BD Pyxis MedStation ES, REF: 323 recallZ-0962-2026 · CareFusion 303, Inc. · initiated Nov 20, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.
FDA root cause category: Device Design
What the company is doing
On 11/20/2025, correction notices were mailed and emailed to customers: 1) If a pocket fails with the error message of "Duplicate Address Detected," users with the proper permissions can leverage the "Recover Storage Space" functionality to eject the affected pocket and clear the error. Note: This does not resolve the Firmware v1.48 issue. This only clears the error message from the "Recover Storage Space" screen. Content stored within the affected CUBIE pocket will still need to be manually accessed once the CUBIE pocket is ejected. 2) Facilities should have comprehensive policies and procedures that outline alternative ways to access items, which may include removing from a different device or from pharmacy. 3. A redundant supply of all life sustaining medication should be kept in a nearby location to help mitigate the length of delay. 4. Firm recommends making a "code-box" or "crash cart" of medications that are readily available outside of an automated storage device in the event of an emergency. Facility policies and procedures should clearly identify when it is appropriate to access emergency medications. Refer to product user guide for additional information. 5. Keys can be used to manually access unavailable drawers. Keys must be readily available and retrievable to constitute risk mitigation. Hospital policies and procedures should be devised by the facility indicating location of keys along with how they are to be obtained and used in downtime scenarios. 6. Share and disseminate this notice within your facility network. 7. Complete and return the response form via email to [email protected] Firm will install the firmware 1.49 upgrade either remotely or via a field service visit. Note: Firmware updates to main devices will update the Aux when connected to the main device. Firm contacts: Remediation Support: [email protected] Technical Support: bd.com/self-service, Phone:1-800-727-6102
Recall details
- Recall number
- Z-0962-2026
- Status
- Open, Classified
- Initiated by the company
- Nov 20, 2025
- Posted by the FDA
- Dec 19, 2025
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 53669
- Product code
- BRY
- Distribution
- Worldwide - US Nationwide distribution including in the states of CO, IN, OH, TX, KS, FL, GA, NC, CA, HI, IL, NY, MI, VA, AL, LA, NE, KY, OK, AR, TN, WY, NJ, WI, SD, AZ, NV, MO, MA, ID, WV, DE, PA, WA, MT, CT, IA, MD, PR, MN, NM, OR, VT, MS, ME, AK, UT, DC, GU, SC, NH, RI, ND, AS and the countries of CH, CA, KW, QA, NL, GB, UY, AU, NZ, BH, TH, SG, CN, FR, LB, BS, HK, TR, PL, AR, BR, CL, CO, MX, ES, BM, SA, IL.
Product description
BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD Pyxis CII Safe ES, REF: 1116-00
Code information (models, lots, serial numbers)
Firmware Version: 1.48. REF/UDI-DI: 323/10885403512667, 324/10885403512681, 352/10885403512674, 1149-00/10885403517822, 1152-00/10885403517839, 1155-00/10885403521300, 1116-00/10885403512605
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- BD Pyxis 323Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA14 recalls2 deaths24,375 reports
- BD Pyxis 324Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA5 recalls1 death5,301 reports
- BD Pyxis 352Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA12 recalls898 reports
- Pyxis Pro MedStation 1155-00Carefusion 303, Inc. · Cabinet, table and tray, anesthesia2 recalls23 reports
- Pyxis Pro MedStation 1149-00Carefusion 303, Inc. · Cabinet, table and tray, anesthesia2 recalls4 reports
- Pyxis Pro MedStation 1152-00Carefusion 303, Inc. · Cabinet, table and tray, anesthesia2 recalls2 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.