Medline Convenience Kits recallZ-2110-2026 · Medline Industries, LP · initiated Jan 7, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA root cause category: Process control
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/14/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-
Recall details
- Recall number
- Z-2110-2026
- Status
- Open, Classified
- Initiated by the company
- Jan 7, 2026
- Posted by the FDA
- May 7, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 12529 units
- Product code
- OHM
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Medline Convenience Kits: 1) C-SECTION PACK, Model Number: CDS820080N; 2) C-SECTION CDS, Model Number: CDS983523G; 3) C-SECTION, Model Number: CDSCCR611O; 4) C-SECTION, Model Number: CDSCCR611P; 5) LUD C-SECTION-LF, Model Number: DYNJ0101616I; 6) C-SECTION TRAY, Model Number: DYNJ15494G; 7) C-SECTION TRAY-LF, Model Number: DYNJ19529O; 8) C-SECTION TRAY-LF, Model Number: DYNJ19529P; 9) C-SECTION TRAY-LF, Model Number: DYNJ19529Q; 10) C-SECTION TRAY-LF, Model Number: DYNJ19529R; 11) C-SECTION TRAY-LF, Model Number: DYNJ19529S; 12) C-SECTION PACK-LF, Model Number: DYNJ21550Q; 13) C-SECTION PACK, Model Number: DYNJ23544F; 14) C-SECTION PACK, Model Number: DYNJ32632F; 15) C-SECTION PACK QVH-LF, Model Number: DYNJ35100I; 16) C-SECTION PACK, Model Number: DYNJ35902F; 17) C-SECTION PACK-LF, Model Number: DYNJ38801F; 18) SCRIPPS SW C-SECTION PACK-LF, Model Number: DYNJ40129C; 19) C-SECTION, Model Number: DYNJ41444F; 20) C-SECTION PACK, Model Number: DYNJ49224I; 21) C-SECTION PACK, Model Number: DYNJ49771A; 22) C-SECTION PACK, Model Number: DYNJ49771B; 23) C SECTION PACK-LF, Model Number: DYNJ52387I; 24) C-SECTION PACK, Model Number: DYNJ53528B; 25) C SECTION PACK-LF, Model Number: DYNJ55367A; 26) C-SECTION PACK, Model Number: DYNJ56523C; 27) C-SECTION PACK, Model Number: DYNJ59387L; 28) C-SECTION PACK, Model Number: DYNJ59387M; 29) C-SECTION PACK, Model Number: DYNJ59387N; 30) C-SECTION PACK, Model Number: DYNJ59387O; 31) C-SECTION PACK, Model Number: DYNJ59387P; 32) C-SECTION TRAY, Model Number: DYNJ62707B; 33) C-SECTION PACK, Model Number: DYNJ64549B; 34) C SECTION PACK, Model Number: DYNJ67347A; 35) C SECTION PACK, Model Number: DYNJ69778A; 36) B-C SECTION BAYSTATE, Model Number: DYNJ900602I; 37) B-C SECTION BAYSTATE, Model Number: DYNJ900602K; 38) PAD C-SECTION, Model Number: DYNJ902134C; 39) RICH C-SECTION, Model Number: DYNJ902182J; 40) C-SECTION, Model Number: DYNJ902938J; 41) C-SECTION PACK, Model Number: DYNJ904329D; 42) OB DELIVERY KIT, Model Number: DYNJ904453C; 43) C-SECTION, Model Number: DYNJ904697F; 44) C-SECTION, Model Number: DYNJ906255B; 45) KIT C SECTION, Model Number: DYNJ906966G; 46) C SECTION, Model Number: DYNJ907999A; 47) KIT C SECTION OR, Model Number: DYNJ908350C; 48) C-SECTION, Model Number: DYNJ909058; 49) C SECTION, Model Number: DYNJ909475A; 50) C-SECTION PACK, Model Number: DYNJ909507; 51) C-SECTION, Model Number: DYNJ910199; 52) C-SECTION, Model Number: DYNJ910199A; 53) C SECTION TRAY, Model Number: DYNJT3407; 54) C-SECTION BASIN PACK, Model Number: OBP211C; 55) TR-C SECTION PACK-LF, Model Number: PHS583549002B
Code information (models, lots, serial numbers)
Show the full code information (14k characters)
1) CDS820080N, UDI-DI: 10195327278014(each), 40195327278015(case), Lot Number: 23IMB818; 2) CDS983523G, UDI-DI: 10195327283681(each), 40195327283682(case), Lot Number: 23IBS984; 3) CDSCCR611O, UDI-DI: 10193489367904(each), 40193489367905(case), Lot Number: 21CBN328; 4) CDSCCR611O, UDI-DI: 10193489367904(each), 40193489367905(case), Lot Number: 21LBC555; 5) CDSCCR611O, UDI-DI: 10193489367904(each), 40193489367905(case), Lot Number: 21WBA953; 6) CDSCCR611P, UDI-DI: 10195327065850(each), 40195327065851(case), Lot Number: 22BBH774; 7) CDSCCR611P, UDI-DI: 10195327065850(each), 40195327065851(case), Lot Number: 22BBT090; 8) CDSCCR611P, UDI-DI: 10195327065850(each), 40195327065851(case), Lot Number: 22EBA155; 9) CDSCCR611P, UDI-DI: 