Tandem Mobi recallZ-0427-2026 · Tandem Diabetes Care, Inc. · initiated Oct 6, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Tandem Diabetes Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.

FDA root cause category: Software design

What the company is doing

On 10/6/2025, correction notices were mailed or emailed to customers who were informed the following: If your pump shows Malfunction 12, insulin delivery will stop, and the pump becomes non-operational: 1) Promptly acknowledge it by pressing "Silence Alarm". 2) Contact Tandem Diabetes Care Customer Technical Support (CTS) via phone at 1-877-801-6901 or email at [email protected] for further assistance. 3) Switch to your backup method of insulin delivery as directed by your physician. 4) Regularly check your blood sugar to ensure you are not having unexpectedly high or low readings. Update your pump software as soon as possible to version 7.9.0.2. You can check your software version within the Mobi Mobile App. Select Settings > Pump > Pump Info. The latest software can be downloaded remotely from within the Tandem Mobi Mobile App. Once logged in to the app, tap "Settings" and then "Pump" to view software updates. Complete the acknowledgement form using link: https://campaign.tandemdiabetes.com/FCA-Malfunction-12 User Guide: https://www.tandemdiabetes.com/docs/default-source/user-guide/user-guide-tandem-mobi-control-iq-7-9-mgdl-en-us-aw1014938.pdf?sfvrsn=872fc4d7_10

Recall details

Recall number
Z-0427-2026
Status
Open, Classified
Initiated by the company
Oct 6, 2025
Posted by the FDA
Nov 5, 2025
Recalling firm
Tandem Diabetes Care, Inc.
Quantity
17,745
Product code
QFG
FDA submissions
K223213, K233044, K240309, K241078
Distribution
US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, PT, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY

Product description

Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). Pump, Tandem Mobi, Control-IQ/1010750, Pump, Tandem Mobi, Control-IQ, Replacement/1012719, Pump, Tandem Mobi, Control-IQ 7.6.0.3, Medicare/1013501, Pump, Tandem Mobi, Control-IQ 7.7/1013655, Pump, Tandem Mobi, Control-IQ, 7.7, Replacement/1013656, Pump, Tandem Mobi, Control-IQ 7.7, Medicare/1013700, Pharmacy Kit, Starter Pack, Tandem Mobi 7.7/1014081

Code information (models, lots, serial numbers)

Software versions: 7.6.0.1, 7.6.0.3, and 7.7.0.1 Catalog/UDI-DI: 1010750/00389152075013, 1012719/00389152271910, 1013501/00389152350110, 1013655/00389152365510, 1013656/00389152365619, 1013700/00389152370019, 1014081/00389152408170

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K223213, K233044, K240309, K241078 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page