BD Alaris infusion Pump Module 8100, REF: 8100 recallZ-0430-2026 · CareFusion 303, Inc. · initiated Oct 17, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
FDA root cause category: Device Design
What the company is doing
On 10/17/2025, correction notices were mailed and emailed: Distribute notice to clinical and non-clinical users. Safety Reminder: https://academy.bd.com/en-us/home. Provide notice to any facilities you may have sold or transferred devices to. Use only BD-authorized companies when renting infusion devices. Complete and return response form to [email protected] If dropped, biomedical engineering should perform testing as specified in Level of Testing Guidelines, including rate accuracy verification testing. If any test fails see troubleshooting section, see Technical Service Manual, Chapter 5 and 6. User Manuals: https://bd.com/self-service. Firm contacts: Technical Support 888-812-3229, [email protected] . Product Complaints: [email protected] BD Alaris Infusion System Safety Reminder: To ensure system remains in good operating condition, if a device or accessory is dropped or severely jarred, take the device out of use. Even if no visible damage is present, internal components may be affected. Return the device to Biomedical Engineering for inspection and repair. Visually inspect the system before each use. Check all visible surfaces and moving parts and if you observe damage or find the device does not function as expected, return it to biomedical engineering for repair. Do not use a device that appears to be damaged or does not function as expected. Return device to Biomedical Engineering for repair. Only biomedical engineering should perform annual maintenance as specified in the System Maintenance Software User Manual. Important in ensuring regular testing of any potentially dropped or damaged device that was not immediately removed from use to prevent infusion errors, protect patient safety, and promote proper function. All should understand the potential safety risks associated with using a device that may have sustained damage.
Recall details
- Recall number
- Z-0430-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 17, 2025
- Posted by the FDA
- Nov 20, 2025
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 2,574,396
- Product code
- FRN
- FDA submissions
- K211218
- Distribution
- Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.
Product description
BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide
Code information (models, lots, serial numbers)
All Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths51,256 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion3 recalls1 death6,859 reports
- Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion1 recall2 deaths5,636 reports
- BD Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page