SHOCKPULSE recallZ-1478-2026 · Olympus Corporation of the Americas · initiated Nov 14, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
FDA root cause category: Device Design
What the company is doing
On or about December 18, 2025 URGENT: MEDICAL DEVICE CORRECTION - UPDATE letters were sent to customers. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification, including the new Supplemental Guidance. Add a copy of this letter with the Supplemental Guidance to your existing Instruction for Use. You may continue to use the device according to this letter and the instruction for use, which cautions users to ensure that a back-up transducer and probe are sterilized and available prior to beginning a procedure. If you have further distributed this product, identify and forward this notification to them.
Recall details
- Recall number
- Z-1478-2026
- Status
- Open, Classified
- Initiated by the company
- Nov 14, 2025
- Posted by the FDA
- Feb 27, 2026
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 1082 units
- Product code
- FEO
- FDA submissions
- K171024
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.
Product description
Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
Code information (models, lots, serial numbers)
Model/Catalog Number: SPL-S; UDI: 00821925044197; Serial Numbers: All;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ShockPulse SPL-SGyrus ACMI, LLC · Lithotriptor, ultrasonic2 recalls159 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K171024 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ShockPulse SPL-TGyrus ACMI, LLC · Lithotriptor, ultrasonic279 reports
- ShockPulse SPL-GGyrus ACMI, LLC · Lithotriptor, ultrasonic2 recalls10 reports
- ShockPulse SPL-FSGyrus ACMI, LLC · Lithotriptor, ultrasonic20 reports
- SHOCKPULSE SPL-PDBX150Gyrus ACMI, LLC · Lithotriptor, Ultrasonic2 reports
- Gyrus ACMI SPL-WGyrus ACMI, LLC · Lithotriptor, ultrasonic1 report
- SHOCKPULSE SPL-PDBX183Gyrus ACMI, LLC · Lithotriptor, Ultrasonic1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 98359).
- Z-1477-2026 · Open, ClassifiedBrand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes Model/Catalog Number: SPL-SR Product Description: The ShockPuls…
- Z-1479-2026 · Open, ClassifiedBrand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product Description: The ShockPulse Lithotripsy …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page