Products that contain the Pericare Wipes recallZ-3116-2026 · C.R. Bard Inc · initiated Jun 25, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from C.R. Bard Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.
FDA root cause category: Under Investigation by firm
What the company is doing
On 06/25/2026, an "URGENT Medical Device Recall" Letter was via regional courier services or other electronic means to inform customers that based on customer feedback, BD has identified that the CliniClean Pericare Wipes provided within Urinary Catheter Trays may be potentially contaminated with an unknown non-biological substance. Customer are instructed to: Do not use CliniClean Pericare Wipes contained within the affected products listed in Attachment 2. The CliniClean Pericare Wipes are the only impacted component within the tray and all remaining components can continue to be safely used. Users should source alternative, cleansing wipes for the impacted product lots listed in Attachment 2 and ensure they continue to follow their facility protocols for providing perineal care for patients with urinary catheters. If affected CliniClean Pericare Wipes have been used without issue, no additional treatment or clinical follow up is required. 1. For affected lots, remove and destroy wipes within the product package at the point of use and do not use. Source alternative cleansing wipes for all impacted urinary catheter trays listed in Attachment 2. 2. Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Correction. 3. Post this notice in all locations where inventory of these trays are stored. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the correction notification, whether or not you have any affected product, so that BD may confirm your receipt of this notification. For Questions - contact: BD Medical Information Services: Available 24/7 Select #2 when prompted 1-800-555-7422 or Email: [email protected] BD Technical Support 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted M-F 8am - 5pm CT or Email: [email protected]
Recall details
- Recall number
- Z-3116-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 25, 2026
- Posted by the FDA
- Sep 4, 2026
- Recalling firm
- C.R. Bard Inc
- Quantity
- 212580
- Product code
- MJC
- FDA submissions
- K040658, K910318, K984136
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY. The countries of Bahrain, Canada, Costs Rica, Guatemala, Mexico, Saudi Arabia, and United Arab Emirates.
Product description
Products that contain the Pericare Wipes: *Bardex I.C. Complete Care Foley Catheter Tray REF: 300316A UDI codes: 00801741024344 10801741024341 *Bardex I.C. Complete Care Foley Catheter Tray and Urine Meter REF: 303314A 303316A 303318A UDI code: 00801741024252 10801741024259 00801741024269 10801741024266 00801741024276 10801741024273 *SureStep" Foley Tray System Bardex IC Complete Care REF: A300314A UDI codes: 00801741073724 10801741073721 REF: A300314AC (Canada Only) UDI-DI code: 00801741143243 *SureStep" Foley Tray System Bardex IC Complete Care REF: A300316A UDI code: 00801741073731 10801741073738 *SureStep" Foley Tray System Bardex IC Complete Care REF: A300318A UDI code: 00801741073748 10801741073745 *SureStep" Foley Tray System Bardex IC Complete Care REF: A303314A A303316A A303318A UDI code: 00801741073786 10801741073783 00801741073793 10801741073790 00801741073809 10801741073806 *SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A303414A UDI code: 00801741073816 10801741073813 *(416A)SureStep" Foley Tray System Bardex IC Complete Care (714A/716A/718A)SureStep" Foley Tray System Bardex IC Complete Care (914)SureStep" Foley Tray System REF: A304716A A304714A A304716A A304718A UDI code: 00801741104206 10801741104203 00801741104268 10801741104265 00801741104299 10801741104296 00801741100376 10801741100373 *SureStep" Foley Tray System Bardex IC Complete Care REF: A319416A UDI code: 