See full list in RES, exceeds character limit recallZ-2109-2026 · Medline Industries, LP · initiated Jan 7, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

FDA root cause category: Process control

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/14/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-

Recall details

Recall number
Z-2109-2026
Status
Open, Classified
Initiated by the company
Jan 7, 2026
Posted by the FDA
May 7, 2026
Recalling firm
Medline Industries, LP
Quantity
15806 units
Product code
OHD
Distribution
Worldwide - US Nationwide distribution.

Product description

See full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 4) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 5) NHP ROBOTIC GYNE CDS, Model Number: CDS984243G; 6) ROBOTIC-LF, Model Number: CDS984543D; 7) ROBOTIC-LF, Model Number: CDS984543F; 8) ROBOTIC-LF, Model Number: CDS984543G; 9) ROBOTIC-LF, Model Number: CDS984543I; 10) GYN CDS, Model Number: CDS984860F; 11) GYN LAP PELVI, Model Number: CDS985423G; 12) GYN KIT, Model Number: DYKM1657A; 13) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL100; 14) KIT ROBOTICS GYN, Model Number: DYKMBNDL109; 15) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116D; 16) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116F; 17) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119; 18) KIT GYN VAGINAL MAJOR, Model Number: DYKMBNDL119A; 19) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168; 20) KIT GYN ABDOMINAL MAJOR, Model Number: DYKMBNDL168A; 21) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184D; 22) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91;

Code information (models, lots, serial numbers)

Show the full code information (35k characters)

