Endo-Model Modular Knee Prosthesis System recallZ-1515-2026 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Jan 12, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
FDA root cause category: Under Investigation by firm
What the company is doing
On or about January 12, 2026 URGENT FIELD SAFETY NOTICE Product Recall letters were sent to customers. Actions to be taken: Review inventory to identify and quarantine any affected product. Return any unused product to the manufacturer. Follow patient management recommendations, if implanted. A custom-made replacement set can be requested. If an affected product is noticed during implantation, another plateau screw from a replacement set or different LINK knee prosthesis package can be used. If this situation occurs, a complaint form must be completed and sent to [email protected]. The replacement of the affected products will begin as soon as replacement products are available. Replacement will not incur any costs to you. Should you have any questions on acquiring replacements for forthcoming surgeries, please contact your local sales representative or customer service for Link products.
Recall details
- Recall number
- Z-1515-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 12, 2026
- Posted by the FDA
- Mar 10, 2026
- Recalling firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Quantity
- 1 unit (OUS only)
- Product code
- KRO
- FDA submissions
- K151008
- Distribution
- Worldwide - US Nationwide distribution in the state of NJ and the countries of Belarus, Canada, Switzerland , Colombia, Germany, Spain, United Kingdom, Hungary, India, Italy, Netherlands, Romania, Saudi Arabia.
Product description
Endo-Model Replacement Plateau; Item Number: 15-2835/12;
Code information (models, lots, serial numbers)
Item Number: 15-2835/12; UDI-DI: 04026575316564; Serial/Lot Number: 241210/1809;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Endo-Model Modular Knee Prosthesis System 15-2835/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
Product lines
Other devices under the same FDA submission
37 devices cleared under K151008 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Endo-Model SL Knee Prosthesis System 16-2817/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model SL Knee Prosthesis System 16-2817/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model SL Knee Prosthesis System 16-2817/07WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model SL Knee Prosthesis System 16-2817/32WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model SL Knee Prosthesis System 16-2817/35WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model SL Knee Prosthesis System 16-2817/37WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls0 reports
- Endo-Model SL Knee Prosthesis System 16-2840/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall17 reports
- Endo-Model SL Knee Prosthesis System 16-2840/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall12 reports
- Endo-Model SL Knee Prosthesis System 16-2840/07WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall2 reports
- Fusion Nail 15-0028/08WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall1 report
- Fusion Nail 15-0028/07WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented5 reports
- Endo-Model Standard Knee Prosthesis System 15-2975/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98342).
- Z-1511-2026 · Open, ClassifiedEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/05;
- Z-1512-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/15;
- Z-1513-2026 · Open, ClassifiedEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;
- Z-1514-2026 · Open, ClassifiedEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/07;
- Z-1516-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-2836/11;
- Z-1517-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8030/12;
- Z-1518-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/09;
- Z-1519-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/11;
- Z-1520-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/15;
- Z-1521-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/11;
- Z-1522-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/12;
- Z-1523-2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/16;
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page