10195327065850(each), 40195327065851(case), Lot Number: 22FBU296; 10) CDSCCR611P, UDI-DI: 10195327065850(each), 40195327065851(case), Lot Number: 22FBU297; 11) CDSCCR611P, UDI-DI: 10195327065850(each), 40195327065851(case), Lot Number: 22FBU298; 12) CDSCCR611P, UDI-DI: 10195327065850(each), 40195327065851(case), Lot Number: 22IBG757; 13) DYNJ0101616I, UDI-DI: 10193489248814(each), 40193489248815(case), Lot Number: 21ADA105; 14) DYNJ15494G, UDI-DI: 10193489961195(each), 40193489961196(case), Lot Number: 22LBS122; 15) DYNJ19529O, UDI-DI: 10195327297206(each), 40195327297207(case), Lot Number: 23CMB465; 16) DYNJ19529O, UDI-DI: 10195327297206(each), 40195327297207(case), Lot Number: 23DMH505; 17) DYNJ19529O, UDI-DI: 10195327297206(each), 40195327297207(case), Lot Number: 23FMF781; 18) DYNJ19529O, UDI-DI: 10195327297206(each), 40195327297207(case), Lot Number: 23IMB944; 19) DYNJ19529O, UDI-DI: 10195327297206(each), 40195327297207(case), Lot Number: 23IME367; 20) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 23JMD920; 21) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 23JMI854; 22) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 23LME909; 23) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 24AMD085; 24) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 24AMI385; 25) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 24BMG152; 26) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 24CMJ610; 27) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 24EMB123; 28) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 24EMD740; 29) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 24FMB372; 30) DYNJ19529P, UDI-DI: 10195327529338(each), 40195327529339(case), Lot Number: 24GMF054; 31) DYNJ19529Q, UDI-DI: 10198459109362(each), 40198459109363(case), Lot Number: 24KMD458; 32) DYNJ19529Q, UDI-DI: 10198459109362(each), 40198459109363(case), Lot Number: 24LMH886; 33) DYNJ19529Q, UDI-DI: 10198459109362(each), 40198459109363(case), Lot Number: 25BMB041; 34) DYNJ19529R, UDI-DI: 10198459311031(each), 40198459311032(case), Lot Number: 25DMA214; 35) DYNJ19529R, UDI-DI: 10198459311031(each), 40198459311032(case), Lot Number: 25EMH398; 36) DYNJ19529R, UDI-DI: 10198459311031(each), 40198459311032(case), Lot Number: 25FMI427; 37) DYNJ19529R, UDI-DI: 10198459311031(each), 40198459311032(case), Lot Number: 25GMH469; 38) DYNJ19529R, UDI-DI: 10198459311031(each), 40198459311032(case), Lot Number: 25IMA090; 39) DYNJ19529S, UDI-DI: 10198459566257(each), 40198459566258(case), Lot Number: 25KMA045; 40) DYNJ21550Q, UDI-DI: 10193489830910(each), 40193489830911(case), Lot Number: 21BBM422; 41) DYNJ21550Q, UDI-DI: 10193489830910(each), 40193489830911(case), Lot Number: 21DBJ153; 42) DYNJ21550Q, UDI-DI: 10193489830910(each), 40193489830911(case), Lot Number: 21DBL994; 43) DYNJ23544F, UDI-DI: 10889942533053(each), 40889942533054(case), Lot Number: 23IMA860; 44) DYNJ32632F, UDI-DI: 10193489888966(each), 40193489888967(case), Lot Number: 22HMC526; 45) DYNJ35100I, UDI-DI: 10195327097608(each), 40195327097609(case), Lot Number: 23IME105; 46) DYNJ35902F, UDI-DI: 10193489920109(each), 40193489920100(case), Lot Number: 23ABF399; 47) DYNJ38801F, UDI-DI: 10195327113117(each), 40195327113118(case), Lot Number: 23IBK585; 48) DYNJ40129C, UDI-DI: 10193489441772(each), 40193489441773(case), Lot Number: 23AMJ303; 49) DYNJ41444F, UDI-DI: 10193489740974(each), 40193489740975(case), Lot Number: 22JMB590; 50) DYNJ49224I, UDI-DI: 10193489371352(each), 40193489371353(case), Lot Number: 21DDA358; 51) DYNJ49771A, UDI-DI: 10198459015717(each), 40198459015718(case), Lot Number: 24GBN350; 52) DYNJ49771B, UDI-DI: 10198459087776(each), 40198459087777(case), Lot Number: 24IBI305; 53) DYNJ52387I, UDI-DI: 10193489248272(each), 40193489248273(case), Lot Number: 21ADA540; 54) DYNJ53528B, UDI-DI: 10195327077303(each), 40195327077304(case), Lot Number: 23GBT514; 55) DYNJ55367A, UDI-DI: 10889942858903(each), 40889942858904(case), Lot Number: 24CBS240; 56) DYNJ56523C, UDI-DI: 