1) CDS981504G, UDI-DI: 10195327157968(each), 40195327157969(case), Lot Number: 25BDB744; 2) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23EBE997; 3) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23GBL921; 4) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23GBO574; 5) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23IBC607; 6) CDS982407P, UDI-DI: 10195327110116(each), 40195327110117(case), Lot Number: 23JBQ095; 7) CDS982893I, UDI-DI: 10889942208050(each), 40889942208051(case), Lot Number: 24IBO240; 8) CDS983300J, UDI-DI: 10193489425048(each), 40193489425049(case), Lot Number: 21CDB781; 9) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21EBA117; 10) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21GBI171; 11) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21HBN807; 12) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21IBM804; 13) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 21VBA272; 14) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22BBL450; 15) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22FBV462; 16) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22GBP114; 17) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22HBV491; 18) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22JBC547; 19) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22KBD152; 20) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 22OBM422; 21) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23ABC549; 22) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23CBN644; 23) CDS984243G, UDI-DI: 10193489872521(each), 40193489872522(case), Lot Number: 23CBX440; 24) CDS984543D, UDI-DI: 10195327072322(each), 40195327072323(case), Lot Number: 22CBC587; 25) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22GBD616; 26) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22HMB259; 27) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22HMG671; 28) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22JMA223; 29) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22KMD619; 30) CDS984543F, UDI-DI: 10195327115647(each), 40195327115648(case), Lot Number: 22OBL925; 31) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 24AMJ461; 32) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23BMD680; 33) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23CMC582; 34) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23DMA508; 35) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23EMI460; 36) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23FMI171; 37) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23GMF425; 38) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23GMH282; 39) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23IMA629; 40) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23JMC376; 41) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 23LMA864; 42) CDS984543G, UDI-DI: 10195327299637(each), 40195327299638(case), Lot Number: 24BMH483; 43) CDS984543I, UDI-DI: 10195327637071(each), 40195327637072(case), Lot Number: 24CMH050; 44) CDS984860F, UDI-DI: 10193489261974(each), 40193489261975(case), Lot Number: 21ADB205; 45) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21CBH282; 46) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21CBH547; 47) CDS985423G, UDI-DI: 10193489319323(each), 40193489319324(case), Lot Number: 21DBV243; 48) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25DBN805; 49) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25FBS108; 50) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25HBS386; 51) DYKM1657A, UDI-DI: 10198459297410(each), 40198459297411(case), Lot Number: 25LBN225; 52) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21ABI962; 53) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21BBG514; 54) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21EBA097; 55) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21GBP406; 56) DYKMBNDL100, UDI-DI: 10193489477955(each), 40193489477956(case), Lot Number: 21JBD206; 57) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABM086; 58) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABO060; 59) DYKMBNDL109, UDI-DI: 10193489478211(each), 40193489478212(case), Lot Number: 21ABW347; 60) DYKMBNDL116D, UDI-DI: 10195327608866(each), 40195327608867(case), Lot Number: 24CBF889; 61) DYKMBNDL116D, UDI-DI: 10195327608866(each), 40195327608867(case), Lot Number: 24DBA349; 62) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24DBS388; 63) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24FBC411; 64) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24GBK948; 65) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24IBV184; 66) DYKMBNDL116F, UDI-DI: 10198459000706(each), 40198459000707(case), Lot Number: 24JBB328; 67) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21ABJ213; 68) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21ABL615; 69) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21BBA024; 70) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21BBP181; 71) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21EBG218; 72) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21FBP762; 73) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21HBK337; 74) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21JBC127; 75) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21JBX803; 76) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21LBP646; 77) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 21LBY301; 78) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22BBR618; 79) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22CBX517; 80) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22FBI695; 81) DYKMBNDL119, UDI-DI: 10193489479317(each), 40193489479318(case), Lot Number: 22HBL832; 82) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 22JBC546; 83) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 23KBA328; 84) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 23LBE369; 85) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24ABX678; 86) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24CBJ465; 87) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24DBV608; 88) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24JBD270; 89) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24KBF533; 90) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 24KBI207; 91) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25FBR662; 92) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25GBX843; 93) DYKMBNDL119A, UDI-DI: 10195327241025(each), 40195327241026(case), Lot Number: 25HBB957; 94) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21ABI984; 95) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21BBG518; 96) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21EBK837; 97) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21GBP369; 98) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21GBU222; 99) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21IBL095; 100) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21JBJ225; 101) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 21KBQ606; 102) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22ABL283; 103) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22ABL299; 104) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22CBX518; 105) DYKMBNDL168, UDI-DI: 10193489477436(each), 40193489477437(case), Lot Number: 22HBP253; 106) DYKMBNDL168A, UDI-DI: 10195327240806(each), 40195327240807(case), Lot Number: 23FBJ346; 107) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24GBJ568; 108) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24GBV008; 109) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 24HBK278; 110) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25ABS671; 111) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25BBF014; 112) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25FBJ143; 113) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25HBC798; 114) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25LBG179; 115) DYKMBNDL184D, UDI-DI: 10198459000485(each), 40198459000486(case), Lot Number: 25LBQ333; 116) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21ABJ086; 117) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21BBG520; 118) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21FBP714; 119) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21HBK341; 120) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21IBO770; 121) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 21LBJ751; 122) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 22GBZ017; 123) DYKMBNDL91, UDI-DI: 10193489477870(each), 40193489477871(case), Lot Number: 22HBB985; 124) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ABJ861; 125) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21BLA496; 126) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21DLA017; 127) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA256; 128) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA352; 129) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21ELA658; 130) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21HLA598; 131) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21JLA896; 132) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21LLA163; 133) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 21LLA284; 134) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ALA125; 135) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ALA331; 136) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22BLA918; 137) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22CLA119; 138) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22DLB036; 139) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ELA227; 140) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22ELB163; 141) DYKMBNDL92, UDI-DI: 10193489477351(each), 40193489477352(case), Lot Number: 22HLA664; 142) DYKMBNDL92A, UDI-DI: 10195327241209(each), 40195327241200(case), Lot Number: 22JLA665; 143) DYKMBNDL92A, UDI-DI: 10195327241209(each), 40195327241200(case), Lot Number: 22KLA731; 144) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24HLB142; 145) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24HLB219; 146) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24JLA881; 147) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24JLB140; 148) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLA194; 149) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLA506; 150) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 24KLB092; 151) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25BLA797; 152) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25DLB036; 153) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25ELA660; 154) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25GLA701; 155) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25HLA275; 156) DYKMBNDL92C, UDI-DI: 10198459000645(each), 40198459000646(case), Lot Number: 25ILA134; 157) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 20LBY985; 158) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21ABM148; 159) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21CBB287; 160) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21CBS045; 161) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21GBB013; 162) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21HBO271; 163) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 21IBV704; 164) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 22FBK394; 165) DYKMBNDL94, UDI-DI: 10193489477894(each), 40193489477895(case), Lot Number: 22HBH516; 166) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 22LBS837; 167) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABJ591; 168) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABK219; 169) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23ABO813; 170) DYKMBNDL94A, UDI-DI: 10195327241216(each), 40195327241217(case), Lot Number: 23BBI491; 171) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24DBR903; 172) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24EBV430; 173) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24IBE086; 174) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24KBA746; 175) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 24LBN994; 176) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25ABH279; 177) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25ABN371; 178) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25BBA805; 179) DYKMBNDL94C, UDI-DI: 10198459000669(each), 40198459000660(case), Lot Number: 25CBR621; 180) DYKMBNDL94D, UDI-DI: 10198459350641(each), 40198459350642(case), Lot Number: 25EBN175; 181) DYKMBNDL94F, UDI-DI: 10198459468926(each), 40198459468927(case), Lot Number: 25HBT315; 182) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24ABI719; 183) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24BBH633; 184) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24CBM999; 185) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24DBL074; 186) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24EBM495; 187) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24EBT021; 188) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24HBA113; 189) DYNJ0281948N, UDI-DI: 10195327600259(each), 40195327600250(case), Lot Number: 24HBX130; 190) DYNJ0347372F, UDI-DI: 10195327213176(each), 40195327213177(case), Lot Number: 22GBR392; 191) DYNJ0347372G, UDI-DI: 10195327266028(each), 40195327266029(case), Lot Number: 22KBU586; 192) DYNJ0347372G, UDI-DI: 10195327266028(each), 40195327266029(case), Lot Number: 23DBP818; 193) DYNJ0347372I, UDI-DI: 10195327418335(each), 40195327418336(case), Lot Number: 23GBO695; 194) DYNJ0347372I, UDI-DI: 10195327418335(each), 40195327418336(case), Lot Number: 23IBK088; 195) DYNJ0384702T, UDI-DI: 10195327049683(each), 40195327049684(case), Lot Number: 23BBN374; 196) DYNJ28252G, UDI-DI: 10198459088124(each), 40198459088125(case), Lot Number: 25BBP572; 197) DYNJ28252G, UDI-DI: 10198459088124(each), 40198459088125(case), Lot Number: 25CBC137; 198) DYNJ30005G, UDI-DI: 10193489601794(each), 40193489601795(case), Lot Number: 21DMA051; 199) DYNJ32706B, UDI-DI: 10195327364304(each), 40195327364305(case), Lot Number: 23EMB628; 200) DYNJ38844N, UDI-DI: 10195327243166(each), 