Tandem Mobi 1010750Recalls & reported problems
Tandem Diabetes Care, Inc. · Tandem Mobi Pump with Control-IQ · Class 2
All models of Tandem Mobi (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Tandem Mobi insulin pump with Control-IQ delivers insulin for people with diabetes. Its 2,050 adverse event reports mostly describe the pump stopping and blockages in the tubing, and some mention high or low blood sugar. The reports include no deaths and 262 injuries. One recall, started in October 2025, is open and concerns a software issue that can falsely report a vibration motor failure, leaving the pump unable to operate.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,030
- Deaths reported
- 0
- Injuries reported
- 259
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2,030 adverse event reports that list this exact device (DI 00389152075013), received between Oct 25, 2023 and Sep 18, 2026 — classified as 259 injuries, 1,771 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (799 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Sep 23, 2026Class IOpen, Classified
Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
Z-0316-2027 · Root cause: Process change control · Recall details · FDA record ↗
- Oct 6, 2025Class IOpen, Classified
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
Z-0427-2026 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Pumping Stopped799
- Obstruction of Flow616
- Partial Blockage121
- Mechanical Problem87
- Adverse Event Without Identified Device or Use Problem64
- Charging Problem55
- Failure to Charge54
- Communication or Transmission Problem52
Patient problems
- No Clinical Signs, Symptoms or Conditions1,773
- Hyperglycemia238
- Hypoglycemia23
- Elevated Ketones/ Diabetic Ketoacidosis13
- Loss of consciousness2
- Vomiting2
- Fall1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 2,030 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50812793 ↗MalfunctionSep 14, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50764041 ↗InjurySep 10, 2026Obstruction of Flow
It was reported that ongoing intermittent occlusion alarms occurred, leading to the customer experiencing intermittent high blood glucose (BG) levels. The customer¿s BG level was displayed as ¿High¿; however, a specific value was not provided. The elevated BG was addressed by a correction injection via manual insulin therapy. No third party assistance was required. Customer changed infusion set and cartridge to address the issue. The customer continued to use the pump for insulin therapy and has an alternate method of insulin therapy available if necessary.
Patient problems: Hyperglycemia
MAUDE report 50725086 ↗MalfunctionSep 8, 2026Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50675209 ↗MalfunctionSep 3, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50594984 ↗MalfunctionSep 3, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50577250 ↗MalfunctionAug 28, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50502609 ↗InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the customer experienced an elevated blood glucose (BG) level displayed as "High"; although specific value was not provided. Cause was not known. Customer administered a correction injection to address the BG. Recommendation was made to consult a healthcare provider in regards to diabetes management.
Patient problems: Hyperglycemia
MAUDE report 50441738 ↗MalfunctionAug 20, 2026Pumping Stopped
It was reported that Tandem Mobi showed cartridge alarm related to motor failure mode while pump was in use, and customer contacted support for assistance. There was no adverse impact to the customer¿s blood glucose level. Customer removed cartridge and delivered 9-unit bolus as instructed, then successfully loaded new cartridge, and issue was resolved with no further assistance needed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50347956 ↗MalfunctionAug 19, 2026Failure to Charge
It was reported that the pump battery could not be charged. There was no adverse impact to the customer's blood glucose. Multiple attempts were made by Tandem Technical Support to follow-up with the customer on the reported issue, but no response was received.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50326725 ↗MalfunctionAug 19, 2026Pumping Stopped
It was reported that pump displayed software malfunction alarm code Malf 0, and caller contacted Technical Support for assistance. There was no adverse impact to the customer¿s blood glucose level. Customer worked with Technical Support to reset pump using provided instructions, malfunction did not recur after reset, customer was advised to continue using pump and to call back if any malfunction alarm appeared again, and caller acknowledged and understood instructions.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50322751 ↗InjuryAug 17, 2026Use of Incorrect Control/Treatment Settings
It was reported that the customer experienced a low blood glucose (BG) level of 40 mg/dL. Reportedly, the customer's settings were entered incorrectly. Customer drank soda to address BG. Customer updated their pump settings to address the event.