10193489430264(each), 40193489430265(case), Lot Number: 21ABS513; 57) DYNJ59387L, UDI-DI: 10195327159856(each), 40195327159857(case), Lot Number: 23CBK420; 58) DYNJ59387M, UDI-DI: 10195327442064(each), 40195327442065(case), Lot Number: 23HBP931; 59) DYNJ59387N, UDI-DI: 10195327492557(each), 40195327492558(case), Lot Number: 23JBO439; 60) DYNJ59387O, UDI-DI: 10195327590079(each), 40195327590070(case), Lot Number: 24ABC320; 61) DYNJ59387O, UDI-DI: 10195327590079(each), 40195327590070(case), Lot Number: 24FMJ621; 62) DYNJ59387O, UDI-DI: 10195327590079(each), 40195327590070(case), Lot Number: 24GMD352; 63) DYNJ59387O, UDI-DI: 10195327590079(each), 40195327590070(case), Lot Number: 24GME349; 64) DYNJ59387O, UDI-DI: 10195327590079(each), 40195327590070(case), Lot Number: 24GMJ147; 65) DYNJ59387O, UDI-DI: 10195327590079(each), 40195327590070(case), Lot Number: 24HMD358; 66) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 24HBU368; 67) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 24JBP235; 68) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 24KBE601; 69) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 24LMB875; 70) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 25AMH030; 71) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 25BMB382; 72) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 25CMG270; 73) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 25FMB960; 74) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 25FMG755; 75) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 25GMK050; 76) DYNJ59387P, UDI-DI: 10198459097447(each), 40198459097448(case), Lot Number: 25HMI033; 77) DYNJ62707B, UDI-DI: 10195327292140(each), 40195327292141(case), Lot Number: 25ABV465; 78) DYNJ64549B, UDI-DI: 10193489940725(each), 40193489940726(case), Lot Number: 23ABL355; 79) DYNJ67347A, UDI-DI: 10195327046156(each), 40195327046157(case), Lot Number: 22JMF954; 80) DYNJ69778A, UDI-DI: 10195327334420(each), 40195327334421(case), Lot Number: 23CBU899; 81) DYNJ900602I, UDI-DI: 10193489487121(each), 40193489487122(case), Lot Number: 20LDB302; 82) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 23ADB906; 83) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 23BDB272; 84) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 23DDA960; 85) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 23GDA465; 86) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 23HDB760; 87) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 23IDB372; 88) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 23JDC028; 89) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 23LDA543; 90) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24DDA865; 91) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24DDC210; 92) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24EDA956; 93) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24FDB598; 94) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24HDB497; 95) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24JDA466; 96) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24JDB666; 97) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24JDC302; 98) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 24LDA598; 99) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 25ADA998; 100) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 25BDB001; 101) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 25DDA885; 102) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 25FDA524; 103) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 25FDB748; 104) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 25GDB840; 105) DYNJ900602K, UDI-DI: 10193489937190(each), 40193489937191(case), Lot Number: 25JDA979; 106) DYNJ902134C, UDI-DI: 10193489452280(each), 40193489452281(case), Lot Number: 21DBP451; 107) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 21BBF750; 108) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 21CBQ743; 109) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 