Impella Heart Pump recallZ-1470-2026 · Abiomed, Inc. · initiated Jan 27, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
FDA root cause category: Under Investigation by firm
What the company is doing
Abiomed notified consignees on about 01/27/2026 via UPS delivery. Additionally, the Customer Letter was posted on Heartrecovery.com: https://www.heartrecovery.com/en-us/resources/downloads/field-safety-notification-impella-rp-differential-pressure-sensor. The notifications informed consignees were instructed clinicians that affected units can continue use and they should: 1. Continue to monitor patient hemodynamics with approved diagnostic devices and continue to verify Impella positioning with imaging before clinical interventions. 2. Refer to the P-level flow rates listed in the Instructions for Use (IFUs), as shown in Figure 1, and rely on these values rather than the AIC display. 3. Trend pump metrics to monitor pump performance and to identify sudden change, noting that abrupt changes in the displayed flow rate need to be assessed both against the flow rate listed in the IFU and across temporal trends. Consignees were instructed that product is NOT being removed from the field and does not need to be returned, patients should be assessed with approved diagnostic devices before clinical interventions, complete and return the provided business reply form (BRF) to Abiomed4437@sedgwick, forward the notification to anyone in the facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products) and if any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice.
Recall details
- Recall number
- Z-1470-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 27, 2026
- Posted by the FDA
- Feb 27, 2026
- Recalling firm
- Abiomed, Inc.
- Quantity
- 4,496 units
- Product code
- PYX
- FDA submissions
- P170011
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Canada, Croatia, Czechia, France, Germany, Hong Kong, Italy, Poland, Serbia, Switzerland.
Product description
Impella RP Flex with SmartAssist. Product Code: 1000323.
Code information (models, lots, serial numbers)
Product Code: 1000323. UDI-DI: 00813502012811.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Impella 1000323Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump6 recalls189 deaths943 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under P170011 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Impella 0042-0000-USAbiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump12 recalls120 deaths2,456 reports
- Impella 0042-0040-USAbiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump8 recalls41 deaths1,305 reports
- Impella 004334Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump3 recalls61 deaths408 reports
- Impella 0046-0035Abiomed, Inc. · Temporary Non-Roller Type Right Heart Support Blood Pump4 recalls37 deaths258 reports
- Impella 1000200Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump9 deaths84 reports
- Impella 0043-0003Abiomed, Inc. · Temporary Non-Roller Type Left Heart Support Blood Pump1 recall7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98276).
- Z-1469-2026 · Open, ClassifiedImpella RP with SmartAssist. Product Code: 0046-0035.
- Z-1471-2026 · Open, ClassifiedImpella RP. Product Code: 0046-0011.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page