Patient problems: Hypoglycemia
MAUDE report 50277924 ↗MalfunctionAug 7, 2026Failure to Charge
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50178510 ↗MalfunctionAug 6, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50162092 ↗MalfunctionAug 3, 2026Pumping Stopped
It was reported that ongoing cartridge alarms occurred during insulin delivery. Reportedly, customer will continue to use the pump for insulin therapy and utilize mobile app as required. There was no reported adverse impact to the customer¿s blood glucose level.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50114393 ↗MalfunctionJul 31, 2026Pumping Stopped
It was reported that the pump experienced a malfunction. There was no reported adverse impact to the customer¿s blood glucose level. The pump was replaced to resolve the issue, and the customer will use multiple daily injections (MDI) as a backup therapy until the replacement arrives.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50104153 ↗MalfunctionJul 28, 2026Pumping Stopped
A malfunction was reported on the pump, and it displayed a non-resettable malfunction code, with no events occurred prior to the issue. There was no reported adverse impact to the customer's blood glucose level. The customer reverted to an alternate method of insulin therapy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50052398 ↗MalfunctionJul 27, 2026Obstruction of Flow
It was reported that ongoing intermittent occlusion alarms occurred. There was no adverse impact to the customer¿s blood glucose level. Customer resolved each event by changing infusion set and cartridge or resuming insulin only. The customer continued to use the pump for insulin therapy and has an alternate method of insulin therapy available if necessary.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50025577 ↗MalfunctionJul 8, 2026Pumping Stopped
IT WAS REPORTED THAT PUMP DISPLAYED MALFUNCTION ALARM CODE MALF 24 WITH FAULT ID 24-0X2219 AND COULD NOT BE RESET, AND THERE WERE NO NOTABLE EVENTS PRIOR TO THE MALFUNCTION. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER PLANNED TO USE MULTIPLE DAILY INJECTIONS AS BACKUP INSULIN THERAPY WHILE TANDEM TECHNICAL SUPPORT ARRANGED SHIPMENT OF REPLACEMENT PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25747709 ↗MalfunctionJul 7, 2026Pumping Stopped
IT WAS REPORTED THAT PUMP DISPLAYED MALFUNCTION ALARM MALF 9 WITH NO NOTABLE EVENTS OCCURRING BEFOREHAND. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER CONTACTED TECHNICAL SUPPORT, WAS GUIDED THROUGH RESETTING PUMP USING PROVIDED INSTRUCTIONS, CONFIRMED THAT MALFUNCTION DID NOT RECUR AFTER RESET, AND WAS ADVISED TO CONTINUE USING PUMP AND CALL BACK IF ANY MALFUNCTION CODE APPEARED AGAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25732124 ↗MalfunctionJul 6, 2026Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25726879 ↗MalfunctionJul 2, 2026Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE WAKE BUTTON WAS NOT WORKING. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. MULTIPLE ATTEMPTS WERE MADE BY TANDEM¿S TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION; HOWEVER, NO RESPONSE WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25692987 ↗MalfunctionJul 2, 2026Pumping Stopped
IT WAS REPORTED THAT PUMP DISPLAYED MALFUNCTION ALARM CODE 9 WITH NO NOTABLE EVENTS OCCURRING BEFOREHAND. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT PROVIDED PUMP RESET INSTRUCTIONS, ASSISTED CALLER WITH RESETTING PUMP, CONFIRMED MALFUNCTION DID NOT RECUR, AND ADVISED CUSTOMER TO CONTINUE USING PUMP AND TO CALL BACK IF ANY MALFUNCTION ALARMS APPEARED AGAIN, WHICH CALLER ACKNOWLEDGED AND UNDERSTOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25690919 ↗MalfunctionJun 30, 2026Failure to Charge
IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. REPORTEDLY, THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25665881 ↗MalfunctionJun 29, 2026Pumping Stopped
IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING INSULIN DELIVERY. CUSTOMER RELOADED THE CARTRIDGE TO RESOLVE THE ISSUE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25659301 ↗
FDA clearance and classification
Cleared through 510(k) K223213 on Jul 10, 2023 (decision: Substantially Equivalent). Submitted by Tandem Diabetes Care, Inc. as “Tandem Mobi Insulin Pump with Interoperable Technology”. The FDA review took 266 days. FDA record ↗
The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗
FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.
Common questions about the Tandem Mobi 1010750
Has the Tandem Mobi 1010750 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0316-2027 (Class I), began on Sep 23, 2026: Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Tandem Mobi 1010750?
The FDA's MAUDE database lists 2,030 adverse event reports for this device, received between Oct 25, 2023 and Sep 18, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (799), “Obstruction of Flow” (616) and “Partial Blockage” (121); the most common patient problems are “Hyperglycemia” (238), “Hypoglycemia” (23) and “Elevated Ketones/ Diabetic Ketoacidosis” (13). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Tandem Mobi 1010750?
None of the reports are classified as deaths, 259 as injuries and 1,771 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Tandem Mobi 1010750 FDA approved?
It was cleared through 510(k) K223213 on Jul 10, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Tandem Mobi 1010750 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Tandem Mobi 1010750, and is it still sold?
The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand Tandem Mobi, model 1010750). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Alternate Controller Enabled Insulin Infusion Pump devices
- t:slim X2 Insulin Pump with Control-IQ 1006402Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls18 deaths159,198 reports
- t:slim X2 Insulin Pump with Control-IQ 1006408Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump2 recalls12 deaths92,505 reports
- t:slim X2 Insulin Pump with Control-IQ 1009637Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall4 deaths83,767 reports
- Omnipod 5 Pod POD-BLE-H1-520Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump1 recall74 deaths71,491 reports
- Omnipod 5 Pod POD-OMNI-I1-6720Insulet Corporation · Alternate Controller Enabled Insulin Infusion Pump2 recalls11 deaths60,668 reports
- t:slim X2 Insulin Pump with Control-IQ 1009643Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall7 deaths51,050 reports
- t:slim X2 Insulin Pump with Control-IQ Technology 1013198Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall8 deaths47,498 reports
- t:slim X2 Insulin Pump with Control-IQ 1006404Tandem Diabetes Care, Inc. · Alternate Controller Enabled Insulin Infusion Pump1 recall18 deaths43,714 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page