AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays… recallZ-2472-2025 · American Contract Systems Inc · initiated Jul 14, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
FDA root cause category: Under Investigation by firm
What the company is doing
Owens & Minor issued an "URGENT: MEDICAL DEVICE CORRECTION" notice to its consignees on 07/15/2025 via email. The notice explained the problem, potential risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of all affected components. " Add warning labels to affected procedure trays to ensure the components are discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. o Customers (end-users): If you receive a product from a distributor, please contact the distributor or your sales representative with any questions you may have. o Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness for your consignees. " Complete the attached Recall Response Form and return as soon as possible to [email protected]. Report any quality problems or adverse events associated with the products listed in this notification to [email protected]. If you have any questions, contact the Manager, Regulatory Affairs at 470-280-4277 or email to: [email protected].
Recall details
- Recall number
- Z-2472-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 14, 2025
- Posted by the FDA
- Aug 28, 2025
- Recalling firm
- American Contract Systems Inc
- Quantity
- 588 units
- Product code
- OJH
- Distribution
- US distribution to states of: IL and LA
Product description
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL KNEE DAN MYER, MODEL/ITEM NUMBER CCKM67F; 3)HIP PROTHESIS PACK, MODEL/ITEM NUMBER HSTH03N; 4)DTC ANTERIOR HIP PACK (PS 209459), MODEL/ITEM NUMBER MIAH65B; 5)DR WHIDD0N TOTAL HIP CPP, MODEL/ITEM NUMBER MPWH01B; 6)KNEE ARTHROSCOPY, MODEL/ITEM NUMBER SPKA57T; 7)ANTERIOR HIP PACK, MODEL/ITEM NUMBER WVAH40H.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
1) MODEL/ITEM NUMBER AMAH40N: UDI/DI 00191072235473, Kit Lot Numbers: 42-7886011, 42-7990911; 2)MODEL/ITEM NUMBER CCKM67F: UDI/DI 00191072233578, Kit Lot Numbers: 70-053862, 70-053457, 70-053042, 70-052715, 70-052358, 70-051850, 70-051243; 3)MODEL/ITEM NUMBER HSTH03N: UDI/DI 00191072189615, Kit Lot Numbers: 42-7115711; 4)MODEL/ITEM NUMBER MIAH65B: UDI/DI 00191072235411, Kit Lot Numbers: 42-8134311; 5)MODEL/ITEM NUMBER MPWH01B: UDI/DI 00191072233271, Kit Lot Numbers: 5436801; 6)MODEL/ITEM NUMBER SPKA57T: UDI/DI 00191072235848, Kit Lot Numbers: 12-7400011, 12-7381311, 12-7358211, 12-7082111, 12-7157111, 12-7309911; 7)MODEL/ITEM NUMBER WVAH40H: UDI/DI 00191072233004, Kit Lot Numbers: 42-7897411, 42-7966211.
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACS HSTH03NAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS WVAH40HAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS MPWH01BAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS CCKM67FAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS MIAH65BAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS AMAH40NAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS SPKA57TAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97358).
- Z-2464-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM NUMBER ANVA82AB…
- Z-2465-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE…
- Z-2466-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO PACK, Model/Item…
- Z-2467-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) A-C SUPPLEMENT PACK - 205946, Model/Item Number ANCV67X; 2) KIDNEY PERFUSION PACK - 208439, Model/Ite…
- Z-2468-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798), MODEL/ITEM NUM…
- Z-2469-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C; 2) LABOR & DELIVERY PACK, Model/Item Number LLLD19I.
- Z-2470-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY PACK, MODEL/ITEM NUMBER EVLP61I; 2) PELVISCOPY, MODEL/ITEM NUMBER EVPV29G.
- Z-2471-2025 · Open, ClassifiedAMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ROBOT HYST PACK, MODEL/ITEM NUMBER GRRB10E; 2) VAG HYST PACK, MODEL/ITEM NUMBER SPVH62P.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page