00801741100307 10801741100304 *SureStep" Foley Tray System Bardex I.C. Complete Care Foley Catheter REF: A319416AM UDI code: 00801741073861 10801741073868 *SureStep" Foley Tray System Bardex IC Complete Care REF: A319418A UDI code: 00801741100314 10801741100311 SureStep" Foley Tray System Bardex IC Complete Care REF: A903316A UDI code: 00801741096600 10801741096607 ***** SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A300414A UDI code: 00801741073755 10801741073752 SureStep" Foley Tray System Bardex I.C. Complete Care Foley Catheter REF: A319414AM UDI code: 00801741073854 10801741073851 SureStep" Foley Tray System Bardex I.C. Complete Care Foley Catheter REF: A319418AM UDI code: 00801741073878 10801741073875 SureStep" Foley Tray System Bardex IC Complete Care REF: A903318A UDI code: 00801741104657 10801741104654 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Bardex¿ I.C. Complete Care¿ Foley Catheter Tray 300316A / 00801741024344¿ 10801741024341 Lot Number: NGKT3657 Bardex¿ I.C. Complete Care¿ Foley Catheter Tray and Urine Meter 303314A / 00801741024252¿ 10801741024259 Lot Number: NGKU4153 303316A / 00801741024269 10801741024266 Lot Number: NGKW2883 303318A / 00801741024276¿ 10801741024273 Lot Number: NGKU0966 *SureStep" Foley Tray System Bardex¿ IC Complete Care¿ A300314A / 00801741073724¿ 10801741073721 Lot Number: NGKU1020 A300314AC (Canada Only) / 00801741143243 Lot Number: **** SureStep" Foley Tray System Bardex¿ IC Complete Care¿ A300316A / 00801741073731 10801741073738 Lot Numbers: NGKT2448 NGKT3221 NGKT3288 NGKT3301 NGKT3315 NGKU2921 NGKV0293 SureStep" Foley Tray System Bardex¿ IC Complete Care¿ A300318A / 00801741073748¿ 10801741073745 Lot Number: NGKT2493 SureStep" Foley Tray System Bardex¿ IC Complete Care¿ A303314A / 00801741073786¿ 10801741073783 Lot Numbers: NGKT3222 NGKV0249 A303316A / 00801741073793¿ 10801741073790 Lot Numbers: NGKU0894 NGKU0947 NGKU0973 NGKU1021 NGKU1748 NGKU1826 NGKU1830 NGKU2034 NGKU2035 NGKU2996 NGKU3042 NGKU3969 NGKU4037 NGKU4077 NGKU4078 NGKU4135 NGKU4713 NGKU4714 NGKU4720 NGKU4721 NGKU4722 NGKU4727 NGKV2390 NGKV3390 A303318A / 00801741073809¿ 10801741073806 Lot Numbers: NGKT3223 NGKU4136 SureStep" Foley Tray System Lubri-Sil¿ IC Complete Care¿ A303414A / 00801741073816¿ 10801741073813 Lot Number: NGKU1022 (416A)SureStep" Foley Tray System Bardex¿ IC Complete Care¿ (714A/716A/718A)SureStep" Foley Tray System Bardex¿ IC Complete Care¿ (914)SureStep" Foley Tray System A304716A / 00801741104206 10801741104203 Lot Number: NGKT2496 A304714A / 00801741104268 10801741104265 Lot Number: NGKU2923 A304716A / 00801741104299 10801741104296 Lot Numbers: NGKU1833 NGKU4729 A304718A / 00801741100376¿ 10801741100373 Lot Numbers: NGKT3224 NGKU3971 SureStep" Foley Tray System Bardex¿ IC Complete Care¿ A319416A / 00801741100307¿ 10801741100304 Lot Number: NGKS1605 SureStep" Foley Tray System Bardex¿ I.C. Complete Care Foley Catheter A319416AM / 00801741073861¿ 10801741073868 Lot Numbers: NGKT2390 NGKT2410 NGKT3225 NGKT3319 NGKU0951 NGKU0977 SureStep" Foley Tray System Bardex¿ IC Complete Care¿ A319418A / 00801741100314¿ 10801741100311 Lot Number: NGKT2411 SureStep" Foley Tray System Bardex¿ IC Complete Care¿ A903316A / 00801741096600¿ 10801741096607 Lot Numbers: NGKT3392 NGKU2020 **** SureStep" Foley Tray System Lubri-Sil¿ IC Complete Care A300414A / 00801741073755 10801741073752 Lot Number: NGKV3524 SureStep" Foley Tray System Bardex¿ I.C. Complete Care Foley Catheter A319414AM / 00801741073854 10801741073851 Lot Number: NGKV3487 SureStep" Foley Tray System Bardex¿ I.C. Complete Care Foley Catheter A319418AM / 00801741073878 10801741073875 Lot Number: NGKV4983 SureStep" Foley Tray System Bardex¿ IC Complete Care A903318A / 00801741104657 10801741104654 Lot Number: NGKV2397
Devices named in this recall
23 devices in the FDA device registry whose identifier appears in the recall's code information.
