ShockPulse SPL-GRecalls & reported problems
Gyrus ACMI, LLC · LITHOTRIPSY GENERATOR ShockPulse-SE · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00821925044203), received between Oct 23, 2023 and Jan 23, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (K171024); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 8, 2026Open, Classified
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
Z-1494-2026 · Root cause: Device Design · FDA record ↗
- Nov 14, 2025Open, Classified
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
Z-1479-2026 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Break2
- Melted2
- Activation Problem1
- Defective Device1
- Electrical /Electronic Property Problem1
- Energy Output Problem1
- Output Problem1
- Power Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Break
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24158345 ↗MalfunctionDec 9, 2025Power Problem; Power Problem
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23757936 ↗MalfunctionNov 6, 2025Melted; Melted
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23485517 ↗MalfunctionSep 12, 2025Output Problem; Output Problem
IT WAS REPORTED THAT THE SUBJECT DEVICE'S POWER-ON LIGHT WAS NOT FUNCTIONING. THE ISSUE OCCURRED DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23039066 ↗MalfunctionSep 2, 2025Energy Output Problem; Energy Output Problem
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD INAPPROPRIATE OUTPUT. THE ISSUE WAS FOUND DURING SET UP / INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22946498 ↗MalfunctionAug 27, 2025Melted; Melted
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A SOCKET CONNECTOR BURN. THE ISSUE OCCURRED DURING SET UP/ INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22903176 ↗MalfunctionAug 4, 2025Activation Problem; Activation Problem
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22701509 ↗MalfunctionJan 2, 2025Defective Device; Defective Device
IT WAS REPORTED THAT THE SUBJECT DEVICE SHOWED AN ERROR ON PANEL AND WOULD NOT WORK. THE ISSUE WAS FOUND DURING SET UP/ INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21059932 ↗MalfunctionDec 27, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THE SHOCKPULSE LITHOTRIPSY GENERATOR EXPERIENCED THAT WHEN PRESSING THE STANDARD BUTTON ON THE TRANSDUCER, THE STATUS ON THE FRONT OF THE MACHINE IS DISPLAYED AT HIGH. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21029701 ↗MalfunctionOct 23, 2023Break; Break
THE CUSTOMER REPORTED THAT DURING A LITHOTRIPSY PROCEDURE THE DEVICE STOPPED WORKING AND COULDN¿T BE ACTIVATED BY THE PEDAL. THE PEDAL WAS DISCONNECTED TO MANUALLY ACTIVATE THROUGH THE TRANSDUCER AND THE SAME ERROR OCCURRED. AN ERROR WITH AN X APPEARED. THEN IT WAS NOTICED THAT THE TRANSDUCER LOOKED LIKE IT WAS BURNED AND BROKEN INSIDE. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE WITH NO DELAY REPORTED. THERE WERE NO REPORTS OF HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17990971 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171024. One submission often covers many models, so check the product and lot details in each record.
- Z-1477-2026 · Nov 14, 2025 · Open, ClassifiedBrand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes Model/Catalog Number: SPL-SR Product Description: The ShockPuls
- Z-1478-2026 · Nov 14, 2025 · Open, ClassifiedBrand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPuls
- Z-0452-2026 · Oct 8, 2025 · Open, ClassifiedBrand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electr
- Z-1514-2022 · Jul 7, 2022 · Open, ClassifiedShockPulse-SE Lithotripsy Generator Model/Serial: SPL-S
Other Lithotriptor, ultrasonic devices
- ShockPulse SPL-TGyrus ACMI, LLC · Lithotriptor, ultrasonic279 reports
- ShockPulse SPL-SGyrus ACMI, LLC · Lithotriptor, ultrasonic2 recalls159 reports
- SHOCKPULSE SPL-SRGyrus ACMI, LLC · Lithotriptor, Ultrasonic2 recalls97 reports
- Lithotripsy Transducer ShockPulse-SE SPL-TCybersonics Inc · Lithotriptor, Ultrasonic1 recall24 reports
- ShockPulse SPL-FSGyrus ACMI, LLC · Lithotriptor, ultrasonic20 reports
- ShockPulse Lithotripsy Generator SPL-GCybersonics Inc · Lithotriptor, Ultrasonic8 reports
- Lithotripsy System ShockPulse-SE SPL-SCybersonics Inc · Lithotriptor, Ultrasonic7 reports
- Lithotripsy Probe, Sterile Single Use ShockPulse -SE, 11.3 Fr (3.76mm) X 396mm SPL-PD376Cybersonics Inc · Lithotriptor, Ultrasonic6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.