Medline Industries, Inc. recallZ-2092-2026 · Medline Industries, LP · initiated Jan 7, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

FDA root cause category: Process control

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/14/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-

Recall details

Recall number
Z-2092-2026
Status
Open, Classified
Initiated by the company
Jan 7, 2026
Posted by the FDA
May 7, 2026
Recalling firm
Medline Industries, LP
Quantity
39341 units
Product code
OES
Distribution
Worldwide - US Nationwide distribution.

Product description

Medline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AB; 2) OR ARTERIOGRAM, Model Number:CDS985094F; 3) INTRAVASOR CATH KIT, Model Number:DYK1060193I; 4) KIT SURG ONC LAP DIAG/CV CATH, Model Number:DYKMBNDL153; 5) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4A; 6) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4B; 7) VEIN PROCEDURE KIT, Model Number:DYNDA2076C; 8) CARDIAC CATH PACK-LF, Model Number:DYNJ0150605AC; 9) CARDIAC CATH PACK-LF, Model Number:DYNJ0368386D; 10) CATH LAB PACK-LF, Model Number:DYNJ0648473Y; 11) NBMC CATH PACK-LF, Model Number:DYNJ0803367I; 12) RADIOLOGY PACK-LF, Model Number:DYNJ0928124C; 13) ANGIO PACK-LF, Model Number:DYNJ0954970I; 14) SPECIALS TRAY RADIOLOGY, Model Number:DYNJ16803D; 15) CARDIAC CATH PACK, Model Number:DYNJ19946M; 16) CATH LAB PACK, Model Number:DYNJ30955I; 17) CATH LAB PACK, Model Number:DYNJ31773K; 18) CATH PACK-LF, Model Number:DYNJ33061D; 19) DB CCL OAKWOOD PACK-LF, Model Number:DYNJ33620D; 20) CARDIAC CATH PACK MRH-LF, Model Number:DYNJ35300M; 21) CATH LAB PK, Model Number:DYNJ35643C; 22) CARDIAC CATH PACK-LF, Model Number:DYNJ39066B; 23) ANGIOGRAPHY PACK, Model Number:DYNJ42134G; 24) CATH PACK-LF, Model Number:DYNJ43101C; 25) OR ANGIO PACK-LF, Model Number:DYNJ43415D; 26) OR ANGIO PACK-LF, Model Number:DYNJ43415F; 27) ANGIOPLASTY PACK, Model Number:DYNJ44700J; 28) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJ44850J; 29) RADIOLOGICAL PACK, Model Number:DYNJ44889O; 30) CARDIAC CATH PACK, Model Number:DYNJ47370B; 31) ANGIOGRAPHY PACK, Model Number:DYNJ53129D; 32) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJ56772G; 33) DR. MOHINDRA PACK, Model Number:DYNJ57602G; 34) NE ENDO VASCULAR PACK, Model Number:DYNJ57940B; 35) CARDIAC CATH PACK - Q1, Model Number:DYNJ58048C; 36) OR ANGIO, Model Number:DYNJ59931B; 37) LPCH PERRY CATH PACK, Model Number:DYNJ67378C; 38) HOSPITAL REGIONAL DE CONCEPCIO, Model Number:DYNJ67918A; 39) CATH LAB PACK, Model Number:DYNJ68633; 40) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197B; 41) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197C; 42) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197D; 43) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197F; 44) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197G; 45) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197I; 46) BASIC CARDIAC CATH PACK, Model Number:DYNJ81593B; 47) CATH LAB PACK, Model Number:DYNJ83073; 48) UMC EL PASO CATH PACK, Model Number:DYNJ83089; 49) DR. MOHINDRA PACK, Model Number:DYNJ83819A; 50) VZ VAS/ANG PACK, Model Number:DYNJ83981; 51) WCPN CARDIOLOGY PACK, Model Number:DYNJ84539; 52) CATH LAB, Model Number:DYNJ85511; 53) CATH LAB, Model Number:DYNJ85511A; 54) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJT1520K; 55) CVC BUNDLE KIT, Model Number:ECVC6225A; 56) CVC 3L 7F 20CM VANTEX, Model Number:ECVC7265A; 57) ANGIOGRAPHY TRAY, Model Number:MNS9130

Code information (models, lots, serial numbers)

Show the full code information (25k characters)

1) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 24IBQ637; 2) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 24KBF279; 3) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 24LBF132; 4) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 25CBJ807; 5) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 25DBO471; 6) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 25GBK506; 7) CDS985094F, UDI-DI: 10195327657642(each), 40195327657643(case), Lot Number: 25AMG933; 8) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 22KBQ113; 9) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23ABU376; 10) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23DBD193; 11) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23GBT819; 12) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23HBQ238; 13) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23IBR569; 14) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23KBH770; 15) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23LBO849; 16) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24ABW181; 17) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24EBM208; 18) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24GBA239; 19) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24IBE619; 20) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24JBA719; 21) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24KBJ312; 22) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24LBM892; 23) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25BBK679; 24) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25CBB540; 25) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25DBA780; 26) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25FBR530; 27) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25GBQ911; 28) DYKMBNDL153, UDI-DI: 10193489478440(each), 40193489478441(case), Lot Number: 21ABI976; 29) DYKMBNDL153, UDI-DI: 10193489478440(each), 40193489478441(case), Lot Number: 21BBG516; 30) DYKMBNDL153, UDI-DI: 10193489478440(each), 40193489478441(case), Lot Number: 21LBJ748; 31) DYKMBNDL153, UDI-DI: 10193489478440(each), 40193489478441(case), Lot Number: 22ABO093; 32) DYKMBNDL4A, UDI-DI: 10193489904512(each), 40193489904513(case), Lot Number: 21GLA287; 33) DYKMBNDL4A, UDI-DI: 10193489904512(each), 40193489904513(case), Lot Number: 21GLA930; 34) DYKMBNDL4A, UDI-DI: 10193489904512(each), 40193489904513(case), Lot Number: 21KLA100; 35) DYKMBNDL4A, UDI-DI: 10193489904512(each), 40193489904513(case), Lot Number: 22ALA279; 36) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 22HLB260; 37) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23BLA583; 38) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23GLA176; 39) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23GLA710; 40) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23HLA316; 41) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23ILA268; 42) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23JLA979; 43) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23KLA402; 44) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 24CLA777; 45) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 24FLA816; 46) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 25ALA014; 47) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 25ELA738; 48) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 25HLA757; 49) DYNDA2076C, UDI-DI: 10198459001208(each), 40198459001209(case), Lot Number: 24HMD928; 50) DYNDA2076C, UDI-DI: 10198459001208(each), 40198459001209(case), Lot Number: 24KMC111; 51) DYNDA2076C, UDI-DI: 10198459001208(each), 40198459001209(case), Lot Number: 24LMB205; 52) DYNJ0150605AC, UDI-DI: 10198459106422(each), 40198459106423(case), Lot Number: 24IMB736; 53) DYNJ0368386D, UDI-DI: 10889942060245(each), 40889942060246(case), Lot Number: 23BMB501; 54) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 24GBT979; 55) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 24HBQ060; 56) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 24IBS786; 57) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 24LBE715; 58) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 25ABH209; 59) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 25BBT175; 60) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 25CBO084; 61) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 25EBR586; 62) DYNJ0803367I, UDI-DI: 10195327461027(each), 40195327461028(case), Lot Number: 23IMC122; 63) DYNJ0803367I, UDI-DI: 10195327461027(each), 40195327461028(case), Lot Number: 23JMA137; 64) DYNJ0928124C, UDI-DI: 10193489484427(each), 40193489484428(case), Lot Number: 23JLB288; 65) DYNJ0954970I, UDI-DI: 10198459297571(each), 40198459297572(case), Lot Number: 25FMJ515; 66) DYNJ16803D, UDI-DI: 10195327503710(each), 40195327503711(case), Lot Number: 23JMC203; 67) DYNJ19946M, UDI-DI: 10193489907933(each), 40193489907934(case), Lot Number: 23HBO857; 68) DYNJ30955I, UDI-DI: 10195327428051(each), 40195327428052(case), Lot Number: 23GBN513; 69) DYNJ31773K, UDI-DI: 10195327245856(each), 40195327245857(case), Lot Number: 23ABB938; 70) DYNJ33061D, UDI-DI: 10193489841428(each), 40193489841429(case), Lot Number: 21CBI315; 71) DYNJ33620D, UDI-DI: 10889942677832(each), 40889942677833(case), Lot Number: 21ADA464; 72) DYNJ35300M, UDI-DI: 10195327101169(each), 40195327101160(case), Lot Number: 22JMB302; 73) DYNJ35643C, UDI-DI: 10195327111946(each), 40195327111947(case), Lot Number: 23CBI301; 74) DYNJ39066B, UDI-DI: 10193489937268(each), 40193489937269(case), Lot Number: 23HMG433; 75) DYNJ42134G, UDI-DI: 10195327644581(each), 40195327644582(case), Lot Number: 24IBS009; 76) DYNJ43101C, UDI-DI: 10888277737327(each), 40888277737328(case), Lot Number: 21ADA535; 77) DYNJ43415D, UDI-DI: 10195327512873(each), 40195327512874(case), Lot Number: 23IBP533; 78) DYNJ43415D, UDI-DI: 10195327512873(each), 40195327512874(case), Lot Number: 24ABI910; 79) DYNJ43415D, UDI-DI: 10195327512873(each), 40195327512874(case), Lot Number: 24CBG256; 80) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24GBV382; 81) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24HBM254; 82) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24IBU874; 83) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24KBR090; 84) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24LBQ306; 85) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 25EBF775; 86) DYNJ44700J, UDI-DI: 10195327277413(each), 40195327277414(case), Lot Number: 25JDA679; 87) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 21KBO116; 88) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22ABI539; 89) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22ABK883; 90) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22CBV346; 91) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22EBP030; 92) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22FBN577; 93) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22HBM547; 94) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22IBT510; 95) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22JBC588; 96) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22JBY064; 97) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22LBG760; 98) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22OBA581; 99) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23ABR589; 100) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23BBD736; 101) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23DBI845; 102) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23EBG893; 103) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23FBF258; 104) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23GBP204; 105) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23IBB504; 106) DYNJ44889O, UDI-DI: 10193489979411(each), 40193489979412(case), Lot Number: 22AME756; 107) DYNJ44889O, UDI-DI: 10193489979411(each), 40193489979412(case), Lot Number: 22CMA383; 108) DYNJ47370B, UDI-DI: 10889942975723(each), 40889942975724(case), Lot Number: 22LBP706; 109) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22GMI134; 110) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22IMG735; 111) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22JME805; 112) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22KMC442; 113) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22LMD047; 114) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23BMC700; 115) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23CMB279; 116) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23DMF448; 117) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23EME724; 118) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23FME772; 119) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23GMI002; 120) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23HMH791; 121) DYNJ56772G, UDI-DI: 10195327054069(each), 40195327054060(case), Lot Number: 22AMD580; 122) DYNJ56772G, UDI-DI: 10195327054069(each), 40195327054060(case), Lot Number: 22BMG182; 123) DYNJ56772G, UDI-DI: 10195327054069(each), 40195327054060(case), Lot Number: 22CMH914; 124) DYNJ57602G, UDI-DI: 10195327338169(each), 40195327338160(case), Lot Number: 23BBK044; 125) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23GBE283; 126) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23GBP181; 127) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23JBB692; 128) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23JBX727; 129) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23KBB465; 130) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24ABP912; 131) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24ABS524; 132) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24ABT394; 133) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24CBC189; 134) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24CBO358; 135) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24DBF723; 136) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24EBH891; 137) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24FBH066; 138) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24GBA907; 139) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24GBQ359; 140) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24JBB314; 141) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24JBC107; 142) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24KBN562; 143) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24LBJ716; 144) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 25BBA430; 145) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 25DBH293; 146) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 25EBL462; 147) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 25FBK782; 148) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23FMF145; 149) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 22LMA967; 150) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23AMI769; 151) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23BME160; 152) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23DMI350; 153) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23FME284; 154) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23GMF491; 155) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23IMD765; 156) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23KMH102; 157) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23LMD057; 158) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24CMF744; 159) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24CMG318; 160) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24DMC509; 161) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24DMG251; 162) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24EMH576; 163) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24GMH576; 164) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24HMF813; 165) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24JMJ426; 166) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24KMD474; 167) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25AMG240; 168) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25CMD739; 169) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25DMK093; 170) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25FMC822; 171) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25GMF770; 172) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25HMD080; 173) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25JME465; 174) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25KMD960; 175) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22BBX590; 176) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22DBM388; 177) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22DBS611; 178) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22FBU003; 179) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22HBI355; 180) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22KBH299; 181) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22LBA917; 182) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 23ABL095; 183) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 23CBU118; 184) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 23EBN753; 185) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 23GBC772; 186) DYNJ67378C, UDI-DI: 10195327697501(each), 40195327697502(case), Lot Number: 24HME465; 187) DYNJ67918A, UDI-DI: 10195327042776(each), 40195327042777(case), Lot Number: 21LBP804; 188) DYNJ67918A, UDI-DI: 10195327042776(each), 