RV, B1P7PF, B1VT00, B1YQXM, B1YR08, B1YR0K, B2G0KC, B1F7LL, B2LN57, B28DRY, B1P7XJ, B2LN4K, B2G0LN, B20WJR, B0D045, B28DC5, B1MB40, 03DDXL, B20WWY, 03729W, B20WTG, B20WTT, F06JZQ, F076ZV, B0WPJC, B1XLYY, B1K53X, B20WR3, B1YQXN, B1YQXZ, B1YQYY, B1YQZF, B1YQZN, B304YR, B1XM2Y, 031NPF, B32MZY, 03GDL2, B1RYYW, B1F891, B1BR63, B20WRH, B3L4KJ, B1F84W, B2RKXN, 02ZFM9, B2LMWT, B1F8BJ, B1F7VJ, B20WQC, B3LD9H, F00G02, B1XNTX, B1P7W9, B20WNC, B35HF3, B1VKY5, B2NP44, B2NPBY, F08G8M, F0BM1F, 02Z973, B1F87F, B28DNK, 02ZDX1, 02Y5J3, 02WYYP, B28DF7, B32N1T, B3FGWB, B2RL71, B1YQRJ, B1YQW1, B2G0H4, B2G0HD, B2G0JT, B2G0JY, 033PPQ, 03H43C, B35CP5, B28D8F, B30596, B28D0N, B2LN12, B1YR0W, B30503, B28D44, B2G0GJ, B1F7VP, B2NPB3, B20WY8, B2G0VR, B28DC2, B078FW, B20WK7, 0397WY, B1YQQJ, B1BR8B, B1YQWT, B28DPL, B2RL63, B3HWZF, B2NP8L, F09LFR, B1YQYH, B28D6D, B28D53, B28D66, B28D6J, B28D6W, B28D6X, B28D70, B28D9W, B28DCP, B28DCT, B2G0L1, B3PZHZ, B1RYYV, B2G0FK, B1BRB0, B2RVBB, B03L5G, B2G0TV, B2G0TX, B2G0VP, B2G0WF, B2RKY4, B3055V, F05MQ0, B2G0JW, B1MB4X, B1MB6B, B1F8C3, B1F8CY, B2G0J6, B1VRX6, B1VRX7, B1RYRR, B3L4KB, B1BR41, B2NPBV, B20WRY, 02ZF2C, B05JX9, B2LMWC, 03BM4Q, B2LN0K, B32N05, B0JH86, B1RYZK, B1MB3V, B18KB9, F03MCV, B1F7LW, B32LX9, B20WJD, 02WQNP, B1F7W9, B1MB3Q, B0NLC0, B20WK2, B20WLG, B20WM5, B28D3T, B1F7VH, B1RZ07, B2RVM2, F019DQ, B2RVDG, B2G0PN, B1MB6W, B1P7WC, B1XLYK, B1XLYL, B20X73, B2G0ZB, B2RVC7, B1F83N, B1F851, B1XM3H, B1F84H, B0T44Q, B3Q04Y, B28DC9, B1P7Q6, B32T7J, B1F83W, B1F7R6, B2G0Q3, B1F801, B20WD6, F0870N, B1TVQK, F00QPT, B2G0DL, 02WV66, F09MJ1, B1P7YP, B1QX1V, B35CWF, B28DBN, B35PZ4, B1YQX5, B35CWC, B1XLZM, B1VRWN, B1YQWP, B20WX1, 02WT34, B2G0XT, B28DDZ, B1F84X, B35CT9, B1F7YL, B28D1G, F00J7P, B2RVCX, B1MB68, F02NZD, B078FR, B2G0C2, 03BM4J, B35D5V, B35CMM, B20WZP, B1MB7C, B2RL7C, F05NM1, B28D05, B2RVC3, B1F7X2, B28D2Y, F00XP6, B1VK49, B20XQV, B1F862, B04H9W, 036DJ8, B2LMZ2, B1VT2G, B1YQW6, B0WHR9, B1XLXL, B1YR1M, B28DN1, B28DN9, B2LMYJ, B1YQX7, F01WV4, B2LN1Q, B3057V, B3HMFM, B1RZ0B, B1RZ0C, 03FP81, B1F7XQ, B20WRD, B20WV8, B02PWL, B1XM15, B18KG6, B1F80D, B1XLZC, B28DBP, 03729T, B1YQRB, B20WVK, B2LMYX, B1P7XB, B0LGG1, B2RVG8, F05MPX, B2RL9M, B28DFL, B20WZL, B20X03, B1BR8V, B1BR82, B1YQY3, B1F7WL, B2G0TQ, B304YH, 0393FY, B117WR, B20WJV, B1RY82, B20WW5, B1F7R4, B3FGWC, 02XYTR, 035H7T, B1F7LC, B1RYR7, B2NP6L, B32LPL, B10PVY, B2RL3F, B20WKD, B2G0LC, B20X6N, B1MB3P, B28D3M, B18KGL, B28DPQ, B1F7Z7, B28DRJ, B1F8BL, B0K7D4, B1VRXC, B1F8C1, B28DK8, B28DLM, B1YQYL, B2NP4X, B28D5J, B2G0WG, F068GK, B1P7TR, B3057P, B1BR3Z, B1RYRV, B1BR81, B2LMWX, B1F7JM, B1F4T1, F063X5, B1F85Q, B1F85K, B1VT15, B1F7LZ, B1F7KC, B304WK, B1MB69, B1F7LG, B1F830, B28DF0, B1F7PH, F06WHX, B1F7PN, B1F7NW, B304X8, B1F7ZK, B28D6H, B1VRY7, B0M27M, F02BB7, B1F85M, B1VRZB, B1F7KD, B28DF2, B3HMC2, B1F879, 