- SureStep™ Foley Tray System Bardex® I.C. Complete Care Foley Catheter A319416AMC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES3 recalls18 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A303316AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES2 recalls62 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A304716AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES2 recalls9 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A300316AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES2 recalls7 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A303318AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES2 recalls4 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A300314AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES2 recalls2 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A300318AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES2 recalls1 report
- SureStep™ Foley Tray System Bardex® IC Complete Care® A303314AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall12 reports
- Bardex® I.C. Complete Care® Foley Catheter Tray and Urine Meter 303316AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall9 reports
- Bardex® I.C. Complete Care® Foley Catheter Tray 300316AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall7 reports
- SureStep™ Foley Tray System Bardex® I.C. Complete Care Foley Catheter A319414AMC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall7 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A304416AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall7 reports
- SureStep™ Foley Tray System Lubri-Sil® IC Complete Care® A303414AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall6 reports
- Bardex® I.C. Complete Care® Foley Catheter Tray and Urine Meter 303318AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall2 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A903316AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall2 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A319416AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall1 report
- SureStep™ Foley Tray System Bardex® IC Complete Care® A304718AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall1 report
- Bardex® I.C. Complete Care® Foley Catheter Tray and Urine Meter 303314AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall0 reports
- SureStep™ Foley Tray System Lubri-Sil® IC Complete Care® A300414AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall0 reports
- SureStep™ Foley Tray System Bardex® I.C. Complete Care Foley Catheter A319418AMC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall0 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A319418AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall0 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care® A304714AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall0 reports
- SureStep™ Foley Tray System Bardex® IC Complete Care A903318AC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall0 reports
Product lines
- SureStep Foley Tray System Bardex IC Complete CareC. R. Bard, Inc. · 14 models · 108 reports · 7 recalls
- Bardex I.C. Complete Care Foley Catheter Tray and Urine MeterC. R. Bard, Inc. · 3 models · 11 reports · 1 recall
- SureStep Foley Tray System Lubri-Sil IC Complete CareC. R. Bard, Inc. · 14 models · 105 reports · 5 recalls
Other products in the same recall event
The FDA groups related products into one recall event (event 98840).
- Z-3114-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: Bardex Lubricath Foley Catheter Tray REF: 899914 899916 899918 UDI codes: 00801741024801 10801741024808 00801741024818 10801741024815 008…
- Z-3115-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: Bard Foley Tray With Lubri-Sil All-Silicone Foley Catheter REF: 947314 947316 947318 UDI codes: 00801741024924 10801741024921 00801741024…
- Z-3117-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: SureStep Intermittent Catheter Tray REF: INTP16 INTS14 INTS14C INTS16 INTS16C PEDINTS06 PEDINTS08 PEDINTS10 UDI code: 00801741183300 1080…
- Z-3118-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: SureStep" Foley Tray System Add-A-Foley Catheter REF: A153200 UDI code: 00801741121067 10801741121064 SureStep" Foley Tray System Comple…
- Z-3119-2026 · Open, ClassifiedProducts that contain the Pericare Wipes: (Canada Only) SureStep" Foley Tray System Bardex IC Complete Care REF: A300314AC UDI-DI code: 00801741143243 SureStep" Foley Tray System …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.