40195327042777(case), Lot Number: 22CBZ026; 189) DYNJ67918A, UDI-DI: 10195327042776(each), 40195327042777(case), Lot Number: 22CBZ027; 190) DYNJ67918A, UDI-DI: 10195327042776(each), 40195327042777(case), Lot Number: 22OBG779; 191) DYNJ68633, UDI-DI: 10193489837667(each), 40193489837668(case), Lot Number: 21BBQ520; 192) DYNJ81197B, UDI-DI: 10195327427818(each), 40195327427819(case), Lot Number: 23FBA862; 193) DYNJ81197B, UDI-DI: 10195327427818(each), 40195327427819(case), Lot Number: 23GBC445; 194) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23GBJ036; 195) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23HBZ547; 196) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23IBT017; 197) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23JBF780; 198) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23JBR165; 199) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24ABW070; 200) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24DBF346; 201) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24EBU925; 202) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24GBA095; 203) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24HBK118; 204) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24IBL386; 205) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24JBE696; 206) DYNJ81197F, UDI-DI: 10198459225000(each), 40198459225001(case), Lot Number: 24LBU027; 207) DYNJ81197F, UDI-DI: 10198459225000(each), 40198459225001(case), Lot Number: 25CBG703; 208) DYNJ81197F, UDI-DI: 10198459225000(each), 40198459225001(case), Lot Number: 25DBT193; 209) DYNJ81197F, UDI-DI: 10198459225000(each), 40198459225001(case), Lot Number: 25EBR556; 210) DYNJ81197G, UDI-DI: 10198459369179(each), 40198459369170(case), Lot Number: 25FBT968; 211) DYNJ81197G, UDI-DI: 10198459369179(each), 40198459369170(case), Lot Number: 25HBT944; 212) DYNJ81197G, UDI-DI: 10198459369179(each), 40198459369170(case), Lot Number: 25LBF577; 213) DYNJ81197G, UDI-DI: 10198459369179(each), 40198459369170(case), Lot Number: 25KBE898; 214) DYNJ81197I, UDI-DI: 10198459640056(each), 40198459640057(case), Lot Number: 26ABG934; 215) DYNJ81593B, UDI-DI: 10195327405298(each), 40195327405299(case), Lot Number: 23EBT723; 216) DYNJ81593B, UDI-DI: 10195327405298(each), 40195327405299(case), Lot Number: 23GBO156; 217) DYNJ81593B, UDI-DI: 10195327405298(each), 40195327405299(case), Lot Number: 23IBJ841; 218) DYNJ83073, UDI-DI: 10195327234645(each), 40195327234646(case), Lot Number: 23DMI662; 219) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 22JMI325; 220) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23BMB766; 221) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23DMA913; 222) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23DMD375; 223) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23EMF492; 224) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23FMC409; 225) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23GMD133; 226) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23IMB693; 227) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23CBS129; 228) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23EBU525; 229) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23FBJ743; 230) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23FBV261; 231) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23HBX967; 232) DYNJ83981, UDI-DI: 10195327324124(each), 40195327324125(case), Lot Number: 23CMI086; 233) DYNJ84539, UDI-DI: 10195327375706(each), 40195327375707(case), Lot Number: 23FMH066; 234) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 23JBP557; 235) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 23JBR927; 236) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24ABR587; 237) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24BBN470; 238) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24CBV309; 239) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24DBS240; 240) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24EBN545; 241) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24FBB518; 242) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24GBK818; 243) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24HBG260; 244) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24IBI836; 245) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24KBR659; 246) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24LBK715; 247) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25BBE395; 248) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25BBP468; 249) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25CBR179; 250) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25EBB233; 251) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25EBR907; 252) DYNJ85511A, UDI-DI: 10198459413735(each), 40198459413736(case), Lot Number: 25GBN456; 253) DYNJ85511A, UDI-DI: 10198459413735(each), 40198459413736(case), Lot Number: 25HBM139; 254) DYNJT1520K, UDI-DI: 10195327436261(each), 40195327436262(case), Lot Number: 23JLA175; 255) ECVC6225A, UDI-DI: 10653160333418(each), 653160333411(case), Lot Number: 2025110690; 256) ECVC7265A, UDI-DI: 10653160345145(each), 653160345148(case), Lot Number: 2023071090; 257) ECVC7265A, UDI-DI: 10653160345145(each), 653160345148(case), Lot Number: 2024021590; 258) ECVC7265A, UDI-DI: 10653160345145(each), 653160345148(case), Lot Number: 2024042390; 259) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 22GBC976; 260) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 22HBO815; 261) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 22JBU760; 262) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 22LBK966; 263) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23ABR642; 264) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23CBI570; 265) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23GBG640; 266) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23JBN131; 267) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23LBC300; 268) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24ABY158; 269) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24DBC089; 270) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24FBF351; 271) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24FBT379; 272) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24IBF858; 273) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24JBW207; 274) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25ABS583; 275) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25DBF439; 276) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25EBA128; 277) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25FBI982; 278) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25HBP222

Devices named in this recall