21CBS046; 110) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 21GBB842; 111) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 21IBL808; 112) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 21JBJ172; 113) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 21JBX738; 114) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 22ABY311; 115) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 22DBN664; 116) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 22EBG610; 117) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 22GBH232; 118) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 22HBF417; 119) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 22IBI254; 120) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 22JBO865; 121) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 22OBJ456; 122) DYNJ902182J, UDI-DI: 10193489452914(each), 40193489452915(case), Lot Number: 23ABM580; 123) DYNJ902938J, UDI-DI: 10193489946116(each), 40193489946117(case), Lot Number: 24CMH453; 124) DYNJ904329D, UDI-DI: 10198459221071(each), 40198459221072(case), Lot Number: 25EBJ650; 125) DYNJ904453C, UDI-DI: 10198459005725(each), 40198459005726(case), Lot Number: 25JBQ639; 126) DYNJ904697F, UDI-DI: 10195327169275(each), 40195327169276(case), Lot Number: 22FDA703; 127) DYNJ906255B, UDI-DI: 10198459196782(each), 40198459196783(case), Lot Number: 25BDA042; 128) DYNJ906255B, UDI-DI: 10198459196782(each), 40198459196783(case), Lot Number: 25DDB842; 129) DYNJ906255B, UDI-DI: 10198459196782(each), 40198459196783(case), Lot Number: 25EDB771; 130) DYNJ906966G, UDI-DI: 10195327557959(each), 40195327557950(case), Lot Number: 24BBV274; 131) DYNJ906966G, UDI-DI: 10195327557959(each), 40195327557950(case), Lot Number: 24DBN053; 132) DYNJ906966G, UDI-DI: 10195327557959(each), 40195327557950(case), Lot Number: 24GBR695; 133) DYNJ906966G, UDI-DI: 10195327557959(each), 40195327557950(case), Lot Number: 24GBV882; 134) DYNJ906966G, UDI-DI: 10195327557959(each), 40195327557950(case), Lot Number: 24IBF270; 135) DYNJ906966G, UDI-DI: 10195327557959(each), 40195327557950(case), Lot Number: 24IBS185; 136) DYNJ907999A, UDI-DI: 10193489996814(each), 40193489996815(case), Lot Number: 21JMC590; 137) DYNJ907999A, UDI-DI: 10193489996814(each), 40193489996815(case), Lot Number: 21LMB616; 138) DYNJ907999A, UDI-DI: 10193489996814(each), 40193489996815(case), Lot Number: 22CMH174; 139) DYNJ908350C, UDI-DI: 10198459154478(each), 40198459154479(case), Lot Number: 24KBJ980; 140) DYNJ908350C, UDI-DI: 10198459154478(each), 40198459154479(case), Lot Number: 24LBT358; 141) DYNJ908350C, UDI-DI: 10198459154478(each), 40198459154479(case), Lot Number: 25ABI501; 142) DYNJ908350C, UDI-DI: 10198459154478(each), 40198459154479(case), Lot Number: 25CBL793; 143) DYNJ908350C, UDI-DI: 10198459154478(each), 40198459154479(case), Lot Number: 25DBP813; 144) DYNJ908350C, UDI-DI: 10198459154478(each), 40198459154479(case), Lot Number: 25FBW044; 145) DYNJ908350C, UDI-DI: 10198459154478(each), 40198459154479(case), Lot Number: 25HBO502; 146) DYNJ909058, UDI-DI: 10195327259327(each), 40195327259328(case), Lot Number: 23EBP308; 147) DYNJ909058, UDI-DI: 10195327259327(each), 40195327259328(case), Lot Number: 23HBF069; 148) DYNJ909058, UDI-DI: 10195327259327(each), 40195327259328(case), Lot Number: 23HBY877; 149) DYNJ909475A, UDI-DI: 10195327647155(each), 40195327647156(case), Lot Number: 25FMI114; 150) DYNJ909507, UDI-DI: 10195327387075(each), 40195327387076(case), Lot Number: 23GME519; 151) DYNJ910199, UDI-DI: 10195327612122(each), 40195327612123(case), Lot Number: 24DBM917; 152) DYNJ910199, UDI-DI: 10195327612122(each), 40195327612123(case), Lot Number: 24DBN447; 153) DYNJ910199A, UDI-DI: 10198459130656(each), 40198459130657(case), Lot Number: 24JBP901; 154) DYNJ910199A, UDI-DI: 10198459130656(each), 40198459130657(case), Lot Number: 24KBD622; 155) DYNJ910199A, UDI-DI: 10198459130656(each), 40198459130657(case), Lot Number: 25BBC088; 156) DYNJT3407, UDI-DI: 10198459324970(each), 40198459324971(case), Lot Number: 25EBK673; 157) OBP211C, UDI-DI: 10884389069745(each), 40884389069746(case), Lot Number: 25DBE903; 158) PHS583549002B, UDI-DI: 10193489643558(each), 40193489643559(case), Lot Number: 21ADA554
