NA TCN-10-3MRecalls & reported problems
Boston Scientific Neuromodulation Corporation · Nitinol TC Electrode, 10cm, 3m Cable · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00813250017502), received between Aug 18, 2017 and May 5, 2022 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Material Discolored” (7 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Apr 29, 2026Class IIOpen, Classified
Reusable electrodes may not meet expected performance levels.
Z-2403-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- May 30, 2019Class IITerminated
The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
Z-1695-2020 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Discolored7
- Break4
- Contamination /Decontamination Problem3
- Adverse Event Without Identified Device or Use Problem1
- Defective Device1
- Insufficient Heating1
Patient problems
- No Known Impact Or Consequence To Patient7
- No Clinical Signs, Symptoms or Conditions1
- No Patient Involvement1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryMay 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT BLOOD SEEPED INTO WHERE THE PROBE ATTACHES TO THE STAINLESS HEAD OF THE ELECTRODE. THE BLOOD COULD NOT BE REMOVED AND KEPT SEEPING OUT DESPITE ATTEMPTS OF STERILIZATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14289701 ↗MalfunctionOct 22, 2019Insufficient Heating; Defective Device; Insufficient Heating; Defective Device
THE DEVICE ANALYSIS PERFORMED ON THE RETURNED ELECTRODE REVEALED THAT THE SHAFT OF THE ELECTRODE WAS FOUND TO BE COMPLETELY SEPARATED FROM THE HUB.
Patient problems: No Patient Involvement
MAUDE report 9219134 ↗MalfunctionAug 9, 2018Material Discolored; Material Discolored
THE DEVICE ANALYSIS PERFORMED ON THE ELECTRODE REVEALED THAT IT HAD DISCOLORED EPOXY AND A CHIP OUT. THE COLOR OF THE NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED. THIS FAILURE MECHANISM IS UNRELATED TO THE REPORTED COMPLAINT AND IS DUE TO EXTERNAL MECHANICAL FORCE AND THERMAL STRESS ASSOCIATED WITH MULTIPLE AUTOCLAVE CYCLES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7770233 ↗MalfunctionOct 3, 2017Break; Material Discolored; Contamination /Decontamination Problem; Break; Material Discolored; Contamination /Decontamination Problem
DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODES SHOWED THAT THE EPOXY HAD A CHIP OUT AND DISCOLORATION WAS NOTED. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES, IT BECOMES BRITTLE AND DISCOLORED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6911523 ↗MalfunctionSep 14, 2017Break; Material Discolored; Break; Material Discolored
DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODES FOUND THAT THE EPOXY HAD A CHIP OUT AND WAS ALSO DISCOLORED. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6864449 ↗MalfunctionSep 8, 2017Break; Material Discolored; Contamination /Decontamination Problem; Break; Material Discolored; Contamination /Decontamination Problem
THE DEVICE ANALYSIS PERFORMED SHOWED THAT THE EPOXY HAS CHIP OUT AND BROWN DISCOLORATION. IN ADDITION, THE ELECTRODE FAILED THE TEMPERATURE TEST AS IT READS NO TEMPERATURE. IT HAD A DEFECTIVE THERMOCOUPLE CIRCUIT. THE OPEN CIRCUIT WAS MEASURED AT THE CONNECTOR. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES, THE EPOXY BECOMES BRITTLE AND DISCOLORED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6853580 ↗MalfunctionAug 29, 2017Material Discolored; Material Discolored
DEVICE ANALYSIS PERFORMED SHOWED THAT BOTH ELECTRODES HAVE DISCOLORED AND CHIPPED OUT EPOXY. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES, THE EPOXY BECOMES BRITTLE AND DISCOLORED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6828412 ↗MalfunctionAug 24, 2017Break; Material Discolored; Contamination /Decontamination Problem; Break; Material Discolored; Contamination /Decontamination Problem
DEVICE EVALUATION FOR THE RETURNED ELECTRODES REVEALED THAT THE EPOXY WAS DISCOLORED AND HAD A CHIP OUT. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES THE EPOXY BECOMES BRITTLE AND DISCOLORED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6819185 ↗MalfunctionAug 18, 2017Material Discolored; Material Discolored
DEVICE ANALYSIS ON THREE ELECTRODES CONFIRMED THAT THE EPOXY WAS VERY DARK.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6803513 ↗
Other PROBE, RADIOFREQUENCY LESION devices
- NA TCN-15Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION2 recalls10 reports
- NA CSK-TC10Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION1 recall1 death58 reports
- NA CLASSICSt. Jude Medical, Inc. · PROBE, RADIOFREQUENCY LESION1 recall40 reports
- Intracept RLV 0055Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall30 reports
- Intracept RLV 0045Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall1 death23 reports
- Intracept RLV0065Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall14 reports
- NA CSK-TC5Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION1 recall3 reports
- Intracept RLV 0075Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page