Lantern recallZ-1630-2026 · Orthalign, Inc · initiated Feb 10, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Orthalign, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
FDA root cause category: Nonconforming Material/Component
What the company is doing
On February 10, 2026, OrthAlign issued a Urgent: Medical device Recall Notification via overnight mail. OrthAlign asked consignees to take the following actions: Please review your inventory for the affected products listed in Table 1. 1. Immediately remove and quarantine the affected products until the products are returned to OrthAlign. 2. Please share this recall notification with all members in your network as well as those where the affected devices have been transferred. 3. If you have further distributed the affected products, please identify and notify consignees. You may provide consignees a copy of this notification letter. 4. Please complete the attached response form within Five (5) business days and email the completed form to : [email protected] 5. Adverse actions are quality issues and can be reported to the Food and Drug Administration s MedWatch program: o Online: By completing and submitting the report online at www.fda.gov/medwatch o Regular mail or fax): download the form from www.fda.gov/MedWatch/getforms.htm or call 800-332-1088 to request a reporting form; then complete the form and mail it to the address on the pre-addressed form or submit by fax to 800-332-0178.
Recall details
- Recall number
- Z-1630-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 10, 2026
- Posted by the FDA
- Mar 24, 2026
- Recalling firm
- Orthalign, Inc
- Quantity
- 7 units
- Product code
- OLO
- FDA submissions
- K200892
- Distribution
- US Nationwide distribution in the state of UT.
Product description
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System
Code information (models, lots, serial numbers)
Lot # 25092204BB/ UDI:00810832032434
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Lantern 402567Orthalign, Inc. · Orthopedic Stereotaxic Instrument1 recall0 reports
Product lines
Other devices under the same FDA submission
5 devices cleared under K200892 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Lantern 402566Orthalign, Inc. · Orthopedic Stereotaxic Instrument1 recall0 reports
- Lantern Surgical Assistant (LSA) 406000-02Orthalign, Inc. · Orthopedic Stereotaxic Instrument13 reports
- Lantern Surgical Assistant (HSA) 405204Orthalign, Inc. · Orthopedic Stereotaxic Instrument3 reports
- Lantern Surgical Assistant (LSA) 406001-01Orthalign, Inc. · Orthopedic Stereotaxic Instrument2 reports
- Lantern Surgical Assistant (LSA) KA406000-02Orthalign, Inc. · Orthopedic Stereotaxic Instrument1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 98516).
- Z-1629-2026 · Open, ClassifiedDrill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page