SWAN-GANZ recallZ-2203-2026 · Becton, Dickinson and Company · initiated Apr 8, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton, Dickinson and Company or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.

FDA root cause category: Process change control

What the company is doing

On 4/8/2026, mailed correction notices were mailed to customers and distributors who were informed of the following: Consider the following to reduce the potential for hub breakage issues. Minimize manipulation of proximal injectate hub and lumen. Pay frequent attention to the (blue) proximal injectate lumen for leakage or breakage, and exchange with another catheter if the issue is encountered. Modify existing clinical workflows to minimize flexing and stress of (blue) proximal injectate lumen. For example: o Securing the catheter lumens to an IV arm board to mitigate bending/flexing of the lumen(s) o Consider using the stopcock on white hub to allow for intermittent transducer pressure tubing changes without twisting white hub on Swan Ganz (blue) proximal injectate lumen. o Prevent CO-Set syringe and proximal injectate lumen from dangling/hanging without support. o Change to automated cardiac output models to mitigate device stress. Providers and institutions may choose to utilize transthoracic or transesophageal echocardiography in combination with techniques to monitor central venous pressure via a central venous catheter in patients in need of right ventricle monitoring and management. Converting to Acumen IQ / FloTrac sensor can provide continuous minimally invasive hemodynamic monitoring from the arterial line. Consider using ClearSight cuff for continuous non-invasive hemodynamic monitoring. Distributors should forward the correction notice to customers who have purchased the impacted product. Complete and return the acknowledgement form via email to [email protected] Questions can be directed to the firm's Customer Service at (800) 424-3278

Recall details

Recall number
Z-2203-2026
Status
Open, Classified
Initiated by the company
Apr 8, 2026
Posted by the FDA
May 15, 2026
Recalling firm
Becton, Dickinson and Company
Quantity
183
Product code
DYG
FDA submissions
K231248
Distribution
Worldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan, Brazil, Colombia, Australia, New Zealand, Poland, Czech Republic, South Africa, Mauritius, Germany, Israel, Iran, Slovakia, Switzerland, Slovenia, Bulgaria, Cyprus, Saudi Arabia, United Arab Emirates (UAE), Kuwait, Malta, Bahrain, Qatar, Hungary, Morocco, Romania, Lebanon, Tunisia, Oman, Turkey, Austria, Belgium, Luxembourg, France, Netherlands, Italy, Spain, Canary Islands, Portugal, United Kingdom, Northern Ireland, Ireland, Denmark, Sweden, Iceland, Norway, Estonia, Finland, Greece, Chile, Puerto Rico, Canada.

Product description

Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P

Code information (models, lots, serial numbers)

Model-UDI-DI(Expiration): SGPT755P/00690103219388/65869018(6/24/2026), 66010501(9/23/2026), 67122784(10/28/2027); SGPT54/00690103219289/65956343(8/19/2026); SGPT64P/00690103219371

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K231248 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98603).