40195327243167(case), Lot Number: 23DBI585; 201) DYNJ38844N, UDI-DI: 10195327243166(each), 40195327243167(case), Lot Number: 23HBI619; 202) DYNJ47550C, UDI-DI: 10195327209964(each), 40195327209965(case), Lot Number: 23DBD093; 203) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23BBC881; 204) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23CBR808; 205) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23FBH830; 206) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23FBU806; 207) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23HBO785; 208) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23IBQ207; 209) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23JBN605; 210) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 23KBQ430; 211) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24CBQ847; 212) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24FBG329; 213) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24HBV800; 214) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24IBE460; 215) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24JBC084; 216) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 24KBM397; 217) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25CBA290; 218) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25DBE603; 219) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25DBE964; 220) DYNJ47714A, UDI-DI: 10193489648867(each), 40193489648868(case), Lot Number: 25KBF492; 221) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21AMA809; 222) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21GMD682; 223) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21JMF391; 224) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21JMH124; 225) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21JMH124A; 226) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 21KMG274; 227) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22BMH522; 228) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22EMA211; 229) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22HMA973; 230) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22HMC351; 231) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 22KMH193; 232) DYNJ49191B, UDI-DI: 10193489497489(each), 40193489497480(case), Lot Number: 23CMB495; 233) DYNJ49705K, UDI-DI: 10195327020347(each), 40195327020348(case), Lot Number: 22IMI306; 234) DYNJ50348F, UDI-DI: 10195327406349(each), 40195327406340(case), Lot Number: 23IMD230; 235) DYNJ50439F, UDI-DI: 10193489704921(each), 40193489704922(case), Lot Number: 22EBC737; 236) DYNJ51795F, UDI-DI: 10193489475937(each), 40193489475938(case), Lot Number: 20LDC148; 237) DYNJ51795F, UDI-DI: 10193489475937(each), 40193489475938(case), Lot Number: 21ADA184; 238) DYNJ52569B, UDI-DI: 10193489685190(each), 40193489685191(case), Lot Number: 23EBS784; 239) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21CBU815; 240) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21IBG462; 241) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 21JBR818; 242) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22ABC476; 243) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22GBJ776; 244) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22HBV228; 245) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 22LBD784; 246) DYNJ55862, UDI-DI: 10889942583997(each), 40889942583998(case), Lot Number: 23BBA462; 247) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21EMB589; 248) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21FMB889; 249) DYNJ57607A, UDI-DI: 10193489206678(each), 40193489206679(case), Lot Number: 21GME150; 250) DYNJ58321A, UDI-DI: 10193489489316(each), 40193489489317(case), Lot Number: 22ABG197; 251) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21DMB839; 252) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21EMF383; 253) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21HMD002; 254) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 21JMF224; 255) DYNJ63079B, UDI-DI: 10193489425536(each), 40193489425537(case), Lot Number: 22AME373; 256) DYNJ63079C, UDI-DI: 10195327102180(each), 40195327102181(case), Lot Number: 23EMI624; 257) DYNJ63079C, UDI-DI: 10195327102180(each), 40195327102181(case), Lot Number: 23GMB716; 258) DYNJ63079D, UDI-DI: 10195327505677(each), 40195327505678(case), Lot Number: 23JMB435; 259) DYNJ63079D, UDI-DI: 10195327505677(each), 40195327505678(case), Lot Number: 24BME368; 260) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24HMC603; 261) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24HMH481; 262) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 24JMD681; 263) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25AMH256; 264) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25JMH582; 265) DYNJ63079F, UDI-DI: 10198459043642(each), 40198459043643(case), Lot Number: 25JMK630; 266) DYNJ64266, UDI-DI: 10193489795516(each), 40193489795517(case), Lot Number: 21ADA447; 267) DYNJ66298D, UDI-DI: 10195327095895(each), 40195327095896(case), Lot Number: 23IBW233; 268) DYNJ67269, UDI-DI: 10193489394696(each), 40193489394697(case), Lot Number: 22KBG504; 269) DYNJ68710, UDI-DI: 10193489842906(each), 40193489842907(case), Lot Number: 22ABM641; 270) DYNJ69419A, UDI-DI: 10195327223588(each), 40195327223589(case), Lot Number: 23ABS144; 271) DYNJ69419A, UDI-DI: 10195327223588(each), 40195327223589(case), Lot Number: 23DBK434; 272) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24EMC427; 273) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24EMD677; 274) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24GMC135; 275) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24GMG805; 276) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24KME386; 277) DYNJ69419D, UDI-DI: 10195327688844(each), 40195327688845(case), Lot Number: 24KMH362; 278) DYNJ81492A, UDI-DI: 10195327348830(each), 40195327348831(case), Lot Number: 23BBT116; 279) DYNJ81761A, UDI-DI: 10195327574468(each), 40195327574469(case), Lot Number: 24FMF687; 280) DYNJ83475A, UDI-DI: 10195327460938(each), 40195327460939(case), Lot Number: 23HBY488; 281) DYNJ84783, UDI-DI: 10195327417949(each), 40195327417940(case), Lot Number: 23IBU517; 282) DYNJ85342, UDI-DI: 10195327491437(each), 40195327491438(case), Lot Number: 23KMF007; 283) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 24EBD237; 284) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 25ABP819; 285) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 25ABT160; 286) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 25ABT161; 287) DYNJ86195A, UDI-DI: 10195327682286(each), 40195327682287(case), Lot Number: 25GBA444; 288) DYNJ900262A, UDI-DI: 10195327477820(each), 40195327477821(case), Lot Number: 23KBU736; 289) DYNJ901046L, UDI-DI: 10195327458614(each), 40195327458615(case), Lot Number: 23KBF626; 290) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21JBM245; 291) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21JBV891; 292) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21KBF079; 293) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 21KBP568; 294) DYNJ901133A, UDI-DI: 10888277949683(each), 40888277949684(case), Lot