Devices named in this recall
54 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ32632FMedline Industries, Inc. · Cesarean section tray4 recalls0 reports
- Medline Industries, Inc. DYNJ904697FMedline Industries, Inc. · Cesarean section tray4 recalls0 reports
- Medline Industries, Inc. DYNJ15494GMedline Industries, Inc. · Cesarean section tray3 recalls0 reports
- Medline Industries, Inc. DYNJ23544FMedline Industries, Inc. · Cesarean section tray2 recalls1 report
- Medline Industries, Inc. DYNJ52387IMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0101616IMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. CDSCCR611OMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ49224IMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ56523CMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ40129CMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902134CMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902182JMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ900602IMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. PHS583549002BMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ41444FMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ21550QMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ900602KMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ64549BMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902938JMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ907999AMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ67347AMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. CDSCCR611PMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ35100IMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ38801FMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ59387LMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ909058Medline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. CDS820080NMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. CDS983523GMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ69778AMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ909507Medline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ19529PMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ906966GMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ19529QMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ904329DMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ19529RMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJT3407Medline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. OBP211CMedline Industries, Inc. · Cesarean section tray2 recalls0 reports
- Medline Industries, Inc. DYNJ908350CMedline Industries, Inc. · Cesarean section tray1 recall1 report
- Medline Industries, Inc. DYNJ35902FMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ53528BMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ62707BMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ19529OMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ59387MMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ59387NMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ59387OMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ910199Medline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ909475AMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ904453CMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ49771AMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ49771BMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ59387PMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ910199AMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ906255BMedline Industries, Inc. · Cesarean section tray1 recall0 reports
- Medline Industries, Inc. DYNJ55367AMedline Industries, Inc. · Cesarean section tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98642).
