Skip to content

ShockPulse SPL-SRecalls & reported problems

Gyrus ACMI, LLC · LITHOTRIPSY SYSTEM ShockPulse-SE · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This ShockPulse lithotripsy system is an ultrasonic lithotripter used to break up stones. Of its 163 adverse event reports, 157 are malfunctions and 6 are injuries, with no deaths. Many describe broken parts, lost output, and power problems. Two open recalls name this device: one over a generator that does not recognize its transducer, and one over a mis-wired part.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
159
Deaths reported
0
Injuries reported
6
Recalls naming it
2

Summary

The FDA's MAUDE database contains 159 adverse event reports that list this exact device (DI 00821925044197), received between Dec 9, 2020 and Sep 29, 2026 — classified as 6 injuries, 153 malfunctions.

The most frequently reported specific device problem is “Break” (33 reports).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (K142428); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 5 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 4 reportsAug 2024: 8 reportsSep 2024: 8 reportsOct 2024: 8 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 9 reportsJul 2025: 8 reportsAug 2025: 13 reportsSep 2025: 4 reportsOct 2025: 8 reportsNov 2025: 7 reportsDec 2025: 7 reportsJan 2026: 1 reportsFeb 2026: 4 reportsMar 2026: 7 reportsApr 2026: 7 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 14, 2025Open, Classified

    Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

    Z-1478-2026 · Root cause: Device Design · FDA record ↗

  2. Jul 7, 2022Open, Classified

    Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment

    Z-1514-2022 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Break33
  • Output Problem19
  • Communication or Transmission Problem12
  • Power Problem11
  • Connection Problem10
  • No Device Output10
  • Defective Device8
  • Energy Output Problem7

Patient problems

  • No Clinical Signs, Symptoms or Conditions144
  • No Patient Involvement11
  • Fistula1
  • Hemorrhage/Blood Loss/Bleeding1
  • No Consequences Or Impact To Patient1
  • Pneumothorax1
  • Sepsis1
  • Unspecified Respiratory Problem1

Recent adverse event reports

The 25 most recent of 159 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Output below Specifications

    It was reported that the shock pulse lithotripsy system exhibited poor power output. The issue occurred during a therapeutic procedure that was delayed for less than 30 minutes under sedation. The procedure was subsequently completed using the same device. No reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935175 ↗
  • MalfunctionSep 22, 2026Activation Problem

    The customer returned the LITHOTRIPSY SYSTEM to the service center for a separate issue. During the device analysis, the investigator indicated that a failed probe test due to faulty control board was found. It was determined that the control board needed to be replaced. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50854807 ↗
  • MalfunctionSep 11, 2026Activation Failure; Complete Loss of Power

    It was reported the Lithotripsy was not powering on. The issue occurred during setup. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50732694 ↗
  • MalfunctionSep 8, 2026Communication or Transmission Problem; Complete Loss of Power

    It was reported the lithotripsy system stopped working and produced an error. This was further clarified to state that a "broken scope" error was produced when the foot pedal was used. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50670237 ↗
  • MalfunctionSep 1, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50548775 ↗
  • MalfunctionAug 12, 2026Thermal Decomposition of Device; Electrical /Electronic Property Problem

    The customer returned the LITHOTRIPSY SYSTEM, which is an Olympus asset with no reported complaint. During the evaluation of the device, faulty transducer receptacles unit at front with burn marks inside was observed. The investigator indicated that the most likely cause of the faulty transducer receptacles with burn marks was due to component failure. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50224410 ↗
  • MalfunctionMay 13, 2026No Device Output

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LITHOTRIPSY SYSTEM FAILED AN OUTPUT TEST DUE TO A FAULTY TRANSDUCER RECEPTACLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25165309 ↗
  • MalfunctionMay 8, 2026Material Twisted/Bent

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC LITHOTRIPTOR HAD BENT PINS ON THE CONNECTOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25125746 ↗
  • MalfunctionApr 15, 2026Break; Key or Button Unresponsive/not Working

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION THE LITHOTRIPSY SYSTEM WAS NOT WORKING WITH THE FOOTSWITCH OR PROBE AND HAD A BROKEN PIN ON THE TRANSDUCER CONNECTOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24888411 ↗
  • MalfunctionApr 15, 2026Output Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION THE LITHOTRIPSY SYSTEM HAD ISSUES WITH ISSUES WITH POWER OUTPUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24886544 ↗
  • MalfunctionApr 15, 2026Power Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SHOCKPULSE LITHOTRIPSY ELECTRICAL SYSTEM HAD INTERMITTENT LOSS OF ENERGY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24884308 ↗
  • MalfunctionApr 14, 2026Output Problem; Positioning Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LITHOTRIPSY SYSTEM WAS NOT GENERATING ENERGY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24877592 ↗
  • MalfunctionApr 14, 2026Power Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SHOCKPULSE LITHOTRIPSY HAD A HIGH POWER: 60 WATTS, A LOW POWER: 37 WATTS, AND POWER WATTS DIFFERENCE: 23 WATTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24874236 ↗
  • MalfunctionApr 6, 2026Material Split, Cut or Torn

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24795497 ↗
  • MalfunctionApr 2, 2026Crack

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24766889 ↗
  • MalfunctionMar 30, 2026No Device Output

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LITHOTRIPSY EXHIBITED FAILED PROBE TEST AND NO OUTPUT DUE TO FAULTY CONTROL BOARD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24736981 ↗
  • MalfunctionMar 26, 2026No Device Output

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC LITHOTRIPTOR'S POWER ON TEST THE POWER ENTRY MODULE IS FAULTY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24699698 ↗
  • MalfunctionMar 24, 2026Output Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LITHOTRIPSY SYSTEM EXHIBITED AN ENERGY OUTPUT PROBLEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24677234 ↗
  • MalfunctionMar 19, 2026Mechanical Problem

    IT WAS REPORTED THE LITHOTRIPSY SYSTEM GENERATOR EXHIBITED AN ¿ERROR¿ SIGNAL AFTER WORKING AT THE BEGINNING OF THE CASE. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24641334 ↗
  • MalfunctionMar 18, 2026Activation Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SHOCKPULSE LITHOTRIPSY SYSTEM RED ERROR LIGHT, AND THE RED CHECK PROBE LIGHTS DID NOT ACTIVATE AFTER FOOTSWITCH WAS HELD DOWN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24630838 ↗
  • MalfunctionMar 10, 2026Communication or Transmission Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24565484 ↗
  • MalfunctionMar 9, 2026Degraded; Output Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE LITHOTRIPSY SYSTEM EXHIBITED AN OUTPUT FAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24548251 ↗
  • MalfunctionFeb 27, 2026Suction Problem; Power Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24474612 ↗
  • MalfunctionFeb 27, 2026Energy Output Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24470963 ↗
  • MalfunctionFeb 27, 2026Energy Output Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24470075 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142428. One submission often covers many models, so check the product and lot details in each record.

  • Z-2197-2020 · Apr 23, 2020 · Terminated
    OLYMPUS ShockPulse-SE Lithotripsy System, Model SPL-T Transducer

Other Lithotriptor, ultrasonic devices

All Lithotriptor, ultrasonic devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.