Number: 22JBR884; 295) DYNJ901559J, UDI-DI: 10193489235890(each), 40193489235891(case), Lot Number: 21CBH289; 296) DYNJ901559J, UDI-DI: 10193489235890(each), 40193489235891(case), Lot Number: 21EBR740; 297) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21ADB190; 298) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21BDA744; 299) DYNJ901812J, UDI-DI: 10193489814163(each), 40193489814164(case), Lot Number: 21FDB236; 300) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25ABS755; 301) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25DBB182; 302) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25EBK653; 303) DYNJ902711M, UDI-DI: 10198459229442(each), 40198459229443(case), Lot Number: 25GBR954; 304) DYNJ905066B, UDI-DI: 10195327098674(each), 40195327098675(case), Lot Number: 23BMB931; 305) DYNJ905066B, UDI-DI: 10195327098674(each), 40195327098675(case), Lot Number: 23BMF722; 306) DYNJ905269G, UDI-DI: 10193489337112(each), 40193489337113(case), Lot Number: 21ABH751; 307) DYNJ905269G, UDI-DI: 10193489337112(each), 40193489337113(case), Lot Number: 21BBD394; 308) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 22GBL545; 309) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23DBT141; 310) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23EBQ671; 311) DYNJ906322D, UDI-DI: 10193489969252(each), 40193489969253(case), Lot Number: 23FBE986; 312) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23BBA689; 313) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23BBB047; 314) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23DBM357; 315) DYNJ906452C, UDI-DI: 10195327228378(each), 40195327228379(case), Lot Number: 23FBP128; 316) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23HBO123; 317) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23IBO509; 318) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23JBT789; 319) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 23LBJ184; 320) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24ABD306; 321) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24BBN302; 322) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24CBS248; 323) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24DBT181; 324) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24FBJ836; 325) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24HBM625; 326) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24IBN246; 327) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24KBJ970; 328) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 24LBN196; 329) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25ABT788; 330) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25BBO090; 331) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25EBA889; 332) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25GBB683; 333) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25HBO792; 334) DYNJ906452D, UDI-DI: 10195327436636(each), 40195327436637(case), Lot Number: 25IBI139; 335) DYNJ906452F, UDI-DI: 10198459575303(each), 40198459575304(case), Lot Number: 25KBG657; 336) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23BBU275; 337) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23DBM354; 338) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23EBN137; 339) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23GBH151; 340) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23IBH652; 341) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23JBB062; 342) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23KBW414; 343) DYNJ906926B, UDI-DI: 10193489921366(each), 40193489921367(case), Lot Number: 23LBN387; 344) DYNJ907753F, UDI-DI: 10195327194482(each), 40195327194483(case), Lot Number: 22LBN185; 345) DYNJ907771C, UDI-DI: 10195327120399(each), 40195327120390(case), Lot Number: 22LBP170; 346) DYNJ908887B, UDI-DI: 10195327620158(each), 40195327620159(case), Lot Number: 24CBP744; 347) DYNJ908887B, UDI-DI: 10195327620158(each), 40195327620159(case), Lot Number: 24DBD796; 348) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24EBM187; 349) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24GBK492; 350) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24HBI648; 351) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24JBP902; 352) DYNJ908887C, UDI-DI: 10195327690793(each), 40195327690794(case), Lot Number: 24LBP187; 353) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23HMI269; 354) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23JMB528; 355) DYNJ909398A, UDI-DI: 10195327338190(each), 40195327338191(case), Lot Number: 23KME175; 356) DYNJ909799, UDI-DI: 10195327483210(each), 40195327483211(case), Lot Number: 23LBC746; 357) DYNJ909799, UDI-DI: 10195327483210(each), 40195327483211(case), Lot Number: 24HBE897; 358) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24BMA251; 359) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24CMI706; 360) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24DMI073; 361) DYNJ910009, UDI-DI: 10195327541149(each), 40195327541140(case), Lot Number: 24HMA202; 362) DYNJ910009A, UDI-DI: 10198459122507(each), 40198459122508(case), Lot Number: 24KMD940; 363) DYNJ910009B, UDI-DI: 10198459240133(each), 40198459240134(case), Lot Number: 25FMD839; 364) DYNJ910476, UDI-DI: 10198459009174(each), 40198459009175(case), Lot Number: 24FMJ212; 365) DYNJ910536, UDI-DI: 10198459029554(each), 40198459029555(case), Lot Number: 24GBA620; 366) DYNJ910536A, UDI-DI: 10198459054358(each), 40198459054359(case), Lot Number: 24IBE833; 367) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 24JBS568; 368) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 24LBM436; 369) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25BBD998; 370) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25DBM184; 371) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25EBD405; 372) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25EBL131; 373) DYNJ910542A, UDI-DI: 10198459142291(each), 40198459142292(case), Lot Number: 25FBV527; 374) DYNJ910589, UDI-DI: 10198459040894(each), 40198459040895(case), Lot Number: 24HBF839; 375) DYNJ910594, UDI-DI: 10198459041327(each), 40198459041328(case), Lot Number: 24HBF837; 376) DYNJ911567, UDI-DI: 10198459409004(each), 40198459409005(case), Lot Number: 25HBB593; 377) DYNJ911567, UDI-DI: 10198459409004(each), 40198459409005(case), Lot Number: 25HBJ084; 378) DYNJ911568, UDI-DI: 10198459409011(each), 40198459409012(case), Lot Number: 25HBB594; 379) DYNJ9682698S, UDI-DI: 10195327257170(each), 40195327257171(case), Lot Number: 23IDC305; 380) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21CKA647; 381) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21CKA812; 382) DYNJ9701096G, UDI-DI: 10193489779257(each), 40193489779258(case), Lot Number: 21DMF113; 383) DYNJ9701096I, UDI-DI: 10193489939651(each), 40193489939652(case), Lot Number: 22AMC654; 384) DYNJT3381, UDI-DI: 10198459325205(each), 40198459325206(case), Lot Number: 25EBK724; 385) DYNJT7540, UDI-DI: 10198459630989(each), 40198459630980(case), Lot Number: 26ABD262

Devices named in this recall

96 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98642).