- Z-2089-2026 · Open, ClassifiedMedline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number:DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number:DYNJ905567A
- Z-2090-2026 · Open, ClassifiedMedline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number:DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number:DYNJ56436B; 3) SCC RF PAIN MGMT, Model Number:DYNJ56436C; 4) CPNB SETUP…
- Z-2091-2026 · Open, ClassifiedMedline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AA; 2) ENDO KIT W/SYRINGE, Model Number:DYKE1743; 3) BGMC ANGIOGRAPHY PACK-LF, Model Number:DYNJ0800934D;…
- Z-2092-2026 · Open, ClassifiedMedline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AB; 2) OR ARTERIOGRAM, Model Number:CDS985094F; 3) INTRAVASOR CATH KIT, Model Number:DYK1060193I; 4) KIT…
- Z-2093-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) OPEN HEART CDS, Model Number:CDS840015AQ; 2) OPEN HEART CDS, Model Number:CDS840015AQ; 3) OPEN HEART CDS, Model Number:CDS84…
- Z-2094-2026 · Open, ClassifiedMedline Convenience Kits: 1) PICC LINE TRAY, Model Number:00-401993O; 2) CODE CART ADULT 4 CENTRAL LINE, Model Number:ACC010268B; 3) CODE CART ADULT 4 CENTRAL LINE, Model Number:…
- Z-2095-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number:DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number:DY…
- Z-2096-2026 · Open, ClassifiedMedline Convenience Kits: 1) DYNDA3003, Model Number: DENTAL PACK; 2) DYNJ87157, Model Number: GULLO SURGICAL PACK
- Z-2097-2026 · Open, ClassifiedSee RES for full list. Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK…
- Z-2098-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, …
- Z-2099-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) BREAST-HERNIA-PORT CDS-LF, Model Number: CDS984853J; 3) KIT UROLOGY URE…
- Z-2100-2026 · Open, ClassifiedMedline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A; 3) LAP BASIN, Model Number: DYNJVB91109D
- Z-2101-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER 6-IV F…
- Z-2102-2026 · Open, ClassifiedMedline Convenience Kits: 1) CVC PACK, Model Number: CVI4500A; 2) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A; 3) TOTE CHANGE OSC HAND REPAIR, Model Number: DYKM1751A; 4…
- Z-2103-2026 · Open, ClassifiedMedline Convenience Kits: 1) CENTRAL LINE UNIVERSAL INSERTI, Model Number: CVI3780; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) KIT SURG ONC INSERT CENT.VENOU, Model Number:…
- Z-2105-2026 · Open, ClassifiedMedline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A
- Z-2107-2026 · Open, ClassifiedMedline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYK…
- Z-2108-2026 · Open, ClassifiedMedline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL…
- Z-2109-2026 · Open, ClassifiedSee full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, …
- Z-2111-2026 · Open, ClassifiedMedline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I; 2) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91A; 3) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ037380…
- Z-2112-2026 · Open, ClassifiedMedline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 3) TOT/ GYN LITHOTOMY CDS, Model…
- Z-2113-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104; 2) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104A; 3) KIT OPHTHALMOLOGY CORNEAL TR…
- Z-2114-2026 · Open, ClassifiedSee RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS98083…
- Z-2115-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55A; 3) TRACH TOTE, Model Number:DYNJ85691
- Z-2116-2026 · Open, ClassifiedMedline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number:PAIN0150C
- Z-2117-2026 · Open, ClassifiedMedline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721
- Z-2118-2026 · Open, ClassifiedMedline Convenience Kits: 1) PICC CATHETER INSERTION TRAY, Model Number: CVI4300; 2) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977C; 3) KIT ACES INSERT PEG, Model Number:…
- Z-2119-2026 · Open, ClassifiedMedline Convenience Kits: 1) WMC I D PACK-LF, Model Number: DYNJ50806L
- Z-2120-2026 · Open, ClassifiedMedline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437
- Z-2121-2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