Tandem Mobi 1013655Recalls & reported problems

Tandem Diabetes Care, Inc. · Tandem Mobi Pump with Control-IQ · Class 2

All models of Tandem Mobi (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is a Tandem Mobi insulin pump with Control-IQ. Its reports mostly describe pump stoppages, blocked flow, repeated occlusion and cartridge alarms, and malfunction codes. There are 29,601 reports, including 3 deaths and 3,893 injuries. One recall, started in October 2025, is open: a software issue can falsely flag a vibration motor failure and stop the pump from working. Reports dropped to 7,560 in the last 12 months, from 20,314 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29,532
Deaths reported
3
Injuries reported
3,891
Recalls naming it
1

Summary

The FDA's MAUDE database contains 29,532 adverse event reports that list this exact device (DI 00389152365510), received between Jun 25, 2024 and Sep 30, 2026 — classified as 3 deaths, 3,891 injuries, 25,638 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (15,688 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 6 reportsJul 2024: 112 reportsAug 2024: 338 reportsSep 2024: 500 reportsOct 2024: 771 reportsNov 2024: 884 reportsDec 2024: 885 reportsJan 2025: 1,938 reportsFeb 2025: 1,273 reportsMar 2025: 1,731 reportsApr 2025: 1,472 reportsMay 2025: 2,318 reportsJun 2025: 2,062 reportsJul 2025: 2,973 reportsAug 2025: 2,206 reportsSep 2025: 1,516 reportsOct 2025: 1,056 reportsNov 2025: 1,168 reportsDec 2025: 1,357 reportsJan 2026: 1,086 reportsFeb 2026: 881 reportsMar 2026: 997 reportsApr 2026: 839 reportsMay 2026: 534 reportsJun 2026: 338 reportsJul 2026: 138 reportsAug 2026: 56 reportsSep 2026: 97 reportsOct 2026: 0 reports2024202520262,973
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 6, 2025Class IOpen, Classified

    Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.

    Z-0427-2026 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped15,688
  • Obstruction of Flow8,458
  • Mechanical Problem1,290
  • Adverse Event Without Identified Device or Use Problem1,128
  • Charging Problem712
  • Partial Blockage483
  • Failure to Charge284
  • Incorrect Measurement247

Patient problems

  • No Clinical Signs, Symptoms or Conditions25,675
  • Hyperglycemia3,340
  • Hypoglycemia505
  • Elevated Ketones/ Diabetic Ketoacidosis165
  • Insufficient Information39
  • Loss of consciousness24
  • Vomiting19
  • Nausea11

Recent adverse event reports

The 25 most recent of 29,532 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50955098 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50955097 ↗
  • InjurySep 30, 2026Obstruction of Flow

    Patient problems: Hyperglycemia

    MAUDE report 50952487 ↗
  • MalfunctionSep 30, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952008 ↗
  • MalfunctionSep 30, 2026Failure to Charge

    It was reported that Mobi pump battery would not charge and patient did not have pump available at the time, so troubleshooting could not be performed. There was no reported impact to the customer¿s blood glucose level. Customer agreed to call back with pump available so battery issue intake could be completed and assistance could be provided with setting up replacement pump, while CTS attempted follow-up by phone and left a message, then sent a follow-up email after no response and closed case with instructions to reopen if patient later contacted support to complete required intake.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949786 ↗
  • MalfunctionSep 30, 2026Obstruction of Flow

    It was reported that customer experienced repeated occlusion alarms on Tandem Mobi pump. There was no adverse impact to the customer¿s blood glucose level. Customer primed tubing, changed infusion set and cartridge, and was able to resume insulin, and later changed supplies again but continued to experience ongoing occlusion issues; CTS advised customer to monitor and seek additional help if alarms recurred, then arranged replacement pump with updated software, provided instructions for storage and return of problematic pump, and guided customer on setting up replacement pump and connecting continuous glucose monitor, while also following up by SMS to obtain supply information before closing case. The customer continued to use the pump for insulin therapy and has an alternate method of insulin therapy available if necessary.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949129 ↗
  • MalfunctionSep 30, 2026Obstruction of Flow

    It was reported that customer experienced repeated occlusion alarms on Mobi pump over several months, including active alarms that recurred during troubleshooting and were associated with supply changes, though customer did not report issues with infusion sites or insulin. There was no adverse impact to the customer¿s blood glucose level. Customer followed Technical Support instructions to disconnect tubing, adjust connections, and deliver test boluses, and clinical support recommended pump replacement; customer was advised to use multiple daily injections until replacement pump arrived, was given education on putting pump in storage mode and returning pump, and Tandem Technical Support proceeded with arranging replacement and follow-up.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948791 ↗
  • MalfunctionSep 30, 2026Obstruction of Flow

    It was reported intermittent occlusion alarms occurred. There was no reported adverse impact to the customer¿s blood glucose level. The customer changed supplies and resumed insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946635 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    It was reported that pump displayed malfunction alarm Malf 24 for undertravel stall and could not be reset. There was no adverse impact to the customer¿s blood glucose level. Customer switched to multiple daily injections as backup insulin therapy, and Technical Support arranged shipment of replacement pump, confirmed shipping details, provided instructions to place problem pump into storage mode, and instructed customer to return problem pump within 10 days using provided return materials, then to enter pump settings and reconnect continuous glucose monitor on replacement pump.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941754 ↗
  • MalfunctionSep 29, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941255 ↗
  • MalfunctionSep 29, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941234 ↗
  • MalfunctionSep 29, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941131 ↗
  • MalfunctionSep 29, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941107 ↗
  • MalfunctionSep 29, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940983 ↗
  • InjurySep 29, 2026Obstruction of Flow

    Patient problems: Hyperglycemia

    MAUDE report 50940955 ↗
  • MalfunctionSep 29, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940952 ↗
  • MalfunctionSep 29, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940840 ↗
  • MalfunctionSep 29, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940823 ↗
  • MalfunctionSep 29, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939736 ↗
  • MalfunctionSep 29, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938523 ↗
  • MalfunctionSep 29, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938507 ↗
  • MalfunctionSep 29, 2026Pumping Stopped

    It was reported that pump displayed malfunction message and customer was not able to reset pump while away from home. There was no adverse impact to the customer¿s blood glucose level. Customer planned to call technical support again when at home to reset pump, and technical support attempted follow-up by phone but reached voicemail, and no additional information was received regarding the outcome of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938457 ↗
  • MalfunctionSep 29, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938450 ↗
  • MalfunctionSep 29, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938240 ↗
  • MalfunctionSep 29, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937872 ↗