SHOCKPULSERecalls & reported problems across all models

Gyrus ACMI, LLC · 11 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

5 open recalls name at least one SHOCKPULSE model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
580
Deaths reported
0
Injuries reported
8
Recalls
5

We found 11 models of SHOCKPULSE by Gyrus ACMI, LLC in the FDA's device registry (GUDID). Together they are named in 580 adverse event reports in the FDA's MAUDE database, received between Dec 9, 2020 and Sep 29, 2026 — classified as 8 injuries, 572 malfunctions.

The most frequently reported specific device problem is “Break” (104 reports).

5 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 11 reportsApr 2024: 4 reportsMay 2024: 13 reportsJun 2024: 12 reportsJul 2024: 12 reportsAug 2024: 19 reportsSep 2024: 19 reportsOct 2024: 29 reportsNov 2024: 8 reportsDec 2024: 5 reportsJan 2025: 8 reportsFeb 2025: 12 reportsMar 2025: 13 reportsApr 2025: 15 reportsMay 2025: 6 reportsJun 2025: 19 reportsJul 2025: 45 reportsAug 2025: 28 reportsSep 2025: 25 reportsOct 2025: 34 reportsNov 2025: 27 reportsDec 2025: 23 reportsJan 2026: 19 reportsFeb 2026: 16 reportsMar 2026: 27 reportsApr 2026: 32 reportsMay 2026: 5 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 19 reportsOct 2026: 0 reports20242025202645
Last 36 months. The most recent month may be incomplete.

SHOCKPULSE recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jan 8, 2026Open, Classified

    Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.

    Z-1494-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  2. Nov 14, 2025Open, Classified

    Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

    Z-1479-2026 · Root cause: Device Design · Recall details · FDA record ↗

  3. Nov 14, 2025Open, Classified

    Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

    Z-1478-2026 · Root cause: Device Design · Recall details · FDA record ↗

  4. Nov 14, 2025Open, Classified

    Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.

    Z-1477-2026 · Root cause: Device Design · Recall details · FDA record ↗

  5. Jul 7, 2022Open, Classified

    Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment

    Z-1514-2022 · Root cause: Process change control · Recall details · FDA record ↗

Most reported SHOCKPULSE problems

Device problems

  • Break104
  • Output Problem56
  • Communication or Transmission Problem52
  • Activation Problem36
  • No Device Output31
  • Energy Output Problem28
  • Connection Problem27
  • Failure to Power Up27

Patient problems

  • No Clinical Signs, Symptoms or Conditions545
  • No Patient Involvement22
  • Insufficient Information7
  • No Consequences Or Impact To Patient4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Fistula1
  • Hemorrhage/Blood Loss/Bleeding1
  • Pneumothorax1

SHOCKPULSE models

11 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent SHOCKPULSE adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Gyrus ACMI, LLC products

All Gyrus ACMI, LLC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page