03BXZW, B16WWB, B1F86F, B16ZBK, B1F88D, B1F8BH, B1F87G, B28D3N, B1F81P, B1F8B9, B1XM3M, B1F7X3, B1F8BW, B1DXH8, B1MB3X, B1MB6G, B1F4T2, B1MB6J, B1YQQ7, 037PTB, B1P7Z0, B1YQRH, B1YQYF, B0JJ3M, B1P7PR, B1P7RJ, F05QVW, B1P7T5, B1FZNP, B1P7TP, B1K53W, B1P7Y5, B28DRD, B1P7YM, B28D5M, B1F8D4, B20WRG, B2RL16, B35D0J, B1RYRP, B28DNM, B304X6, B1VT1F, B1QTWW, B2G0P6, B2LN1X, B2RKWN, B2RKX3, B1YQQF, B1YQT4, 036W4B, B1XLXM, B1RYTD, B1ZRN7, B1YR10, B1YR0L, B1VT1X, B1YQYQ, B1YQY1, B1P7RQ, B20WK0, B1XLYV, B1XLYZ, B1XLZ7, B1XM03, B1XM07, B2RVBR, B2RVBV, B28DT6, B20WL1, B20WJN, B1YQRD, B2G0XX, B20WT9, B20X14, B20WRR, B20WTQ, B2G0D0, B20X0D, F05XHG, B1VRX5, B20WXF, B20X7L, B20X18, B20X0W, B28D7N, B1XLZ2, B20X04, B28D38, B28D3K, B20X70, B1VRZF, B1P7PX, B2RVFP, B28D2Z, B28D6C, B28DCF, F0153C, B28D45, B28DF1, B28DFF, B28DG3, B28DD4, B1GQJ3, B28DMK, B28DHM, B28DLD, B28DFZ, B28DN3, B1VRZP, B1VT0L, B2G0P9, B2G0Y6, B2RV6L, F016BB, B1VT02, B2G0M4, B2G0M3, B1MB7T, B1ZQNT, F03BFT, B2LN8W, B2NP3X, B2LMX8, B1F88L, B2RVCZ, B2G0L9, B0Q5D4, B2LMZG, F09C1Z, B30517, B2LN1D, B304YC, B2G0F4, B2G0MT, B2NP3W, B2LN3Z, B2ZQ92, F05XHP, B2NP3N, B2LN27, B28DG8, B2NP68, B2NP45, B1MB5K, B2LN94, B2RV7Q, B2RV8Y, B1P7QQ, B1P7QX, B2G0RP, B2RL6P, F08V30, B2LN3B, B1RYRC, B2RL59, B1YQQB, B2RVJM, B2RVK9, B2RVLX, B1MB7N, B2RL27, B28CZX, F093NN, B2LN97, B2RVDW, B2RVFJ, B2RV9P, B2RVCR, B2RVFY, B2RVGK, B28D49, B28D7R, B18K90, B16WY3, B304YL, B304Y1, B2RVGR, B2RVHG, B3052G, B304XY, B304Y0, B32LK9, B3052R, B2LN2G, B3053L, B20WRN, B20WT8, B3051J, B3058M, B2G0WJ, B30597, F05VJB, B1F7PG, B3057H, F00QPV, B2RL08, B28D41, F096JH, B2NP66, F03GW0, F03GW4, F03GW5, B2NP4B, B2RVC5, F00J81, F00J8F, F05Q85, F063X9, F0846C, B1RYTY, B2RV9T, B2RKWC, B2RL0C, B1VPBM, B2G0NV, B2G0PT, B35CQV, B35C45, B2RKYF, B1YQZM, F04F2T, B2RVD7, B39LX6, B3PYH3, B28DR9, B2RKVR, B28DNZ, B2LMXL, B3FHD3, B1MB6C, B2G0WY, B1P7XY, B2RKYY, B2ZQ90, B1P7TV, B1BR8R, B28D9H, B2RKXC, B3L458, B1F7Z9, F00J7R, B1RYX2, B28DG9, B20WY2, B3L598, F03K5J, F03LBW, F00QPK, B28CZN, B3PYC1, B3LDCP, B2LN11, B1F81W, B2G0VY, B32T7M, 03H39R, F08R8T, B2G0NG, F0149G, B2NP97, F012WG, F012WH, F0846F, F03LBQ, B304XF, B3LDCF, F012W7, F012VM, F012WB, B1YQQ5, F02CN0, F05VJC, 036JB8, F02MX1, F069DF, F05N82, F02MHG, F063XL, B2NP2G, F0393T, F068GJ, F05QVY, F0881R, F03BFQ, F01N4B, B28D4R, F0881N, F06K2B, F07DJF, F02YBQ, F01PRJ, F07J8Q, B0T4GV, F02NL3, F02KZM, 0316RY, F05XHD, F05M62, F06K2H, F06TT5, F07Y4W, F07DJD, B1YQXX, F05NJQ, 0335K3, B0QHY9, 0397TQ, F06K23, B0WHRB, F077WP, B1F843, B20WWN, B28DNJ, B20Y21, 037WFX, F07DJC, B1P7W6, B1YQWF, F07CN4, F0BZ00, F08ZRD, F07J8N, F02YWK, B2RVDM, F032WZ, F02WJ6, F09R7L, F09RYY, F09R7H, F09VHD, B1P7XT, B1P7VL, B1F84G, B1F7XY, B1F7XB, F09LFP, B32MDF, B1P7Q0, B20X20, B20X2J, B20X2T, B20X3B, B20X3D, B20X3V, B20X4L, 032447, 032057, 039398, B1F8BF, B1MB1Y, B20WDH, 036GQH, B3058Z, B35P0N, F0816G, B32RZ6, B304Y6, F01ZGB, B2RV9R, B35D94, B2NP9T, F03RP9, B35CMQ, B2RV7V, B35J5Q, B39MCT, B2NP80, B2RL78.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Transducer V6-2 989605437982Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls0 reports
- Transducer C5-2 989605454101Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic3 recalls0 reports
- Transducer V6-2 989605415042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic2 recalls0 reports
- Transducer V6-2 989605357563Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic2 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97217).
- Z-2347-2025 · Open, Classified3D9-3v Transducer Probe
- Z-2349-2025 · Open, ClassifiedBP10-5ec Transducer Probe
- Z-2350-2025 · Open, ClassifiedC10-3v Transducer Probe
- Z-2351-2025 · Open, ClassifiedC10-4ec Transducer Probe
- Z-2352-2025 · Open, Classifiedc5-1 Transducer Probe
- Z-2353-2025 · Open, ClassifiedC5-2 Transducer Probe
- Z-2354-2025 · Open, ClassifiedC5-2 Lumify Transducer Probe
- Z-2355-2025 · Open, ClassifiedC6-2 Transducer Probe
- Z-2356-2025 · Open, ClassifiedC6-3 Transducer Probe
- Z-2357-2025 · Open, ClassifiedC8-4v Transducer Probe
- Z-2358-2025 · Open, ClassifiedC8-4v Transducer Probe
- Z-2359-2025 · Open, ClassifiedC8-5 Transducer Probe
- Z-2360-2025 · Open, ClassifiedC9-2 Transducer Probe
- Z-2361-2025 · Open, ClassifiedC9-3io Transducer Probe
- Z-2362-2025 · Open, ClassifiedC9-3v Transducer Probe
- Z-2363-2025 · Open, ClassifiedC9-4 Transducer Probe
- Z-2364-2025 · Open, ClassifiedC9-4v Transducer Probe
- Z-2365-2025 · Open, ClassifiedC9-5ec Transducer Probe
- Z-2366-2025 · Open, ClassifiedD2cwc Transducer Probe
- Z-2367-2025 · Open, ClassifiedD2TCD Transducer Probe
- Z-2368-2025 · Open, ClassifiedD5cwc Transducer Probe
- Z-2369-2025 · Open, ClassifiedeL18-4 Transducer Probe
- Z-2370-2025 · Open, ClassifiedL10-4LAP Transducer Probe
- Z-2371-2025 · Open, ClassifiedL11-3 Transducer Probe
- Z-2372-2025 · Open, ClassifiedL12-3 Transducer Probe
- Z-2373-2025 · Open, ClassifiedL12-4 Transducer Probe
- Z-2374-2025 · Open, ClassifiedL12-4 Lumify Transducer Probe
- Z-2375-2025 · Open, ClassifiedL12-5 38mm Transducer Probe
- Z-2376-2025 · Open, ClassifiedL12-5 50 Transducer Probe
- Z-2377-2025 · Open, ClassifiedL15-7io Transducer Probe
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.